The powers are:

PS1 and PS2:  -1000 V (the old 2-channel TENNELEC TC952)
n's: -1100 V (the new 2-channel C.A.E.N N471)

The thresholds are:

PS1 and PS2:  -80 mV
n's: -200 mV

The discriminator widths are:

PS1: -20 ns
PS2: -10 ns
n's: -20 ns


Let's do the random coincidence cosmic ray test. PS1 and PS2 is about 30 cm away from each other. And let's look on the coincidence signal between PS1 and PS2:

total counts time counts/min counts/sec
PS1 [math]\displaystyle{ 10247\pm101.2 }[/math] 70 min [math]\displaystyle{ 146.4\pm1.44 }[/math] [math]\displaystyle{ 2.44\pm0.02 }[/math]
PS1 [math]\displaystyle{ 12969\pm113.9 }[/math] 70 min [math]\displaystyle{ 185.3\pm1.63 }[/math] [math]\displaystyle{ 3.09\pm0.03 }[/math]
PS1+PS2 [math]\displaystyle{ 3\pm1.7 }[/math] 70 min [math]\displaystyle{ 0.043\pm0.024 }[/math] [math]\displaystyle{ 0.0007\pm0.0004 }[/math]


And below my scope pictures of this rear event:


Let's do the 2n correlation test with CF252 source. The voltages and thresholds are the same. Set up is like that:

the distance between PS1 and PS2 is 12.5 cm and they faced to the source
the CF252 source in the middle
total counts time counts/min counts/sec
PS1 [math]\displaystyle{ 255977\pm505.9 }[/math] 10 min [math]\displaystyle{ 25597.7\pm50.6 }[/math] [math]\displaystyle{ 426.63\pm0.84 }[/math]
PS1 [math]\displaystyle{ 421917\pm649.5 }[/math] 10 min [math]\displaystyle{ 42191.7\pm64.9 }[/math] [math]\displaystyle{ 703.19\pm1.08 }[/math]
PS1+PS2 [math]\displaystyle{ 1224\pm34.9 }[/math] 10 min [math]\displaystyle{ 122.4\pm3.5 }[/math] [math]\displaystyle{ 2.04\pm0.06 }[/math]