Pump downs
Date of pump down |
Components |
Reason
|
05/2021 |
Cold head and poptop |
Detector had not been used in years
|
07/20/2021 |
Poptop |
Detector had begun to warm up
|
How to Read HPGe Thermistor
Make sure detector high voltage is turned off
Unplug pre-amp power
Unscrew detector electronics housing (8 screws total)
Arrows pointing to screws on detector electronics housing
Slide electronics housing towards the crystal end of the detector
Slide the sectioned area in the direction of the arrow to expose the thermistor terminals
Turn multimeter to resistance (Ω) setting
Measure resistance across the white/black wires the exit the white nylon 4 pinned terminal
Check resistance in tables below for corresponding detector temperature
Resistance and Corresponding Temperatures
100-109Ω
Thermistor Resistance (Ω) |
Detector Temperature (Κ)
|
100 |
76.5780
|
101 |
77.0421
|
102 |
77.5063
|
103 |
77.9706
|
104 |
78.4352
|
105 |
78.9000
|
106 |
79.3649
|
107 |
79.8300
|
108 |
80.2952
|
109 |
80.7607
|
110-119Ω
Thermistor Resistance (Ω) |
Detector Temperature (Κ)
|
110 |
81.2263
|
111 |
81.6921
|
112 |
82.1581
|
113 |
82.6242
|
114 |
83.0906
|
115 |
83.5571
|
116 |
84.0238
|
117 |
84.4906
|
118 |
84.9576
|
119 |
85.4248
|
120-129Ω
Thermistor Resistance (Ω) |
Detector Temperature (Κ)
|
120 |
85.8922
|
121 |
86.3597
|
122 |
86.8275
|
123 |
87.2954
|
124 |
87.7634
|
125 |
88.2317
|
126 |
88.7001
|
127 |
89.1686
|
128 |
89.6374
|
129 |
90.1063
|
130-139Ω
Thermistor Resistance (Ω) |
Detector Temperature (Κ)
|
130 |
90.5754
|
131 |
91.0446
|
132 |
91.5141
|
133 |
91.9837
|
134 |
92.4524
|
135 |
92.9234
|
136 |
93.3935
|
137 |
93.8637
|
138 |
94.3342
|
139 |
94.8048
|
140-149Ω
Thermistor Resistance (Ω) |
Detector Temperature (Κ)
|
140 |
95.2755
|
141 |
95.7465
|
142 |
96.2176
|
143 |
96.6888
|
144 |
97.1603
|
145 |
97.6319
|
146 |
98.1036
|
147 |
98.5756
|
148 |
99.0477
|
149 |
99.5199
|
150-159Ω
Thermistor Resistance (Ω) |
Detector Temperature (Κ)
|
150 |
99.9923
|
151 |
100.4649
|
152 |
100.9377
|
153 |
101.4106
|
154 |
101.8836
|
155 |
102.3569
|
156 |
102.8303
|
157 |
103.3038
|
158 |
103.7775
|
159 |
104.2514
|
160-169Ω
Thermistor Resistance (Ω) |
Detector Temperature (Κ)
|
160 |
104.7254
|
161 |
105.1996
|
162 |
105.6740
|
163 |
106.1485
|
164 |
106.6232
|
165 |
107.0980
|
166 |
107.5730
|
167 |
108.0482
|
168 |
108.5235
|
169 |
108.9989
|
170-179Ω
Thermistor Resistance (Ω) |
Detector Temperature (Κ)
|
170 |
109.4746
|
171 |
109.9503
|
172 |
110.4263
|
173 |
110.9024
|
174 |
111.3786
|
175 |
111.8550
|
176 |
112.3316
|
177 |
112.8083
|
178 |
113.2851
|
179 |
113.7622
|
180-189Ω
Thermistor Resistance (Ω) |
Detector Temperature (Κ)
|
180 |
114.2393
|
181 |
114.7167
|
182 |
115.1941
|
183 |
115.6718
|
184 |
116.1496
|
185 |
116.6275
|
186 |
117.1056
|
187 |
117.5838
|
188 |
118.0622
|
189 |
118.5408
|
190-199Ω
Thermistor Resistance (Ω) |
Detector Temperature (Κ)
|
190 |
119.0194
|
191 |
119.4983
|
192 |
119.9773
|
193 |
120.4564
|
194 |
120.9357
|
195 |
121.4152
|
196 |
121.8947
|
197 |
122.3745
|
198 |
122.8544
|
199 |
123.3344
|
Thermistor Reading Log
Date & Time |
Thermistor Measurement (Ω) |
Detector Thermal Cycled Between Previous Measurement |
Notes
|
4/13/22 11:45am |
241.2 |
Yes |
Detector casing cold to touch, has been cooling down >72hrs
|
4/13/22 1:50pm |
229.3 |
No |
|
4/14/22 11:26am |
176.0 |
No |
|
6/2/22 11:46am |
185.0 |
No |
Cooler was off for a few minutes prior due to detector being moved to other side of room
|
6/16/22 10:00am |
176.8 |
No |
Measured before use of high voltage
|
6/30/22 11:00am |
184.1 |
No |
Preliminary shutdown cable measurement of -11.97, .1V away from normal of -11.87V
|
7/3/22 12:45pm |
182.2 |
No |
|
7/6/22 11:40am |
181.7 |
No |
|
PAS HPGe Detector