Attenuation Experiment using New Modules in Electronics Chain and Inside Borated Poly Box
Electronics Chain (Poly Box)
Include schematic of electronics chain
Components
Each component of electronics chain with make/model, function, and brief description of why it's present
Ortec 142PC Pre-amp
Pre-amp for the He3 tube
CAEN Quad Linear Fan In - Fan Out Model N25 Tag #15687
- The fan in/out is used to duplicate the pulse to run the spec amp and watch the signal on the scope for any anomalies.
- The fan in/out is also used to invert the signal from the spec amp to send a negative pulse to the discriminator to trigger the scalar.
Ortec Spec Amp Model 672
- Manual Pole Zero adjustment
LeCroy Discriminator Model 821 Serial #B16303
- Pulse width as measured on scope: ~43ns
Scope Images
Plateau Measurements
| 1250V Plateau Measurements |
|
| Run Number |
Configuration |
Discriminator Threshold (mV) |
Cf-252 Source |
Counts |
Run Time (Sec) |
Count Rate (Hz) |
Uncertainty
|
| 1 |
Detector on poly blocks at +1250V |
30 |
Yes |
|
100 |
|
|
| 2 |
Detector on poly blocks at +1250V |
40 |
Yes |
|
100 |
|
|
| 3 |
Detector on poly blocks at +1250V |
50 |
Yes |
|
100 |
|
|
| 4 |
Detector on poly blocks at +1250V |
60 |
Yes |
|
100 |
|
|
| 5 |
Detector on poly blocks at +1250V |
70 |
Yes |
|
100 |
|
|
| 6 |
Detector on poly blocks at +1250V |
80 |
Yes |
|
100 |
|
|
| 7 |
Detector on poly blocks at +1250V |
90 |
Yes |
|
100 |
|
|
| 8 |
Detector on poly blocks at +1250V |
100 |
Yes |
|
100 |
|
|
Background Measurements
| 1250V Background Measurements |
|
| Run Number |
Configuration |
Discriminator Threshold (mV) |
Cf-252 Source |
Counts |
Run Time (Sec) |
Count Rate (Hz) |
Uncertainty
|
| 1 |
Detector on 3D printed stands at +1250V |
255 |
No |
|
100 |
|
|
| 2 |
Detector on 3D printed stands at +1250V |
255 |
No |
|
100 |
|
|
| 3 |
Detector on 3D printed stands at +1250V |
255 |
No |
|
100 |
|
|
| 4 |
Detector on 3D printed stands at +1250V |
255 |
No |
|
100 |
|
|
| 5 |
Detector on 3D printed stands at +1250V |
255 |
No |
|
100 |
|
|
Poly Box Attenuation Measurements
Open Source
| 1250V source in |
|
| Run Number |
Configuration |
Discriminator Threshold (mV) |
Cf-252 Source |
Counts |
Run Time (Sec) |
Count Rate (Hz) |
Uncertainty
|
| 1 |
Detector on poly at +1250V |
80 |
Yes |
|
100 |
|
|
| 2 |
Detector on poly at +1250V |
80 |
Yes |
|
100 |
|
|
| 3 |
Detector on poly at +1250V |
80 |
Yes |
|
100 |
|
|
| 4 |
Detector on poly at +1250V |
80 |
Yes |
|
100 |
|
|
| 5 |
Detector on poly at +1250V |
80 |
Yes |
|
100 |
|
|
2" Green poly (one brick) in Box
| 1250V source in |
|
| Run Number |
Configuration |
Discriminator Threshold (mV) |
Cf-252 Source |
Counts |
Run Time (Sec) |
Count Rate (Hz) |
Uncertainty
|
| 1 |
Detector on poly at +1250V |
80 |
Yes |
764 |
100 |
|
|
| 2 |
Detector on poly at +1250V |
80 |
Yes |
819 |
100 |
|
|
| 3 |
Detector on poly at +1250V |
80 |
Yes |
818 |
100 |
|
|
| 4 |
Detector on poly at +1250V |
80 |
Yes |
815 |
100 |
|
|
| 5 |
Detector on poly at +1250V |
80 |
Yes |
913 |
100 |
|
|
4" Green poly (two bricks) in Box
| 1250V source in |
|
| Run Number |
Configuration |
Discriminator Threshold (mV) |
Cf-252 Source |
Counts |
Run Time (Sec) |
Count Rate (Hz) |
Uncertainty
|
| 1 |
Detector on poly at +1250V |
80 |
Yes |
486 |
100 |
|
|
| 2 |
Detector on poly at +1250V |
80 |
Yes |
509 |
100 |
|
|
| 3 |
Detector on poly at +1250V |
80 |
Yes |
531 |
100 |
|
|
| 4 |
Detector on poly at +1250V |
80 |
Yes |
544 |
100 |
|
|
| 5 |
Detector on poly at +1250V |
80 |
Yes |
530 |
100 |
|
|
6" Green poly (three bricks) in Box
| 1250V source in |
|
| Run Number |
Configuration |
Discriminator Threshold (mV) |
Cf-252 Source |
Counts |
Run Time (Sec) |
Count Rate (Hz) |
Uncertainty
|
| 1 |
Detector on poly at +1250V |
80 |
Yes |
415 |
100 |
|
|
| 2 |
Detector on poly at +1250V |
80 |
Yes |
373 |
100 |
|
|
| 3 |
Detector on poly at +1250V |
80 |
Yes |
380 |
100 |
|
|
| 4 |
Detector on poly at +1250V |
80 |
Yes |
397 |
100 |
|
|
| 5 |
Detector on poly at +1250V |
80 |
Yes |
419 |
100 |
|
|
8" Green poly (four bricks) in Box
| 1250V source in |
|
| Run Number |
Configuration |
Discriminator Threshold (mV) |
Cf-252 Source |
Counts |
Run Time (Sec) |
Count Rate (Hz) |
Uncertainty
|
| 1 |
Detector on poly at +1250V |
80 |
Yes |
371 |
100 |
|
|
| 2 |
Detector on poly at +1250V |
80 |
Yes |
396 |
100 |
|
|
| 3 |
Detector on poly at +1250V |
80 |
Yes |
351 |
100 |
|
|
| 4 |
Detector on poly at +1250V |
80 |
Yes |
366 |
100 |
|
|
| 5 |
Detector on poly at +1250V |
80 |
Yes |
394 |
100 |
|
|
Previous Page
CaGaS Phase II