Pump downs
Date of pump down |
Components |
Reason
|
05/2021 |
X-Cooler and Poptop |
Detector had not been used in years.
|
07/20/2022 |
Poptop |
Detector temp increased (~5 degree increase in ~14 days, according to thermistor).
|
11/09/2022 |
Poptop |
Detector resolution changed dramatically from 1.4KeV to 2.1KeV then to 5KeV with what looks to be double humped peaks.
|
Maintenance
Date |
Components |
Reason |
Outcome
|
11/09/2022 |
X-Cooler II |
Ortec sent new o-ring, sieve pill, and high vacuum grease to help try and remedy issues listed above under pumpdown for 11/09/2022. |
|
11/10/2022 |
Poptop |
Detector thermal contacts were cleaned using small amount of acetone and q-tips. Left to air dry before reassembly. Old sieve pill looks discolored but will be saved and stored as they can be refurbished via baking at IAC. |
|
How to Read HPGe Thermistor
Make sure detector high voltage is turned off
Unplug pre-amp power
Unscrew detector electronics housing (8 screws total)
Arrows pointing to screws on detector electronics housing
Slide electronics housing towards the crystal end of the detector
Slide the sectioned area in the direction of the arrow to expose the thermistor terminals
Turn multimeter to resistance (Ω) setting
Measure resistance across the white/black wires that exit the white nylon 4 pinned terminal
Check resistance in tables below for corresponding detector temperature
Resistance and Corresponding Temperatures
100-109Ω
Thermistor Resistance (Ω) |
Detector Temperature (Κ)
|
100 |
76.5780
|
101 |
77.0421
|
102 |
77.5063
|
103 |
77.9706
|
104 |
78.4352
|
105 |
78.9000
|
106 |
79.3649
|
107 |
79.8300
|
108 |
80.2952
|
109 |
80.7607
|
110-119Ω
Thermistor Resistance (Ω) |
Detector Temperature (Κ)
|
110 |
81.2263
|
111 |
81.6921
|
112 |
82.1581
|
113 |
82.6242
|
114 |
83.0906
|
115 |
83.5571
|
116 |
84.0238
|
117 |
84.4906
|
118 |
84.9576
|
119 |
85.4248
|
120-129Ω
Thermistor Resistance (Ω) |
Detector Temperature (Κ)
|
120 |
85.8922
|
121 |
86.3597
|
122 |
86.8275
|
123 |
87.2954
|
124 |
87.7634
|
125 |
88.2317
|
126 |
88.7001
|
127 |
89.1686
|
128 |
89.6374
|
129 |
90.1063
|
130-139Ω
Thermistor Resistance (Ω) |
Detector Temperature (Κ)
|
130 |
90.5754
|
131 |
91.0446
|
132 |
91.5141
|
133 |
91.9837
|
134 |
92.4524
|
135 |
92.9234
|
136 |
93.3935
|
137 |
93.8637
|
138 |
94.3342
|
139 |
94.8048
|
140-149Ω
Thermistor Resistance (Ω) |
Detector Temperature (Κ)
|
140 |
95.2755
|
141 |
95.7465
|
142 |
96.2176
|
143 |
96.6888
|
144 |
97.1603
|
145 |
97.6319
|
146 |
98.1036
|
147 |
98.5756
|
148 |
99.0477
|
149 |
99.5199
|
150-159Ω
Thermistor Resistance (Ω) |
Detector Temperature (Κ)
|
150 |
99.9923
|
151 |
100.4649
|
152 |
100.9377
|
153 |
101.4106
|
154 |
101.8836
|
155 |
102.3569
|
156 |
102.8303
|
157 |
103.3038
|
158 |
103.7775
|
159 |
104.2514
|
160-169Ω
Thermistor Resistance (Ω) |
Detector Temperature (Κ)
|
160 |
104.7254
|
161 |
105.1996
|
162 |
105.6740
|
163 |
106.1485
|
164 |
106.6232
|
165 |
107.0980
|
166 |
107.5730
|
167 |
108.0482
|
168 |
108.5235
|
169 |
108.9989
|
170-179Ω
Thermistor Resistance (Ω) |
Detector Temperature (Κ)
|
170 |
109.4746
|
171 |
109.9503
|
172 |
110.4263
|
173 |
110.9024
|
174 |
111.3786
|
175 |
111.8550
|
176 |
112.3316
|
177 |
112.8083
|
178 |
113.2851
|
179 |
113.7622
|
180-189Ω
Thermistor Resistance (Ω) |
Detector Temperature (Κ)
|
180 |
114.2393
|
181 |
114.7167
|
182 |
115.1941
|
183 |
115.6718
|
184 |
116.1496
|
185 |
116.6275
|
186 |
117.1056
|
187 |
117.5838
|
188 |
118.0622
|
189 |
118.5408
|
190-199Ω
Thermistor Resistance (Ω) |
Detector Temperature (Κ)
|
190 |
119.0194
|
191 |
119.4983
|
192 |
119.9773
|
193 |
120.4564
|
194 |
120.9357
|
195 |
121.4152
|
196 |
121.8947
|
197 |
122.3745
|
198 |
122.8544
|
199 |
123.3344
|
Thermistor Reading Log
Date |
Time |
Thermistor Measurement (Ω) |
Detector Thermal Cycled Between Previous Measurement |
Notes
|
04/13/22 |
11:45am |
241.2 |
Yes |
Detector casing cold to touch, has been cooling down >72hrs
|
04/13/22 |
1:50pm |
229.3 |
No |
|
04/14/22 |
11:26am |
176.0 |
No |
|
06/02/22 |
11:46am |
185.0 |
No |
Cooler was off for a few minutes prior due to detector being moved to other side of room
|
06/16/22 |
10:00am |
176.8 |
No |
Measured before use of high voltage
|
06/30/22 |
11:00am |
184.1 |
No |
Shutdown cable measurement of -11.97, change from normal of -11.87V
|
07/03/22 |
12:45pm |
182.2 |
No |
|
07/06/22 |
11:40am |
181.7 |
No |
Detector warmed after this measurement in prep for pumpdown
|
07/21/22 |
10:50am |
177.2 |
Yes |
Poptop pumped down by IAC, chiller turned on 3:50pm 07/20/22
|
07/21/22 |
11:00am |
176.2 |
No |
Repeat of above
|
07/21/22 |
11:30am |
174.7 |
No |
Repeat of above
|
07/21/22 |
12:30pm |
173.4 |
No |
Repeat of above
|
07/21/22 |
1:22pm |
172.0 |
No |
Repeat of above
|
07/21/22 |
2:11pm |
171.2 |
No |
Repeat of above
|
07/21/22 |
3:30pm |
170.3 |
No |
Repeat of above
|
07/25/22 |
10:10am |
169.5 |
No |
Detector continued to cool over a weekend, shutdown cable = -11.99V
|
10/27/22 |
10:55am |
181.2 |
No |
Resolution of detector changed from ~1.4KeV FWHM to ~2.1KeV FWHM
|
11/01/22 |
10:48am |
171.8 |
Yes |
Thermal cycle of detector in attempt to bring temp back down, attempt was successful.
|
11/03/22 |
12:16pm |
175.1 |
No |
Cs137 peak was wide, poor energy resolution. HV/Preamp turned off, detector beginning to cool again.
|
11/03/22 |
1:00pm |
170.8 |
No |
Preamp & HV off for 45 mins, detector cooled again.
|
11/06/22 |
12:58pm |
184.2 |
Yes |
Thermal cycled detector due to >5KeV sigma in energy peaks.
|
11/07/22 |
7:45am |
192.6 |
No |
Detector cooling for >36hrs
|
PAS HPGe Detector