05/12/2011
The powers are:
PS1 and PS2: -1000 V (the old 2-channel TENNELEC TC952) n's: -1100 V (the new 2-channel C.A.E.N N471)
The thresholds are:
PS1 and PS2: -80 mV n's: -200 mV
The discriminator widths are:
PS1: -20 ns PS2: -10 ns n's: -20 ns
Let's do the random coincidence cosmic ray test. PS1 and PS2 is about 30 cm away from each other. And let's look on the coincidence signal between PS1 and PS2:
| total counts | time | counts/min | counts/sec | |
|---|---|---|---|---|
| PS1 | [math]\displaystyle{ 10247\pm101.2 }[/math] | 70 min | [math]\displaystyle{ 146.4\pm1.44 }[/math] | [math]\displaystyle{ 2.44\pm0.02 }[/math] |
| PS1 | [math]\displaystyle{ 12969\pm113.9 }[/math] | 70 min | [math]\displaystyle{ 185.3\pm1.63 }[/math] | [math]\displaystyle{ 3.09\pm0.03 }[/math] |
| PS1+PS2 | [math]\displaystyle{ 3\pm1.7 }[/math] | 70 min | [math]\displaystyle{ 0.043\pm0.024 }[/math] | [math]\displaystyle{ 0.0007\pm0.0004 }[/math] |
And below my scope pictures of this rear event:
Let's do the 2n correlation test with CF252 source. The voltages and thresholds are the same. Set up is like that:
- the distance between PS1 and PS2 is 12.5 cm and they faced to the source
- the CF252 source in the middle
| total counts | time | counts/min | counts/sec | |
|---|---|---|---|---|
| PS1 | [math]\displaystyle{ 255977\pm505.9 }[/math] | 10 min | [math]\displaystyle{ 25597.7\pm50.6 }[/math] | [math]\displaystyle{ 426.63\pm0.84 }[/math] |
| PS1 | [math]\displaystyle{ 421917\pm649.5 }[/math] | 10 min | [math]\displaystyle{ 42191.7\pm64.9 }[/math] | [math]\displaystyle{ 703.19\pm1.08 }[/math] |
| PS1+PS2 | [math]\displaystyle{ 1224\pm34.9 }[/math] | 10 min | [math]\displaystyle{ 122.4\pm3.5 }[/math] | [math]\displaystyle{ 2.04\pm0.06 }[/math] |