NaI JD-766 Plateau
Jump to navigation
Jump to search
Timing filter Amp Configuration
Course gain: x6
Fine gain: at minimum
Integrate (nS): 500nS
Diff (nS): Out
Invert/Noninverted: Non-Inv
Dual Timer Configuration
Setup such that input discriminated pulse is at width ~200ns
Out feeds into scalar for counts
Plateau Data
700V
700V Source Out Data | ||||||
Run Number | Configuration | Discriminator Threshold (mV) | Cs-137 Button Source | Counts | Run Time (Sec) | Count Rate (Hz) |
1 | Unshielded detector at +700V | 05 | No | 958 | 60 | 15.97 |
2 | Unshielded detector at +700V | 15 | No | 278 | 60 | 4.63 |
3 | Unshielded detector at +700V | 35 | No | 107 | 60 | 1.78 |
4 | Unshielded detector at +700V | 55 | No | 85 | 60 | 1.42 |
5 | Unshielded detector at +700V | 75 | No | 80 | 60 | 1.33 |
6 | Unshielded detector at +700V | 95 | No | 80 | 60 | 1.33 |
7 | Unshielded detector at +700V | 115 | No | 75 | 60 | 1.25 |
8 | Unshielded detector at +700V | 135 | No | 78 | 60 | 1.3 |
9 | Unshielded detector at +700V | 155 | No | 82 | 60 | 1.37 |
10 | Unshielded detector at +700V | 175 | No | 77 | 60 | 1.28 |
11 | Unshielded detector at +700V | 195 | No | 66 | 60 | 1.1 |
12 | Unshielded detector at +700V | 215 | No | 68 | 60 | 1.13 |
13 | Unshielded detector at +700V | 235 | No | 53 | 60 | 0.88 |
14 | Unshielded detector at +700V | 255 | No | 59 | 60 | 0.98 |
700V Source In Data | ||||||
Run Number | Configuration | Discriminator Threshold (mV) | Cs-137 Button Source | Counts | Run Time (Sec) | Count Rate (Hz) |
1 | Unshielded detector at +700V | 05 | Yes | 1033 | 60 | |
2 | Unshielded detector at +700V | 15 | Yes | 291 | 60 | |
3 | Unshielded detector at +700V | 35 | Yes | 102 | 60 | |
4 | Unshielded detector at +700V | 55 | Yes | 87 | 60 | |
5 | Unshielded detector at +700V | 75 | Yes | 88 | 60 | |
6 | Unshielded detector at +700V | 95 | Yes | 103 | 60 | |
7 | Unshielded detector at +700V | 115 | Yes | 101 | 60 | |
8 | Unshielded detector at +700V | 135 | Yes | 84 | 60 | |
9 | Unshielded detector at +700V | 155 | Yes | 72 | 60 | |
10 | Unshielded detector at +700V | 175 | Yes | 57 | 60 | |
11 | Unshielded detector at +700V | 195 | Yes | 72 | 60 | |
12 | Unshielded detector at +700V | 215 | Yes | 68 | 60 | |
13 | Unshielded detector at +700V | 235 | Yes | 74 | 60 | |
14 | Unshielded detector at +700V | 255 | Yes | 52 | 60 |
700V Plots | ||
750V
750V Source Out Data | ||||||
Run Number | Configuration | Discriminator Threshold (mV) | Cs-137 Button Source | Counts | Run Time (Sec) | Count Rate (Hz) |
1 | Unshielded detector at +750V | 05 | No | 2211 | 60 | 36.85 |
2 | Unshielded detector at +750V | 15 | No | 1125 | 60 | 18.75 |
3 | Unshielded detector at +750V | 35 | No | 293 | 60 | 4.88 |
4 | Unshielded detector at +750V | 55 | No | 105 | 60 | 1.75 |
5 | Unshielded detector at +750V | 75 | No | 80 | 60 | 1.33 |
6 | Unshielded detector at +750V | 95 | No | 97 | 60 | 1.62 |
7 | Unshielded detector at +750V | 115 | No | 99 | 60 | 1.65 |
8 | Unshielded detector at +750V | 135 | No | 89 | 60 | 1.48 |
9 | Unshielded detector at +750V | 155 | No | 74 | 60 | 1.23 |
10 | Unshielded detector at +750V | 175 | No | 80 | 60 | 1.33 |
11 | Unshielded detector at +750V | 195 | No | 80 | 60 | 1.33 |
12 | Unshielded detector at +750V | 215 | No | 64 | 60 | 1.07 |
13 | Unshielded detector at +750V | 235 | No | 89 | 60 | 1.48 |
14 | Unshielded detector at +750V | 255 | No | 69 | 60 | 1.15 |
750V Source In Data | ||||||
Run Number | Configuration | Discriminator Threshold (mV) | Cs-137 Button Source | Counts | Run Time (Sec) | Count Rate (Hz) |
1 | Unshielded detector at +750V | 05 | Yes | 4055 | 60 | |
2 | Unshielded detector at +750V | 15 | Yes | 1160 | 60 | |
3 | Unshielded detector at +750V | 35 | Yes | 301 | 60 | |
4 | Unshielded detector at +750V | 55 | Yes | 102 | 60 | |
5 | Unshielded detector at +750V | 75 | Yes | 82 | 60 | |
6 | Unshielded detector at +750V | 95 | Yes | 114 | 60 | |
7 | Unshielded detector at +750V | 115 | Yes | 85 | 60 | |
8 | Unshielded detector at +750V | 135 | Yes | 93 | 60 | |
9 | Unshielded detector at +750V | 155 | Yes | 87 | 60 | |
10 | Unshielded detector at +750V | 175 | Yes | 87 | 60 | |
11 | Unshielded detector at +750V | 195 | Yes | 79 | 60 | |
12 | Unshielded detector at +750V | 215 | Yes | 90 | 60 | |
13 | Unshielded detector at +750V | 235 | Yes | 84 | 60 | |
14 | Unshielded detector at +750V | 255 | Yes | 73 | 60 |
750V Plots | ||
800V
800V Source Out Data | ||||||
Run Number | Configuration | Discriminator Threshold (mV) | Cs-137 Button Source | Counts | Run Time (Sec) | Count Rate (Hz) |
1 | Unshielded detector at +800V | 05 | No | 3866 | 60 | 64.43 |
2 | Unshielded detector at +800V | 15 | No | 2356 | 60 | 39.27 |
3 | Unshielded detector at +800V | 35 | No | 1091 | 60 | 18.18 |
4 | Unshielded detector at +800V | 55 | No | 384 | 60 | 6.4 |
5 | Unshielded detector at +800V | 75 | No | 244 | 60 | 4.07 |
6 | Unshielded detector at +800V | 95 | No | 167 | 60 | 2.78 |
7 | Unshielded detector at +800V | 115 | No | 100 | 60 | 1.67 |
8 | Unshielded detector at +800V | 135 | No | 88 | 60 | 1.47 |
9 | Unshielded detector at +800V | 155 | No | 89 | 60 | 1.48 |
10 | Unshielded detector at +800V | 175 | No | 86 | 60 | 1.43 |
11 | Unshielded detector at +800V | 195 | No | 90 | 60 | 1.5 |
12 | Unshielded detector at +800V | 215 | No | 81 | 60 | 1.35 |
13 | Unshielded detector at +800V | 235 | No | 95 | 60 | 1.58 |
14 | Unshielded detector at +800V | 255 | No | 89 | 60 | 1.48 |
800V Source In Data | ||||||
Run Number | Configuration | Discriminator Threshold (mV) | Cs-137 Button Source | Counts | Run Time (Sec) | Count Rate (Hz) |
1 | Unshielded detector at +800V | 05 | Yes | 10055 | 60 | |
2 | Unshielded detector at +800V | 15 | Yes | 5438 | 60 | |
3 | Unshielded detector at +800V | 35 | Yes | 1078 | 60 | |
4 | Unshielded detector at +800V | 55 | Yes | 375 | 60 | |
5 | Unshielded detector at +800V | 75 | Yes | 217 | 60 | |
6 | Unshielded detector at +800V | 95 | Yes | 126 | 60 | |
7 | Unshielded detector at +800V | 115 | Yes | 100 | 60 | |
8 | Unshielded detector at +800V | 135 | Yes | 93 | 60 | |
9 | Unshielded detector at +800V | 155 | Yes | 99 | 60 | |
10 | Unshielded detector at +800V | 175 | Yes | 96 | 60 | |
11 | Unshielded detector at +800V | 195 | Yes | 102 | 60 | |
12 | Unshielded detector at +800V | 215 | Yes | 99 | 60 | |
13 | Unshielded detector at +800V | 235 | Yes | 62 | 60 | |
14 | Unshielded detector at +800V | 255 | Yes | 85 | 60 |
800V Plots | ||
850V
850V Source Out Data | ||||||
Run Number | Configuration | Discriminator Threshold (mV) | Cs-137 Button Source | Counts | Run Time (Sec) | Count Rate (Hz) |
1 | Unshielded detector at +850V | 05 | No | 6402 | 60 | 106.7 |
2 | Unshielded detector at +850V | 15 | No | 3867 | 60 | 64.45 |
3 | Unshielded detector at +850V | 35 | No | 2058 | 60 | 34.3 |
4 | Unshielded detector at +850V | 55 | No | 1233 | 60 | 20.55 |
5 | Unshielded detector at +850V | 75 | No | 787 | 60 | 13.12 |
6 | Unshielded detector at +850V | 95 | No | 391 | 60 | 6.52 |
7 | Unshielded detector at +850V | 115 | No | 256 | 60 | 4.27 |
8 | Unshielded detector at +850V | 135 | No | 193 | 60 | 3.22 |
9 | Unshielded detector at +850V | 155 | No | 151 | 60 | 2.52 |
10 | Unshielded detector at +850V | 175 | No | 97 | 60 | 1.62 |
11 | Unshielded detector at +850V | 195 | No | 105 | 60 | 1.75 |
12 | Unshielded detector at +850V | 215 | No | 97 | 60 | 1.62 |
13 | Unshielded detector at +850V | 235 | No | 87 | 60 | 1.45 |
14 | Unshielded detector at +850V | 255 | No | 101 | 60 | 1.68 |
850V Source In Data | ||||||
Run Number | Configuration | Discriminator Threshold (mV) | Cs-137 Button Source | Counts | Run Time (Sec) | Count Rate (Hz) |
1 | Unshielded detector at +850V | 05 | Yes | 15370 | 60 | |
2 | Unshielded detector at +850V | 15 | Yes | 10298 | 60 | |
3 | Unshielded detector at +850V | 35 | Yes | 2963 | 60 | |
4 | Unshielded detector at +850V | 55 | Yes | 1366 | 60 | |
5 | Unshielded detector at +850V | 75 | Yes | 842 | 60 | |
6 | Unshielded detector at +850V | 95 | Yes | 402 | 60 | |
7 | Unshielded detector at +850V | 115 | Yes | 281 | 60 | |
8 | Unshielded detector at +850V | 135 | Yes | 196 | 60 | |
9 | Unshielded detector at +850V | 155 | Yes | 143 | 60 | |
10 | Unshielded detector at +850V | 175 | Yes | 98 | 60 | |
11 | Unshielded detector at +850V | 195 | Yes | 88 | 60 | |
12 | Unshielded detector at +850V | 215 | Yes | 101 | 60 | |
13 | Unshielded detector at +850V | 235 | Yes | 104 | 60 | |
14 | Unshielded detector at +850V | 255 | Yes | 101 | 60 |
850V Plots | ||
1000V
1000V Source Out Data | ||||||
Run Number | Configuration | Discriminator Threshold (mV) | Cs-137 Button Source | Counts | Run Time (Sec) | Count Rate (Hz) |
1 | Unshielded detector at +1000V | 05 | No | 21167 | 60 | |
2 | Unshielded detector at +1000V | 10 | No | 18739 | 60 | |
3 | Unshielded detector at +1000V | 15 | No | 16091 | 60 | |
4 | Unshielded detector at +1000V | 20 | No | 13773 | 60 | |
5 | Unshielded detector at +1000V | 25 | No | 11983 | 60 | |
6 | Unshielded detector at +1000V | 30 | No | 10358 | 60 | |
7 | Unshielded detector at +1000V | 35 | No | 9274 | 60 | |
8 | Unshielded detector at +1000V | 45 | No | 7423 | 60 | |
9 | Unshielded detector at +1000V | 55 | No | 6067 | 60 | |
10 | Unshielded detector at +1000V | 65 | No | 5258 | 60 | |
11 | Unshielded detector at +1000V | 75 | No | 4451 | 60 | |
12 | Unshielded detector at +1000V | 80 | No | 4321 | 60 | |
13 | Unshielded detector at +1000V | 85 | No | 4111 | 60 | |
14 | Unshielded detector at +1000V | 90 | No | 3823 | 60 | |
15 | Unshielded detector at +1000V | 95 | No | 3666 | 60 | |
16 | Unshielded detector at +1000V | 100 | No | 3466 | 60 | |
17 | Unshielded detector at +1000V | 105 | No | 3241 | 60 | |
18 | Unshielded detector at +1000V | 110 | No | 3180 | 60 | |
19 | Unshielded detector at +1000V | 115 | No | 2968 | 60 | |
20 | Unshielded detector at +1000V | 120 | No | 2812 | 60 | |
21 | Unshielded detector at +1000V | 125 | No | 2769 | 60 | |
22 | Unshielded detector at +1000V | 130 | No | 2548 | 60 | |
23 | Unshielded detector at +1000V | 135 | No | 2426 | 60 | |
24 | Unshielded detector at +1000V | 155 | No | 2065 | 60 | |
25 | Unshielded detector at +1000V | 175 | No | 1843 | 60 | |
26 | Unshielded detector at +1000V | 195 | No | 1582 | 60 | |
27 | Unshielded detector at +1000V | 215 | No | 1374 | 60 | |
28 | Unshielded detector at +1000V | 235 | No | 1172 | 60 | |
29 | Unshielded detector at +1000V | 255 | No | 1012 | 60 |
1000V Source In Data | ||||||
Run Number | Configuration | Discriminator Threshold (mV) | Cs-137 Button Source | Counts | Run Time (Sec) | Count Rate (Hz) |
1 | Unshielded detector at +1000V | 05 | Yes | 34443 | 60 | |
2 | Unshielded detector at +1000V | 10 | Yes | 30427 | 60 | |
3 | Unshielded detector at +1000V | 15 | Yes | 26882 | 60 | |
4 | Unshielded detector at +1000V | 20 | Yes | 23779 | 60 | |
5 | Unshielded detector at +1000V | 25 | Yes | 21672 | 60 | |
6 | Unshielded detector at +1000V | 30 | Yes | 19611 | 60 | |
7 | Unshielded detector at +1000V | 35 | Yes | 18115 | 60 | |
8 | Unshielded detector at +1000V | 45 | Yes | 15119 | 60 | |
9 | Unshielded detector at +1000V | 55 | Yes | 12824 | 60 | |
10 | Unshielded detector at +1000V | 65 | Yes | 11575 | 60 | |
11 | Unshielded detector at +1000V | 75 | Yes | 10320 | 60 | |
12 | Unshielded detector at +1000V | 80 | Yes | 9658 | 60 | |
13 | Unshielded detector at +1000V | 85 | Yes | 9328 | 60 | |
14 | Unshielded detector at +1000V | 90 | Yes | 8885 | 60 | |
15 | Unshielded detector at +1000V | 95 | Yes | 8328 | 60 | |
16 | Unshielded detector at +1000V | 100 | Yes | 7906 | 60 | |
17 | Unshielded detector at +1000V | 105 | Yes | 7777 | 60 | |
18 | Unshielded detector at +1000V | 110 | Yes | 7478 | 60 | |
19 | Unshielded detector at +1000V | 115 | Yes | 7313 | 60 | |
20 | Unshielded detector at +1000V | 120 | Yes | 7206 | 60 | |
21 | Unshielded detector at +1000V | 125 | Yes | 6947 | 60 | |
22 | Unshielded detector at +1000V | 130 | Yes | 6936 | 60 | |
23 | Unshielded detector at +1000V | 135 | Yes | 6692 | 60 | |
24 | Unshielded detector at +1000V | 155 | Yes | 2242 | 60 | |
25 | Unshielded detector at +1000V | 175 | Yes | 1878 | 60 | |
26 | Unshielded detector at +1000V | 195 | Yes | 1596 | 60 | |
27 | Unshielded detector at +1000V | 215 | Yes | 1378 | 60 | |
28 | Unshielded detector at +1000V | 235 | Yes | 1280 | 60 | |
29 | Unshielded detector at +1000V | 255 | Yes | 1092 | 60 |
1000V Plots | ||
1050V
Operating voltage when detectors used at IAC
1050V Source Out Data | ||||||
Run Number | Configuration | Discriminator Threshold (mV) | Cs-137 Button Source | Counts | Run Time (Sec) | Count Rate (Hz) |
1 | Unshielded detector at +1050V | 05 | No | 22001 | 60 | |
2 | Unshielded detector at +1050V | 10 | No | 20338 | 60 | |
3 | Unshielded detector at +1050V | 15 | No | 18679 | 60 | |
4 | Unshielded detector at +1050V | 20 | No | 16810 | 60 | |
5 | Unshielded detector at +1050V | 25 | No | 15498 | 60 | |
6 | Unshielded detector at +1050V | 30 | No | 14096 | 60 | |
7 | Unshielded detector at +1050V | 35 | No | 13005 | 60 | |
8 | Unshielded detector at +1050V | 45 | No | 10948 | 60 | |
9 | Unshielded detector at +1050V | 55 | No | 9276 | 60 | |
10 | Unshielded detector at +1050V | 65 | No | 8022 | 60 | |
11 | Unshielded detector at +1050V | 75 | No | 7012 | 60 | |
12 | Unshielded detector at +1050V | 80 | No | 6618 | 60 | |
13 | Unshielded detector at +1050V | 85 | No | 6169 | 60 | |
14 | Unshielded detector at +1050V | 90 | No | 5942 | 60 | |
15 | Unshielded detector at +1050V | 95 | No | 5559 | 60 | |
16 | Unshielded detector at +1050V | 100 | No | 5272 | 60 | |
17 | Unshielded detector at +1050V | 105 | No | 4871 | 60 | |
18 | Unshielded detector at +1050V | 110 | No | 4783 | 60 | |
19 | Unshielded detector at +1050V | 115 | No | 4451 | 60 | |
20 | Unshielded detector at +1050V | 120 | No | 4287 | 60 | |
21 | Unshielded detector at +1050V | 125 | No | 4168 | 60 | |
22 | Unshielded detector at +1050V | 130 | No | 4020 | 60 | |
23 | Unshielded detector at +1050V | 135 | No | 3860 | 60 | |
24 | Unshielded detector at +1050V | 155 | No | 3405 | 60 | |
25 | Unshielded detector at +1050V | 175 | No | 3029 | 60 | |
26 | Unshielded detector at +1050V | 195 | No | 2634 | 60 | |
27 | Unshielded detector at +1050V | 215 | No | 2298 | 60 | |
28 | Unshielded detector at +1050V | 235 | No | 2153 | 60 | |
29 | Unshielded detector at +1050V | 255 | No | 2002 | 60 |
1050V Source In Data | ||||||
Run Number | Configuration | Discriminator Threshold (mV) | Cs-137 Button Source | Counts | Run Time (Sec) | Count Rate (Hz) |
1 | Unshielded detector at +1050V | 05 | Yes | 34250 | 60 | |
2 | Unshielded detector at +1050V | 10 | Yes | 31439 | 60 | |
3 | Unshielded detector at +1050V | 15 | Yes | 29093 | 60 | |
4 | Unshielded detector at +1050V | 20 | Yes | 26912 | 60 | |
5 | Unshielded detector at +1050V | 25 | Yes | 25185 | 60 | |
6 | Unshielded detector at +1050V | 30 | Yes | 23568 | 60 | |
7 | Unshielded detector at +1050V | 35 | Yes | 21921 | 60 | |
8 | Unshielded detector at +1050V | 45 | Yes | 19174 | 60 | |
9 | Unshielded detector at +1050V | 55 | Yes | 17314 | 60 | |
10 | Unshielded detector at +1050V | 65 | Yes | 15339 | 60 | |
11 | Unshielded detector at +1050V | 75 | Yes | 13939 | 60 | |
12 | Unshielded detector at +1050V | 80 | Yes | 13280 | 60 | |
13 | Unshielded detector at +1050V | 85 | Yes | 12649 | 60 | |
14 | Unshielded detector at +1050V | 90 | Yes | 12203 | 60 | |
15 | Unshielded detector at +1050V | 95 | Yes | 11962 | 60 | |
16 | Unshielded detector at +1050V | 100 | Yes | 11217 | 60 | |
17 | Unshielded detector at +1050V | 105 | Yes | 10733 | 60 | |
18 | Unshielded detector at +1050V | 110 | Yes | 10347 | 60 | |
19 | Unshielded detector at +1050V | 115 | Yes | 9953 | 60 | |
20 | Unshielded detector at +1050V | 120 | Yes | 9691 | 60 | |
21 | Unshielded detector at +1050V | 125 | Yes | 9551 | 60 | |
22 | Unshielded detector at +1050V | 130 | Yes | 9147 | 60 | |
23 | Unshielded detector at +1050V | 135 | Yes | 8840 | 60 | |
24 | Unshielded detector at +1050V | 155 | Yes | 7735 | 60 | |
25 | Unshielded detector at +1050V | 175 | Yes | 6970 | 60 | |
26 | Unshielded detector at +1050V | 195 | Yes | 6587 | 60 | |
27 | Unshielded detector at +1050V | 215 | Yes | 5559 | 60 | |
28 | Unshielded detector at +1050V | 235 | Yes | 2341 | 60 | |
29 | Unshielded detector at +1050V | 255 | Yes | 1966 | 60 |
1050V Plots | ||
1100V
1100V Source Out Data | ||||||
Run Number | Configuration | Discriminator Threshold (mV) | Cs-137 Button Source | Counts | Run Time (Sec) | Count Rate (Hz) |
1 | Unshielded detector at +1100V | 05 | No | 21898 | 60 | |
2 | Unshielded detector at +1100V | 10 | No | 60 | ||
3 | Unshielded detector at +1100V | 15 | No | 60 | ||
4 | Unshielded detector at +1100V | 20 | No | 60 | ||
5 | Unshielded detector at +1100V | 25 | No | 60 | ||
6 | Unshielded detector at +1100V | 30 | No | 60 | ||
7 | Unshielded detector at +1100V | 35 | No | 60 | ||
8 | Unshielded detector at +1100V | 45 | No | 60 | ||
9 | Unshielded detector at +1100V | 55 | No | 60 | ||
10 | Unshielded detector at +1100V | 75 | No | 60 | ||
11 | Unshielded detector at +1100V | 95 | No | 60 | ||
12 | Unshielded detector at +1100V | 115 | No | 60 | ||
13 | Unshielded detector at +1100V | 135 | No | 60 | ||
14 | Unshielded detector at +1100V | 155 | No | 60 | ||
15 | Unshielded detector at +1100V | 175 | No | 60 | ||
16 | Unshielded detector at +1100V | 195 | No | 60 | ||
17 | Unshielded detector at +1100V | 215 | No | 60 | ||
18 | Unshielded detector at +1100V | 235 | No | 60 | ||
19 | Unshielded detector at +1100V | 255 | No | 60 |
1100V Source In Data | ||||||
Run Number | Configuration | Discriminator Threshold (mV) | Cs-137 Button Source | Counts | Run Time (Sec) | Count Rate (Hz) |
1 | Unshielded detector at +1100V | 05 | Yes | 35704 | 60 | |
2 | Unshielded detector at +1100V | 10 | Yes | 32022 | 60 | |
3 | Unshielded detector at +1100V | 15 | Yes | 30121 | 60 | |
4 | Unshielded detector at +1100V | 20 | Yes | 28777 | 60 | |
5 | Unshielded detector at +1100V | 25 | Yes | 27499 | 60 | |
6 | Unshielded detector at +1100V | 30 | Yes | 26482 | 60 | |
7 | Unshielded detector at +1100V | 35 | Yes | 25219 | 60 | |
8 | Unshielded detector at +1100V | 45 | Yes | 23069 | 60 | |
9 | Unshielded detector at +1100V | 55 | Yes | 21675 | 60 | |
10 | Unshielded detector at +1100V | 75 | Yes | 18171 | 60 | |
11 | Unshielded detector at +1100V | 95 | Yes | 15749 | 60 | |
12 | Unshielded detector at +1100V | 115 | Yes | 13095 | 60 | |
13 | Unshielded detector at +1100V | 135 | Yes | 11453 | 60 | |
14 | Unshielded detector at +1100V | 155 | Yes | 10482 | 60 | |
15 | Unshielded detector at +1100V | 175 | Yes | 9366 | 60 | |
16 | Unshielded detector at +1100V | 195 | Yes | 8606 | 60 | |
17 | Unshielded detector at +1100V | 215 | Yes | 7853 | 60 | |
18 | Unshielded detector at +1100V | 235 | Yes | 7148 | 60 | |
19 | Unshielded detector at +1100V | 255 | Yes | 6642 | 60 |
1100V Plots | ||
1200V
1200V Source Out Data | ||||||
Run Number | Configuration | Discriminator Threshold (mV) | Cs-137 Button Source | Counts | Run Time (Sec) | Count Rate (Hz) |
1 | Unshielded detector at +1200V | 05 | No | 24052 | 60 | |
2 | Unshielded detector at +1200V | 15 | No | 20684 | 60 | |
3 | Unshielded detector at +1200V | 35 | No | 18324 | 60 | |
4 | Unshielded detector at +1200V | 55 | No | 17266 | 60 | |
5 | Unshielded detector at +1200V | 75 | No | 15643 | 60 | |
6 | Unshielded detector at +1200V | 95 | No | 14223 | 60 | |
7 | Unshielded detector at +1200V | 115 | No | 12792 | 60 | |
8 | Unshielded detector at +1200V | 135 | No | 11691 | 60 | |
9 | Unshielded detector at +1200V | 155 | No | 10414 | 60 | |
10 | Unshielded detector at +1200V | 175 | No | 9245 | 60 | |
11 | Unshielded detector at +1200V | 195 | No | 8574 | 60 | |
12 | Unshielded detector at +1200V | 215 | No | 7665 | 60 | |
13 | Unshielded detector at +1200V | 235 | No | 7008 | 60 | |
14 | Unshielded detector at +1200V | 255 | No | 6386 | 60 |
1200V Source In Data | ||||||
Run Number | Configuration | Discriminator Threshold (mV) | Cs-137 Button Source | Counts | Run Time (Sec) | Count Rate (Hz) |
1 | Unshielded detector at +1200V | 05 | Yes | 39450 | 60 | |
2 | Unshielded detector at +1200V | 15 | Yes | 34061 | 60 | |
3 | Unshielded detector at +1200V | 35 | Yes | 29029 | 60 | |
4 | Unshielded detector at +1200V | 55 | Yes | 27672 | 60 | |
5 | Unshielded detector at +1200V | 75 | Yes | 25935 | 60 | |
6 | Unshielded detector at +1200V | 95 | Yes | 24104 | 60 | |
7 | Unshielded detector at +1200V | 115 | Yes | 22223 | 60 | |
8 | Unshielded detector at +1200V | 135 | Yes | 20670 | 60 | |
9 | Unshielded detector at +1200V | 155 | Yes | 19168 | 60 | |
10 | Unshielded detector at +1200V | 175 | Yes | 17920 | 60 | |
11 | Unshielded detector at +1200V | 195 | Yes | 16383 | 60 | |
12 | Unshielded detector at +1200V | 215 | Yes | 15066 | 60 | |
13 | Unshielded detector at +1200V | 235 | Yes | 14130 | 60 | |
14 | Unshielded detector at +1200V | 255 | Yes | 13382 | 60 |
1200V Plots | ||
1300V
1300V Source Out Data | ||||||
Run Number | Configuration | Discriminator Threshold (mV) | Cs-137 Button Source | Counts | Run Time (Sec) | Count Rate (Hz) |
1 | Unshielded detector at +1300V | 05 | No | 24651 | 60 | |
2 | Unshielded detector at +1300V | 10 | No | 22487 | 60 | |
3 | Unshielded detector at +1300V | 15 | No | 21029 | 60 | |
4 | Unshielded detector at +1300V | 20 | No | 20535 | 60 | |
5 | Unshielded detector at +1300V | 25 | No | 19313 | 60 | |
6 | Unshielded detector at +1300V | 30 | No | 18947 | 60 | |
7 | Unshielded detector at +1300V | 35 | No | 18654 | 60 | |
8 | Unshielded detector at +1300V | 45 | No | 18405 | 60 | |
9 | Unshielded detector at +1300V | 55 | No | 18078 | 60 | |
10 | Unshielded detector at +1300V | 65 | No | 17830 | 60 | |
11 | Unshielded detector at +1300V | 75 | No | 17463 | 60 | |
12 | Unshielded detector at +1300V | 80 | No | 17303 | 60 | |
13 | Unshielded detector at +1300V | 85 | No | 17542 | 60 | |
14 | Unshielded detector at +1300V | 90 | No | 17342 | 60 | |
15 | Unshielded detector at +1300V | 95 | No | 17276 | 60 | |
16 | Unshielded detector at +1300V | 100 | No | 17350 | 60 | |
17 | Unshielded detector at +1300V | 105 | No | 16885 | 60 | |
18 | Unshielded detector at +1300V | 110 | No | 16842 | 60 | |
19 | Unshielded detector at +1300V | 115 | No | 16801 | 60 | |
20 | Unshielded detector at +1300V | 120 | No | 16599 | 60 | |
21 | Unshielded detector at +1300V | 125 | No | 16283 | 60 | |
22 | Unshielded detector at +1300V | 130 | No | 16292 | 60 | |
23 | Unshielded detector at +1300V | 135 | No | 16165 | 60 | |
24 | Unshielded detector at +1300V | 155 | No | 15329 | 60 | |
25 | Unshielded detector at +1300V | 175 | No | 14452 | 60 | |
26 | Unshielded detector at +1300V | 195 | No | 14130 | 60 | |
27 | Unshielded detector at +1300V | 215 | No | 12954 | 60 | |
28 | Unshielded detector at +1300V | 235 | No | 12155 | 60 | |
29 | Unshielded detector at +1300V | 255 | No | 11392 | 60 |
1300V Source In Data | ||||||
Run Number | Configuration | Discriminator Threshold (mV) | Cs-137 Button Source | Counts | Run Time (Sec) | Count Rate (Hz) |
1 | Unshielded detector at +1300V | 05 | Yes | 36726 | 60 | |
2 | Unshielded detector at +1300V | 10 | Yes | 33627 | 60 | |
3 | Unshielded detector at +1300V | 15 | Yes | 31514 | 60 | |
4 | Unshielded detector at +1300V | 20 | Yes | 30182 | 60 | |
5 | Unshielded detector at +1300V | 25 | Yes | 29448 | 60 | |
6 | Unshielded detector at +1300V | 30 | Yes | 28211 | 60 | |
7 | Unshielded detector at +1300V | 35 | Yes | 27912 | 60 | |
8 | Unshielded detector at +1300V | 45 | Yes | 26890 | 60 | |
9 | Unshielded detector at +1300V | 55 | Yes | 26347 | 60 | |
10 | Unshielded detector at +1300V | 65 | Yes | 26055 | 60 | |
11 | Unshielded detector at +1300V | 75 | Yes | 25154 | 60 | |
12 | Unshielded detector at +1300V | 80 | Yes | 25277 | 60 | |
13 | Unshielded detector at +1300V | 85 | Yes | 25219 | 60 | |
14 | Unshielded detector at +1300V | 90 | Yes | 25103 | 60 | |
15 | Unshielded detector at +1300V | 95 | Yes | 25104 | 60 | |
16 | Unshielded detector at +1300V | 100 | Yes | 25059 | 60 | |
17 | Unshielded detector at +1300V | 105 | Yes | 24800 | 60 | |
18 | Unshielded detector at +1300V | 110 | Yes | 24749 | 60 | |
19 | Unshielded detector at +1300V | 115 | Yes | 24944 | 60 | |
20 | Unshielded detector at +1300V | 120 | Yes | 24233 | 60 | |
21 | Unshielded detector at +1300V | 125 | Yes | 24260 | 60 | |
22 | Unshielded detector at +1300V | 130 | Yes | 23815 | 60 | |
23 | Unshielded detector at +1300V | 135 | Yes | 23766 | 60 | |
24 | Unshielded detector at +1300V | 155 | Yes | 23158 | 60 | |
25 | Unshielded detector at +1300V | 175 | Yes | 22176 | 60 | |
26 | Unshielded detector at +1300V | 195 | Yes | 21476 | 60 | |
27 | Unshielded detector at +1300V | 215 | Yes | 20375 | 60 | |
28 | Unshielded detector at +1300V | 235 | Yes | 19628 | 60 | |
29 | Unshielded detector at +1300V | 255 | Yes | 18512 | 60 |
1300V Plots | ||
1400V
1400V Source Out Data | ||||||
Run Number | Configuration | Discriminator Threshold (mV) | Cs-137 Button Source | Counts | Run Time (Sec) | Count Rate (Hz) |
1 | Unshielded detector at +1400V | 05 | No | 22873 | 60 | |
2 | Unshielded detector at +1400V | 10 | No | 20892 | 60 | |
3 | Unshielded detector at +1400V | 15 | No | 19899 | 60 | |
4 | Unshielded detector at +1400V | 20 | No | 19398 | 60 | |
5 | Unshielded detector at +1400V | 25 | No | 18861 | 60 | |
6 | Unshielded detector at +1400V | 30 | No | 18587 | 60 | |
7 | Unshielded detector at +1400V | 35 | No | 18053 | 60 | |
8 | Unshielded detector at +1400V | 45 | No | 17563 | 60 | |
9 | Unshielded detector at +1400V | 55 | No | 17356 | 60 | |
10 | Unshielded detector at +1400V | 65 | No | 17347 | 60 | |
11 | Unshielded detector at +1400V | 75 | No | 17329 | 60 | |
12 | Unshielded detector at +1400V | 80 | No | 17164 | 60 | |
13 | Unshielded detector at +1400V | 85 | No | 17146 | 60 | |
14 | Unshielded detector at +1400V | 90 | No | 17268 | 60 | |
15 | Unshielded detector at +1400V | 95 | No | 17234 | 60 | |
16 | Unshielded detector at +1400V | 100 | No | 17231 | 60 | |
17 | Unshielded detector at +1400V | 105 | No | 17119 | 60 | |
18 | Unshielded detector at +1400V | 110 | No | 17119 | 60 | |
19 | Unshielded detector at +1400V | 115 | No | 16956 | 60 | |
20 | Unshielded detector at +1400V | 120 | No | 17135 | 60 | |
21 | Unshielded detector at +1400V | 125 | No | 16680 | 60 | |
22 | Unshielded detector at +1400V | 130 | No | 17093 | 60 | |
23 | Unshielded detector at +1400V | 135 | No | 16893 | 60 | |
24 | Unshielded detector at +1400V | 155 | No | 16909 | 60 | |
25 | Unshielded detector at +1400V | 175 | No | 16489 | 60 | |
26 | Unshielded detector at +1400V | 195 | No | 16340 | 60 | |
27 | Unshielded detector at +1400V | 215 | No | 15683 | 60 | |
28 | Unshielded detector at +1400V | 235 | No | 15707 | 60 | |
29 | Unshielded detector at +1400V | 255 | No | 15411 | 60 |
1400V Source In Data | ||||||
Run Number | Configuration | Discriminator Threshold (mV) | Cs-137 Button Source | Counts | Run Time (Sec) | Count Rate (Hz) |
1 | Unshielded detector at +1400V | 05 | Yes | 36948 | 60 | |
2 | Unshielded detector at +1400V | 10 | Yes | 33637 | 60 | |
3 | Unshielded detector at +1400V | 15 | Yes | 31769 | 60 | |
4 | Unshielded detector at +1400V | 20 | Yes | 30477 | 60 | |
5 | Unshielded detector at +1400V | 25 | Yes | 28817 | 60 | |
6 | Unshielded detector at +1400V | 30 | Yes | 28523 | 60 | |
7 | Unshielded detector at +1400V | 35 | Yes | 27931 | 60 | |
8 | Unshielded detector at +1400V | 45 | Yes | 26824 | 60 | |
9 | Unshielded detector at +1400V | 55 | Yes | 26639 | 60 | |
10 | Unshielded detector at +1400V | 65 | Yes | 25810 | 60 | |
11 | Unshielded detector at +1400V | 75 | Yes | 26154 | 60 | |
12 | Unshielded detector at +1400V | 80 | Yes | 26157 | 60 | |
13 | Unshielded detector at +1400V | 85 | Yes | 25888 | 60 | |
14 | Unshielded detector at +1400V | 90 | Yes | 25679 | 60 | |
15 | Unshielded detector at +1400V | 95 | Yes | 25026 | 60 | |
16 | Unshielded detector at +1400V | 100 | Yes | 25533 | 60 | |
17 | Unshielded detector at +1400V | 105 | Yes | 25389 | 60 | |
18 | Unshielded detector at +1400V | 110 | Yes | 25223 | 60 | |
19 | Unshielded detector at +1400V | 115 | Yes | 24894 | 60 | |
20 | Unshielded detector at +1400V | 120 | Yes | 25054 | 60 | |
21 | Unshielded detector at +1400V | 125 | Yes | 25116 | 60 | |
22 | Unshielded detector at +1400V | 130 | Yes | 25117 | 60 | |
23 | Unshielded detector at +1400V | 135 | Yes | 24980 | 60 | |
24 | Unshielded detector at +1400V | 155 | Yes | 24495 | 60 | |
25 | Unshielded detector at +1400V | 175 | Yes | 24411 | 60 | |
26 | Unshielded detector at +1400V | 195 | Yes | 24345 | 60 | |
27 | Unshielded detector at +1400V | 215 | Yes | 23909 | 60 | |
28 | Unshielded detector at +1400V | 235 | Yes | 23306 | 60 | |
29 | Unshielded detector at +1400V | 255 | Yes | 23191 | 60 |
1400V Plots | ||