Accelerator Configuration
- Accelerator Settings
| beam energy | 6.8 MeV | 
| Beam type: | Electron | 
| I_average: | 0.3 nA | 
| I_peak: | 0.05[math] \mu[/math] A 0.006 duty factor | 
| pulse width: | 10 [math]\mu[/math]s  (it seems on 200 [math] \mu[/math]s is possible?) | 
| Rate: | 60 Hz | 
| PEPPO Spectrometer: | 151.3 A | 
Set HV to Detectors
|  | Signal channel | HV dac | HV (V) | 
| Detectors #1 | A1 | 2.4 | 960 | 
| Detectors #2 | A2 | 3.0 | 1200 | 
How to turn HV On
Scope Pictures
CH1: Beam Sync
CH2: A1
CH3: A2
| Target | Pictures | 
| No Target |     | 
| Chromox Target |     | 
Annihilation Counter Runs
- Signal Definitions for Trigger logic
- SYNC:=60 Hz sync pulse 200 micro seconds long
- Det1:= Annihilation Counter #1 (HV1)
- Det2:=Annihilation Counter #2 (HV2)
| Run# | run mode | Run length (sec) | Beam | Target | Trig Logic | VME discriminator Threshold (mV) | HV1 (setting/V) | HV2 (setting/V) | Deadtime |  | 
| 2902 | sample | 0 | e[math]^-[/math] | Chromox | (Det1 OR Det2) | 50 | 2.4/960 | 3/1200 | 93.1818% [math]\pm[/math] 10.3874% |  | 
| 2903 | sample | 203 | e[math]^-[/math] | Empty | (Det1 OR Det2) | 50 | 2.4/960 | 3/1200 | 84.9788% [math]\pm[/math]0.106622% |  | 
| 2904 | sample | 452 | e[math]^-[/math] | Empty | Det1 AND Det2 | 50 | 2.4/960 | 3/1200 | 12.2563% [math]\pm[/math] 2.32772% |  | 
| 2905 | sample | 556 | e[math]^-[/math] | Chromox | Det1 AND Det2 | 50 | 2.4/960 | 3/1200 | 3.16901% [math]\pm[/math] 4.2301% |     | 
| 2908 | sample | 150 | off | out | (Det1 or Det2) | 50 | 2.4/960 | 3/1200 | 74.0667% [math]\pm[/math]  0.157206% |  | 
| 2909 | sample | 1118 | off | out | (Det1 AND Det2) | 50 | 2.4/960 | 3/1200 | 3.57143% [math]\pm[/math]26.9725% |  | 
data analysis
Negative Energy ?
This run 2902, Int_Ch14 histogram. Why there is negative? 
 
Pulses
run 2905
1st 100 puleses
 
pulse that only partially read-in: 
 
 
first 10 pulses, some pulses that only partially read-in: 
 
looks like some pulses are only partially read-in. We need to delay signal more?
One possible cut may to determine if the signal is increasing within the first 100 - 300 $Iterations.  This would be like creating a leading edge discriminator in software.
T->Draw("CH14:Iteration$","Iteration$<300 & CH14<100","L",10,1) 
looks like it just won't draw. I need to know the to way exclude that event does not satisfy "Iteration$<300 & CH14<100" requirement. 
 
Why CH14 histogram is all positive, but int_14 histogram has negatives?
 
 
TF Analysis
Using the following ROOT programs
Media:03212012_T.C.txt
Media:03212012_T.h.txt
| Run# | run mode | Run length (sec) | Beam | Target | Trig Logic | VME discriminator Threshold (mV) | HV1 (setting/V) | HV2 (setting/V) | Deadtime |  | 
| 2902 | sample | 0 | e[math]^-[/math] | Chromox | (Det1 OR Det2) | 50 | 2.4/960 | 3/1200 | 93.1818% [math]\pm[/math] 10.3874% |  | 
| 2903 | sample | 203 | e[math]^-[/math] | Empty | (Det1 OR Det2) | 50 | 2.4/960 | 3/1200 | 84.9788% [math]\pm[/math]0.106622% |  | 
| 2904 | sample | 452 | e[math]^-[/math] | Empty | Det1 AND Det2 | 50 | 2.4/960 | 3/1200 | 12.2563% [math]\pm[/math] 2.32772% |   | 
| 2905 | sample | 556 | e[math]^-[/math] | Chromox | Det1 AND Det2 | 50 | 2.4/960 | 3/1200 | 3.16901% [math]\pm[/math] 4.2301% |   | 
| 2908 | sample | 150 | off | out | (Det1 or Det2) | 50 | 2.4/960 | 3/1200 | 74.0667% [math]\pm[/math]  0.157206% |  | 
| 2909 | sample | 1118 | off | out | (Det1 AND Det2) | 50 | 2.4/960 | 3/1200 | 3.57143% [math]\pm[/math]26.9725% |  | 
Run 2904 and 2905 coincidence overlay (NOT Deadtime corrected)