CH HPGe Efficiency

From New IAC Wiki
Jump to navigation Jump to search

Efficiency Measurements Summer 2022

Electronics Chain

Click here to view the electronics chain used for these efficiency measurements.

Past Work

Most likely incorrect

Eu152 Calibration

  • HPGe calibrated using Eu152 source
CH Eu152 Calibration Spectrum.png

As of 9/28/21

The calibration used to map from ADC channel number to KeV is:

[math]Photon\ Energy\ =\ -36.2928\ +\ 0.449519\ x\ -\ 0.000000861277\ x^2[/math]

Where x = channel 9 of the ADC and is input as PADC.PADC785N[9].

Total Efficiency

This is total efficiency using the Gadolinium foil as the source.

[math] (line\ intensity)\ (incident\ thermal\ neutrons\ per\ second)\ (\sigma_{total})\ (MCNP\ fudge\ factor\ from\ Paul\ approx\ 4)\ =\ Theoretical\ Rate [/math]

For the ease of calculation, I will use the outgoing photon rate from the Gd foil as tabulated by MCNP, thanks to Paul Stonaha. The number given was into 2pi, towards the HPGe detector, multiplying by 2 should give the rate into 4pi.

The 897 KeV peak found in the Gd in/out measurements can be attributed to two different lines created by neutron capture on Gd157. The high purity germanium detector used in this experiment would not be able to resolve the two distinct lines as the resolution is not fine enough. However, adding the intensities together should yield the theoretical rate for the peak measured at that energy.

The first peak is 897.506 KeV with a relative intensity of 17.846%, according the capgam nndc data repository.

The second peak is 897.62 KeV with a relative intensity of 16%, according the capgam nndc data repository.

(0.17846 + 0.16) (2*1374Hz) (234000) = Rate

234000 is the total thermal neutron capture cross section for 157Gd

Note: the fudge factor comes from the nucleus being excited and then de-exciting from the neutron absorption and emits on average, 4 photons per neutron capture

Then find detected rate using HPGe

Divide (detected rate) by (theoretical rate) for total detector efficiency

An easier way to calculate the detector efficiency is

[math]Det.\ Eff.\ =\ \frac{Detected\ Rate}{(Normalized\ Line\ Intensity)(Theoretical\ Total\ Rate)}[/math]

Theoretical total rate = 2748 Hz from simulations

Line Energy Line Weighting Detected Rate Detector Efficiency
897 KeV (2 lines) 0.03967382135 0.357 Hz 0.33% [math]\pm[/math] 0.03
1107 KeV 0.02885336857 0.264 Hz 0.33% [math]\pm[/math] 0.03
~1186 KeV (2 lines) 0.05145546699 0.511 Hz 0.36% [math]\pm[/math] 0.03

Previous Page

CaGaS Phase II

PAS_Experiment

PAS HPGe Detector