Below are the resulting transparency measurements for PbF2 block # 16
12/3/07
- Accelerator Settings
beam energy |
20 MeV
|
Rep Rate: |
30 Hz
|
I_peak: |
60 mA/pulse
|
pulse width: |
0.1 [math]\mu[/math]s (100 ns)
|
beam start: |
15:39
|
beam stop: |
16:39
|
- Pick-up Coil Pulse Recording (PbF2 crystal at z=30cm away)
O-scope pulses were recorded on the pick-up coil with dimensions in inches and the intensity measured in Volt [math]\times[/math] ns
@16:00
@16:05
@16:10
@16:15
@16:20
@16:25
@16:30
@16:39
- Dose rate measurements (@1643 hours)
[math]\gamma[/math]-radiation |
3.5 mR/hr
|
[math]\beta[/math]-radiation: |
12 mR/hr
|
- Dose rate measurements (@1532 hours)
[math]\gamma[/math]-radiation |
3 mR/hr
|
[math]\beta[/math]-radiation: |
7 mR/hr
|
12/4/07
Dose rate measurements (@0957 hours)
Under Blue light for 19 hours before next irradiation.
12/5/07
- Accelerator Settings
beam energy |
20 MeV
|
Rep Rate: |
30 Hz
|
I_peak: |
60 mA/pulse
|
pulse width: |
0.1 [math]\mu[/math]s (100 ns)
|
beam start: |
08:38
|
beam stop: |
|
- Pick-up Coil Pulse Recording (PbF2 crystal at z=30cm away)
O-scope pulses were recorded on the pick-up coil with dimensions in inches and the intensity measured in Volt [math]\times[/math] ns
@8:57
@9:07
@9:17
@9:27
@9:37
@9:47
@9:57
@10:07
@10:17