| Date | 
Time (24h) | 
Guard (V) | 
Sense (V) | 
Field (V) | 
Sense ([math]\mu[/math]A) | 
Field ([math]\mu[/math]A) | 
comments
 | 
| 05/28/13 | 
14:08 | 
100 | 
100 | 
-100 | 
0.02 | 
0.00 | 
Confirmed with initial setting of 30/40/50V that there are no crossed connections between wire types and that voltage supply is connected correctly.
 | 
| 05/28/13 | 
16:34 | 
536 | 
1500 | 
-750 | 
45.5 | 
45.5 | 
Someone upped the voltage to full the full operating level without recording it. This is the current reading right now.
 | 
| 05/28/13 | 
17:56 | 
536 | 
1500 | 
-750 | 
41.5 | 
41.6 | 
 | 
| 05/29/13 | 
07:56 | 
536 | 
1500 | 
-750 | 
22.5 | 
22.6 | 
 | 
| 05/29/13 | 
15:02 | 
536 | 
1500 | 
-750 | 
17.9 | 
18.0 | 
 | 
| 05/29/13 | 
18:07 | 
536 | 
1500 | 
-750 | 
16.7 | 
16.8 | 
 | 
| 05/30/13 | 
07:55 | 
536 | 
1500 | 
-750 | 
10.6 | 
10.7 | 
 | 
| 05/30/13 | 
12:51 | 
536 | 
1500 | 
-750 | 
9.5 | 
9.6 | 
 | 
| 05/30/13 | 
17:23 | 
536 | 
1500 | 
-750 | 
8.7 | 
8.8 | 
 | 
| 06/03/13 | 
08:23 | 
536 | 
1500 | 
-750 | 
1.7 | 
1.8 | 
 | 
| 06/04/13 | 
07:42 | 
536 | 
1500 | 
-750 | 
1.4 | 
1.5 | 
 | 
| 06/05/13 | 
12:01 | 
536 | 
1500 | 
-750 | 
1.8 | 
1.9 | 
had to turn off HV to organize cables
 | 
| 06/05/13 | 
17:56 | 
536 | 
1500 | 
-750 | 
1.8 | 
1.9 | 
 | 
| 06/06/13 | 
07:56 | 
536 | 
1500 | 
-750 | 
1.8 | 
1.9 | 
seem to be stuck at this current draw?
 | 
| 06/10/13 | 
17:29 | 
536 | 
1500 | 
-750 | 
1.5 | 
1.6 | 
I think we ran out of gas on the weekend, the windows were deflated a bit but now they deflected outward
 | 
| 06/11/13 | 
07:50 | 
536 | 
1500 | 
-750 | 
3.2 | 
3.3 | 
current doubles?  Humidity dropped from 40 to 25%.
 | 
| 06/12/13 | 
13:39 | 
536 | 
1500 | 
-750 | 
3.6 | 
3.7 | 
Humidity rose from 25 to 40%.
 | 
| 06/13/13 | 
11:00 | 
536 | 
1500 | 
-750 | 
4.5 | 
4.6 | 
Humidity is 40 +/- 5 %.
 | 
| 06/14/13 | 
08:56 | 
536 | 
1500 | 
-750 | 
3.5 | 
3.6 | 
 | 
| 06/17/13 | 
07:58 | 
536 | 
1500 | 
-750 | 
3.6 | 
3.7 | 
 | 
Why is the current increasing?
is the current going to ground ?
Try raising the sense to 1750 and lowering
the field to 500 (same voltage difference),
and see if the sense current still equals the
field current - want to know if the current
is inside the chamber from wire to wire or if
it's a current to ground, for example ...
If current is from field to sense:
if it is current from field
to sense the current should increase exponentially
with voltage, doubling with 100V increase; i.e.
2250 -> 2350.  You could check to see if this
is the case.  If so, you could then try to
locate where the problem is: which wire(s) are
responsible?
Increasing Sense
| Date | 
Time (24h) | 
Guard (V) | 
Sense (V) | 
Field (V) | 
Sense ([math]\mu[/math]A) | 
Field ([math]\mu[/math]A) | 
comments
 | 
| 06/17/13 | 
12:20 | 
536 | 
1510 | 
-750 | 
8 | 
8.1 | 
 | 
| 06/17/13 | 
12:16 | 
536 | 
1520 | 
-750 | 
10.7 | 
10.8 | 
 | 
| 06/17/13 | 
12:34 | 
536 | 
1530 | 
-750 | 
22.3 | 
22.4 | 
 | 
| 06/17/13 | 
12:43 | 
536 | 
1540 | 
-750 | 
37.3 | 
37.4 | 
 | 
| 06/17/13 | 
15:53 | 
536 | 
1540 | 
-750 | 
45.2 | 
45.3 | 
 | 
| 06/17/13 | 
15:55 | 
536 | 
1550 | 
-750 | 
54.9 | 
55.1 | 
 | 
Lower the field to -500 V
| Date | 
Time (24h) | 
Guard (V) | 
Sense (V) | 
Field (V) | 
Sense ([math]\mu[/math]A) | 
Field ([math]\mu[/math]A) | 
comments
 | 
| 06/17/13 | 
15:55 | 
536 | 
1550 | 
-750 | 
54.9 | 
55.1 | 
 | 
| 06/17/13 | 
15:58 | 
536 | 
1550 | 
-740 | 
45.9 | 
46.1 | 
 | 
| 06/17/13 | 
15:59 | 
536 | 
1550 | 
-730 | 
37.6 | 
37.9 | 
 | 
| 06/17/13 | 
14:00 | 
536 | 
1550 | 
-710 | 
24.2 | 
24.3 | 
 | 
| 06/17/13 | 
14:02 | 
536 | 
1550 | 
-690 | 
13.9 | 
14.0 | 
 | 
| 06/17/13 | 
14:03 | 
536 | 
1550 | 
-650 | 
3.8 | 
3.9 | 
 | 
| 06/17/13 | 
14:04 | 
536 | 
1550 | 
-600 | 
1.33 | 
1.41 | 
 | 
| 06/17/13 | 
14:05 | 
536 | 
1550 | 
-550 | 
0.88 | 
0.97 | 
 | 
| 06/17/13 | 
14:06 | 
536 | 
1550 | 
-500 | 
0.66 | 
0.74 | 
 | 
Increase sense when Field is at -500 V
| Date | 
Time (24h) | 
Guard (V) | 
Sense (V) | 
Field (V) | 
Sense ([math]\mu[/math]A) | 
Field ([math]\mu[/math]A) | 
comments
 | 
| 06/17/13 | 
14:06 | 
536 | 
1550 | 
-500 | 
0.66 | 
0.74 | 
 | 
| 06/17/13 | 
14:10 | 
536 | 
1650 | 
-500 | 
1.5 | 
1.6 | 
 | 
| 06/17/13 | 
14:13 | 
536 | 
1750 | 
-500 | 
16.2 | 
16.3 | 
 | 
SL2 current draw search
| Date | 
Time (24h) | 
Guard (V) | 
Sense (V) | 
Field (V) | 
Sense ([math]\mu[/math]A) | 
Field ([math]\mu[/math]A) | 
comments
 | 
| 06/18/13 | 
08:57 | 
536 | 
1550 | 
-550 | 
0.59 | 
0.69 | 
no voltage to HVB1
 | 
| 06/18/13 | 
09:00 | 
536 | 
1550 | 
-500 | 
0.46 | 
0.55 | 
no voltage to HVB1
 | 
| 06/18/13 | 
09:12 | 
536 | 
1650 | 
-500 | 
1.1 | 
1.2 | 
no voltage to HVB1
 | 
| 06/18/13 | 
09:17 | 
536 | 
1750 | 
-500 | 
9.5 | 
9.6 | 
no voltage to HVB1, less current measured
 | 
| 06/18/13 | 
10:31 | 
536 | 
1550 | 
-500 | 
0.32 | 
0.42 | 
no voltage to HVB1 and HVB2
 | 
| 06/18/13 | 
10:40 | 
536 | 
1650 | 
-500 | 
1.0 | 
1.1 | 
no voltage to HVB1 and HVB2
 | 
| 06/18/13 | 
10:45 | 
536 | 
1750 | 
-500 | 
8.8 | 
8.9 | 
no voltage to HVB1 and HVB2
 | 
| 06/18/13 | 
11:07 | 
536 | 
1550 | 
-500 | 
0.32 | 
0.42 | 
no voltage to HVB1 and HVB3
 | 
| 06/18/13 | 
11:11 | 
536 | 
1650 | 
-500 | 
1.0 | 
1.1 | 
no voltage to HVB1 and HVB3
 | 
| 06/18/13 | 
11:16 | 
536 | 
1750 | 
-500 | 
6.8 | 
6.9 | 
no voltage to HVB1 and HVB3
 | 
| 06/18/13 | 
11:31 | 
536 | 
1550 | 
-500 | 
0.34 | 
0.43 | 
no voltage to HVB1,HVB3,and HVB4
 |