Attenuation Experiment using New Modules in Electronics Chain and Inside Borated Poly Box
Electronics Chain (Poly Box)
Include schematic of electronics chain
Components
Each component of electronics chain with make/model, function, and brief description of why it's present
Ortec 142PC Pre-amp
Pre-amp for the He3 tube
CAEN Quad Linear Fan In - Fan Out Model N25 Tag #15687
- The fan in/out is used to duplicate the pulse to run the spec amp and watch the signal on the scope for any anomalies.
- The fan in/out is also used to invert the signal from the spec amp to send a negative pulse to the discriminator to trigger the scalar.
Ortec Spec Amp Model 672
- Manual Pole Zero adjustment
LeCroy Discriminator Model 821 Serial #B16303
- Pulse width as measured on scope: ~43ns
Background Measurements
1250V Background Measurements |
|
Run Number |
Configuration |
Discriminator Threshold (mV) |
Cf-252 Source |
Counts |
Run Time (Sec) |
Count Rate (Hz) |
Uncertainty
|
1 |
Detector on 3D printed stands at +1250V |
255 |
No |
|
100 |
|
|
2 |
Detector on 3D printed stands at +1250V |
255 |
No |
|
100 |
|
|
3 |
Detector on 3D printed stands at +1250V |
255 |
No |
|
100 |
|
|
4 |
Detector on 3D printed stands at +1250V |
255 |
No |
|
100 |
|
|
5 |
Detector on 3D printed stands at +1250V |
255 |
No |
|
100 |
|
|
High Activity Source Poly Box Attenuation Measurements
472[math]\mu[/math]Ci at time of measurement.
Open Source
1250V source in |
|
Run Number |
Configuration |
Discriminator Threshold (mV) |
Cf-252 Source |
Counts |
Run Time (Sec)
|
1 |
Detector on poly at +1250V |
80 |
Yes |
34456 |
10
|
2 |
Detector on poly at +1250V |
80 |
Yes |
34920 |
10
|
3 |
Detector on poly at +1250V |
80 |
Yes |
34422 |
10
|
4 |
Detector on poly at +1250V |
80 |
Yes |
34007 |
10
|
5 |
Detector on poly at +1250V |
80 |
Yes |
34420 |
10
|
2" Green poly (one brick) in Box
1250V source in |
|
Run Number |
Configuration |
Discriminator Threshold (mV) |
Cf-252 Source |
Counts |
Run Time (Sec)
|
1 |
Detector on poly at +1250V |
80 |
Yes |
7538 |
10
|
2 |
Detector on poly at +1250V |
80 |
Yes |
7515 |
10
|
3 |
Detector on poly at +1250V |
80 |
Yes |
7613 |
10
|
4 |
Detector on poly at +1250V |
80 |
Yes |
7566 |
10
|
5 |
Detector on poly at +1250V |
80 |
Yes |
7510 |
10
|
4" Green poly (two bricks) in Box
1250V source in |
|
Run Number |
Configuration |
Discriminator Threshold (mV) |
Cf-252 Source |
Counts |
Run Time (Sec)
|
1 |
Detector on poly at +1250V |
80 |
Yes |
4758 |
10
|
2 |
Detector on poly at +1250V |
80 |
Yes |
4798 |
10
|
3 |
Detector on poly at +1250V |
80 |
Yes |
4722 |
10
|
4 |
Detector on poly at +1250V |
80 |
Yes |
4859 |
10
|
5 |
Detector on poly at +1250V |
80 |
Yes |
4672 |
10
|
6" Green poly (three bricks) in Box
1250V source in |
|
Run Number |
Configuration |
Discriminator Threshold (mV) |
Cf-252 Source |
Counts |
Run Time (Sec)
|
1 |
Detector on poly at +1250V |
80 |
Yes |
3664 |
10
|
2 |
Detector on poly at +1250V |
80 |
Yes |
3627 |
10
|
3 |
Detector on poly at +1250V |
80 |
Yes |
3699 |
10
|
4 |
Detector on poly at +1250V |
80 |
Yes |
3688 |
10
|
5 |
Detector on poly at +1250V |
80 |
Yes |
3500 |
10
|
8" Green poly (four bricks) in Box
1250V source in |
|
Run Number |
Configuration |
Discriminator Threshold (mV) |
Cf-252 Source |
Counts |
Run Time (Sec)
|
1 |
Detector on poly at +1250V |
80 |
Yes |
3140 |
10
|
2 |
Detector on poly at +1250V |
80 |
Yes |
3064 |
10
|
3 |
Detector on poly at +1250V |
80 |
Yes |
3030 |
10
|
4 |
Detector on poly at +1250V |
80 |
Yes |
3095 |
10
|
5 |
Detector on poly at +1250V |
80 |
Yes |
3113 |
10
|
Undetermined Activity Source Poly Box Attenuation Measurements
Open Source
1250V source in |
|
Run Number |
Configuration |
Discriminator Threshold (mV) |
Cf-252 Source |
Counts |
Run Time (Sec)
|
1 |
Detector on poly at +1250V |
80 |
Yes |
3873 |
100
|
2 |
Detector on poly at +1250V |
80 |
Yes |
3909 |
100
|
3 |
Detector on poly at +1250V |
80 |
Yes |
3859 |
100
|
4 |
Detector on poly at +1250V |
80 |
Yes |
3944 |
100
|
5 |
Detector on poly at +1250V |
80 |
Yes |
3812 |
100
|
2" Green poly (one brick) in Box
1250V source in |
|
Run Number |
Configuration |
Discriminator Threshold (mV) |
Cf-252 Source |
Counts |
Run Time (Sec)
|
1 |
Detector on poly at +1250V |
80 |
Yes |
764 |
100
|
2 |
Detector on poly at +1250V |
80 |
Yes |
819 |
100
|
3 |
Detector on poly at +1250V |
80 |
Yes |
818 |
100
|
4 |
Detector on poly at +1250V |
80 |
Yes |
815 |
100
|
5 |
Detector on poly at +1250V |
80 |
Yes |
913 |
100
|
4" Green poly (two bricks) in Box
1250V source in |
|
Run Number |
Configuration |
Discriminator Threshold (mV) |
Cf-252 Source |
Counts |
Run Time (Sec)
|
1 |
Detector on poly at +1250V |
80 |
Yes |
486 |
100
|
2 |
Detector on poly at +1250V |
80 |
Yes |
509 |
100
|
3 |
Detector on poly at +1250V |
80 |
Yes |
531 |
100
|
4 |
Detector on poly at +1250V |
80 |
Yes |
544 |
100
|
5 |
Detector on poly at +1250V |
80 |
Yes |
530 |
100
|
6" Green poly (three bricks) in Box
1250V source in |
|
Run Number |
Configuration |
Discriminator Threshold (mV) |
Cf-252 Source |
Counts |
Run Time (Sec)
|
1 |
Detector on poly at +1250V |
80 |
Yes |
415 |
100
|
2 |
Detector on poly at +1250V |
80 |
Yes |
373 |
100
|
3 |
Detector on poly at +1250V |
80 |
Yes |
380 |
100
|
4 |
Detector on poly at +1250V |
80 |
Yes |
397 |
100
|
5 |
Detector on poly at +1250V |
80 |
Yes |
419 |
100
|
8" Green poly (four bricks) in Box
1250V source in |
|
Run Number |
Configuration |
Discriminator Threshold (mV) |
Cf-252 Source |
Counts |
Run Time (Sec)
|
1 |
Detector on poly at +1250V |
80 |
Yes |
371 |
100
|
2 |
Detector on poly at +1250V |
80 |
Yes |
396 |
100
|
3 |
Detector on poly at +1250V |
80 |
Yes |
351 |
100
|
4 |
Detector on poly at +1250V |
80 |
Yes |
366 |
100
|
5 |
Detector on poly at +1250V |
80 |
Yes |
394 |
100
|
Lead Attenuation Measurements
The purpose of these measurements was to determine if photon build up in the He3 tube was a significant effect to consider when trying to determine the activity of the Unknown Source
4/5/22
Lead Open Source
1250V source in |
|
Run Number |
Configuration |
Discriminator Threshold (mV) |
Cf-252 Source |
Counts |
Run Time (Sec)
|
1 |
Detector on poly at +1250V |
80 |
Yes |
3873 |
100
|
2 |
Detector on poly at +1250V |
80 |
Yes |
3909 |
100
|
3 |
Detector on poly at +1250V |
80 |
Yes |
3859 |
100
|
4 |
Detector on poly at +1250V |
80 |
Yes |
3944 |
100
|
5 |
Detector on poly at +1250V |
80 |
Yes |
3812 |
100
|
1" Lead
1250V source in |
|
Run Number |
Configuration |
Discriminator Threshold (mV) |
Cf-252 Source |
Counts |
Run Time (Sec)
|
1 |
Detector on poly at +1250V |
80 |
Yes |
4152 |
100
|
2 |
Detector on poly at +1250V |
80 |
Yes |
3945 |
100
|
3 |
Detector on poly at +1250V |
80 |
Yes |
4141 |
100
|
4 |
Detector on poly at +1250V |
80 |
Yes |
3974 |
100
|
5 |
Detector on poly at +1250V |
80 |
Yes |
4154 |
100
|
2" Lead
1250V source in |
|
Run Number |
Configuration |
Discriminator Threshold (mV) |
Cf-252 Source |
Counts |
Run Time (Sec)
|
1 |
Detector on poly at +1250V |
80 |
Yes |
3509 |
100
|
2 |
Detector on poly at +1250V |
80 |
Yes |
3472 |
100
|
3 |
Detector on poly at +1250V |
80 |
Yes |
3387 |
100
|
4 |
Detector on poly at +1250V |
80 |
Yes |
3488 |
100
|
5 |
Detector on poly at +1250V |
80 |
Yes |
3463 |
100
|
3" Lead
1250V source in |
|
Run Number |
Configuration |
Discriminator Threshold (mV) |
Cf-252 Source |
Counts |
Run Time (Sec)
|
1 |
Detector on poly at +1250V |
80 |
Yes |
2995 |
100
|
2 |
Detector on poly at +1250V |
80 |
Yes |
2921 |
100
|
3 |
Detector on poly at +1250V |
80 |
Yes |
2956 |
100
|
4 |
Detector on poly at +1250V |
80 |
Yes |
2916 |
100
|
5 |
Detector on poly at +1250V |
80 |
Yes |
3015 |
100
|
4" Lead
1250V source in |
|
Run Number |
Configuration |
Discriminator Threshold (mV) |
Cf-252 Source |
Counts |
Run Time (Sec)
|
1 |
Detector on poly at +1250V |
80 |
Yes |
2630 |
100
|
2 |
Detector on poly at +1250V |
80 |
Yes |
2601 |
100
|
3 |
Detector on poly at +1250V |
80 |
Yes |
2558 |
100
|
4 |
Detector on poly at +1250V |
80 |
Yes |
2629 |
100
|
5 |
Detector on poly at +1250V |
80 |
Yes |
2650 |
100
|
Lowest Activity Source Poly Box Attenuation Measurements
Approximately 1.15[math]\mu[/math]Ci at time of measurement, according to ISU Radiation Safety
Open Source
1250V source in |
|
Run Number |
Configuration |
Discriminator Threshold (mV) |
Cf-252 Source |
Counts |
Run Time (Sec)
|
1 |
Detector on poly at +1250V |
80 |
Yes |
31 |
100
|
2 |
Detector on poly at +1250V |
80 |
Yes |
32 |
100
|
3 |
Detector on poly at +1250V |
80 |
Yes |
26 |
100
|
4 |
Detector on poly at +1250V |
80 |
Yes |
23 |
100
|
5 |
Detector on poly at +1250V |
80 |
Yes |
29 |
100
|
2" Green poly (one brick) in Box
1250V source in |
|
Run Number |
Configuration |
Discriminator Threshold (mV) |
Cf-252 Source |
Counts |
Run Time (Sec)
|
1 |
Detector on poly at +1250V |
80 |
Yes |
18 |
100
|
2 |
Detector on poly at +1250V |
80 |
Yes |
27 |
100
|
3 |
Detector on poly at +1250V |
80 |
Yes |
21 |
100
|
4 |
Detector on poly at +1250V |
80 |
Yes |
21 |
100
|
5 |
Detector on poly at +1250V |
80 |
Yes |
22 |
100
|
4" Green poly (two bricks) in Box
1250V source in |
|
Run Number |
Configuration |
Discriminator Threshold (mV) |
Cf-252 Source |
Counts |
Run Time (Sec)
|
1 |
Detector on poly at +1250V |
80 |
Yes |
31 |
100
|
2 |
Detector on poly at +1250V |
80 |
Yes |
18 |
100
|
3 |
Detector on poly at +1250V |
80 |
Yes |
19 |
100
|
4 |
Detector on poly at +1250V |
80 |
Yes |
31 |
100
|
5 |
Detector on poly at +1250V |
80 |
Yes |
30 |
100
|
6" Green poly (three bricks) in Box
1250V source in |
|
Run Number |
Configuration |
Discriminator Threshold (mV) |
Cf-252 Source |
Counts |
Run Time (Sec)
|
1 |
Detector on poly at +1250V |
80 |
Yes |
35 |
100
|
2 |
Detector on poly at +1250V |
80 |
Yes |
24 |
100
|
3 |
Detector on poly at +1250V |
80 |
Yes |
23 |
100
|
4 |
Detector on poly at +1250V |
80 |
Yes |
21 |
100
|
5 |
Detector on poly at +1250V |
80 |
Yes |
22 |
100
|
8" Green poly (four bricks) in Box
1250V source in |
|
Run Number |
Configuration |
Discriminator Threshold (mV) |
Cf-252 Source |
Counts |
Run Time (Sec)
|
1 |
Detector on poly at +1250V |
80 |
Yes |
20 |
100
|
2 |
Detector on poly at +1250V |
80 |
Yes |
28 |
100
|
3 |
Detector on poly at +1250V |
80 |
Yes |
29 |
100
|
4 |
Detector on poly at +1250V |
80 |
Yes |
16 |
100
|
5 |
Detector on poly at +1250V |
80 |
Yes |
22 |
100
|
Previous Page
CaGaS Phase II