MiniPrototype CAEN V792 Calibration
I injected a signal into the Drift chamber preamps and read it out using the CAEN V792 ADC.
12/11/07
7 mV Signal
Here is a scope picture of the signal coming out of the preamp and going into the ADC (Blue) as well as the integration gate used. The signal is about 7 mv as shown below. There is a 2 mv peak to peak noise level which has structure.
The data was then collected using coda and recorded as run #189. A histogram of channel #1 is shown below.
20 mV signal
Reducing the attenuation, a 20 mV signal was sent into the ADC.
The data was then collected using coda and recorded as run #190. A histogram of channel #1 is shown below.
100 mV signal
Reducing the attenuation, a 100 mV signal was sent into the ADC.
The data was then collected using coda and recorded as run #191. A histogram of channel #1 is shown below.
Calibration
Input (mV) | Chan 1 | Chan 2 | Chan 3 | Chan 4 | Chan 5 | Chan 6 | Chan 7 | Chan 8 | Chan 9 | Chan 10 | Chan 11 | Chan 12 | Chan 13 | Chan 14 | Chan 15 | Chan 16 |
7 | ||||||||||||||||
20 | ||||||||||||||||
100 |