Testing procedure
Things that may cause "noise" that are actually due to your testing setup:
- STB boards not securely attached to chamber. Make sure center screws (ones attaching directly to the chamber, not the nylon stand-offs) are tightened to 14 in. lb's.
- Scope's termination set to 1Mohm. Set the scope to 50ohm termination.
- Lower bandwidth of 1 Ghz scope to 200 Mhz. This attenuates peak height, but also rejects high frequency noise and reduces the noise level from 55-65 mV to ~11-15 mV.
Super Layer 1
STB-1 G1202220
Layer |
Wire |
Channel |
Date |
Comments
|
|
|
U1 |
|
This channel presents higher noise level than the other ones, which makes difficult to get a signal with amplitude below 30mV. This behaviour extends to all the first channels of each connector in the board. however, if a high enough trigger level is set in the scope, it is always possible to get a signal in these channels.
|
STB-2
STB-3
STB-4
STB-5
STB-6
STB-7
Super Layer 2
STB-1
STB-2
STB-3
STB-4
STB-5
STB-6
STB-7
CLAS12_R1_S6_QA