05/12/2011

From New IAC Wiki
Revision as of 19:45, 12 May 2011 by Shaproma (talk | contribs)
Jump to navigation Jump to search

Go Back


The powers are:

PS1 and PS2:  -1000 V (the old 2-channel TENNELEC TC952)
n's: -1100 V (the new 2-channel C.A.E.N N471)

The thresholds are:

PS1 and PS2:  -80 mV
n's: -200 mV

The discriminator widths are:

PS1: -20 ns
PS2: -10 ns
n's: -20 ns


Let's do the random coincidence cosmic ray test. PS1 and PS2 is about 30 cm away from each other. And let's look on the coincidence signal between PS1 and PS2:

total counts time counts/min counts/sec
PS1 10247[math]\pm[/math]101.2 70 min 146.4[math]\pm[/math]1.44 2.44[math]\pm[/math]0.02
PS1 12969[math]\pm[/math]113.9 70 min
PS1+PS2 3[math]\pm[/math]1.7 70 min