Timing filter Amp Configuration
Course gain: x6
Fine gain: at minimum
Integrate (nS): 500nS
Diff (nS): Out
Invert/Noninverted: Non-Inv
Dual Timer Configuration
Setup such that input discriminated pulse is at width ~200ns
Out feeds into scalar for counts
Top half of dual timer used to gate signal for scalar in plateau measurements
hello
i'm guessing this doesn't work
Plateau Data
700V
700V Source Out Data |
|
Run Number |
Configuration |
Discriminator Threshold (mV) |
Cs-137 Button Source |
Counts |
Run Time (Sec) |
Count Rate (Hz)
|
1 |
Unshielded detector at +700V |
05 |
No |
958 |
60 |
15.97
|
2 |
Unshielded detector at +700V |
15 |
No |
278 |
60 |
4.63
|
3 |
Unshielded detector at +700V |
35 |
No |
107 |
60 |
1.78
|
4 |
Unshielded detector at +700V |
55 |
No |
85 |
60 |
1.42
|
5 |
Unshielded detector at +700V |
75 |
No |
80 |
60 |
1.33
|
6 |
Unshielded detector at +700V |
95 |
No |
80 |
60 |
1.33
|
7 |
Unshielded detector at +700V |
115 |
No |
75 |
60 |
1.25
|
8 |
Unshielded detector at +700V |
135 |
No |
78 |
60 |
1.3
|
9 |
Unshielded detector at +700V |
155 |
No |
82 |
60 |
1.37
|
10 |
Unshielded detector at +700V |
175 |
No |
77 |
60 |
1.28
|
11 |
Unshielded detector at +700V |
195 |
No |
66 |
60 |
1.1
|
12 |
Unshielded detector at +700V |
215 |
No |
68 |
60 |
1.13
|
13 |
Unshielded detector at +700V |
235 |
No |
53 |
60 |
0.88
|
14 |
Unshielded detector at +700V |
255 |
No |
59 |
60 |
0.98
|
700V Source In Data |
|
Run Number |
Configuration |
Discriminator Threshold (mV) |
Cs-137 Button Source |
Counts |
Run Time (Sec) |
Count Rate (Hz)
|
1 |
Unshielded detector at +700V |
05 |
Yes |
1033 |
60 |
|
2 |
Unshielded detector at +700V |
15 |
Yes |
291 |
60 |
|
3 |
Unshielded detector at +700V |
35 |
Yes |
102 |
60 |
|
4 |
Unshielded detector at +700V |
55 |
Yes |
87 |
60 |
|
5 |
Unshielded detector at +700V |
75 |
Yes |
88 |
60 |
|
6 |
Unshielded detector at +700V |
95 |
Yes |
103 |
60 |
|
7 |
Unshielded detector at +700V |
115 |
Yes |
101 |
60 |
|
8 |
Unshielded detector at +700V |
135 |
Yes |
84 |
60 |
|
9 |
Unshielded detector at +700V |
155 |
Yes |
72 |
60 |
|
10 |
Unshielded detector at +700V |
175 |
Yes |
57 |
60 |
|
11 |
Unshielded detector at +700V |
195 |
Yes |
72 |
60 |
|
12 |
Unshielded detector at +700V |
215 |
Yes |
68 |
60 |
|
13 |
Unshielded detector at +700V |
235 |
Yes |
74 |
60 |
|
14 |
Unshielded detector at +700V |
255 |
Yes |
52 |
60 |
|
700V Plots |
|
Plateau measurements at 700V, source in and source out overlaid. |
Plateau measurements at 700V, source in minus source out.
|
750V
750V Source Out Data |
|
Run Number |
Configuration |
Discriminator Threshold (mV) |
Cs-137 Button Source |
Counts |
Run Time (Sec) |
Count Rate (Hz)
|
1 |
Unshielded detector at +750V |
05 |
No |
2211 |
60 |
36.85
|
2 |
Unshielded detector at +750V |
15 |
No |
1125 |
60 |
18.75
|
3 |
Unshielded detector at +750V |
35 |
No |
293 |
60 |
4.88
|
4 |
Unshielded detector at +750V |
55 |
No |
105 |
60 |
1.75
|
5 |
Unshielded detector at +750V |
75 |
No |
80 |
60 |
1.33
|
6 |
Unshielded detector at +750V |
95 |
No |
97 |
60 |
1.62
|
7 |
Unshielded detector at +750V |
115 |
No |
99 |
60 |
1.65
|
8 |
Unshielded detector at +750V |
135 |
No |
89 |
60 |
1.48
|
9 |
Unshielded detector at +750V |
155 |
No |
74 |
60 |
1.23
|
10 |
Unshielded detector at +750V |
175 |
No |
80 |
60 |
1.33
|
11 |
Unshielded detector at +750V |
195 |
No |
80 |
60 |
1.33
|
12 |
Unshielded detector at +750V |
215 |
No |
64 |
60 |
1.07
|
13 |
Unshielded detector at +750V |
235 |
No |
89 |
60 |
1.48
|
14 |
Unshielded detector at +750V |
255 |
No |
69 |
60 |
1.15
|
750V Source In Data |
|
Run Number |
Configuration |
Discriminator Threshold (mV) |
Cs-137 Button Source |
Counts |
Run Time (Sec) |
Count Rate (Hz)
|
1 |
Unshielded detector at +750V |
05 |
Yes |
4055 |
60 |
|
2 |
Unshielded detector at +750V |
15 |
Yes |
1160 |
60 |
|
3 |
Unshielded detector at +750V |
35 |
Yes |
301 |
60 |
|
4 |
Unshielded detector at +750V |
55 |
Yes |
102 |
60 |
|
5 |
Unshielded detector at +750V |
75 |
Yes |
82 |
60 |
|
6 |
Unshielded detector at +750V |
95 |
Yes |
114 |
60 |
|
7 |
Unshielded detector at +750V |
115 |
Yes |
85 |
60 |
|
8 |
Unshielded detector at +750V |
135 |
Yes |
93 |
60 |
|
9 |
Unshielded detector at +750V |
155 |
Yes |
87 |
60 |
|
10 |
Unshielded detector at +750V |
175 |
Yes |
87 |
60 |
|
11 |
Unshielded detector at +750V |
195 |
Yes |
79 |
60 |
|
12 |
Unshielded detector at +750V |
215 |
Yes |
90 |
60 |
|
13 |
Unshielded detector at +750V |
235 |
Yes |
84 |
60 |
|
14 |
Unshielded detector at +750V |
255 |
Yes |
73 |
60 |
|
750V Plots |
|
Plateau measurements at 750V, source in and source out overlaid. |
Plateau measurements at 750V, source in minus source out.
|
800V
800V Source Out Data |
|
Run Number |
Configuration |
Discriminator Threshold (mV) |
Cs-137 Button Source |
Counts |
Run Time (Sec) |
Count Rate (Hz)
|
1 |
Unshielded detector at +800V |
05 |
No |
3866 |
60 |
64.43
|
2 |
Unshielded detector at +800V |
15 |
No |
2356 |
60 |
39.27
|
3 |
Unshielded detector at +800V |
35 |
No |
1091 |
60 |
18.18
|
4 |
Unshielded detector at +800V |
55 |
No |
384 |
60 |
6.4
|
5 |
Unshielded detector at +800V |
75 |
No |
244 |
60 |
4.07
|
6 |
Unshielded detector at +800V |
95 |
No |
167 |
60 |
2.78
|
7 |
Unshielded detector at +800V |
115 |
No |
100 |
60 |
1.67
|
8 |
Unshielded detector at +800V |
135 |
No |
88 |
60 |
1.47
|
9 |
Unshielded detector at +800V |
155 |
No |
89 |
60 |
1.48
|
10 |
Unshielded detector at +800V |
175 |
No |
86 |
60 |
1.43
|
11 |
Unshielded detector at +800V |
195 |
No |
90 |
60 |
1.5
|
12 |
Unshielded detector at +800V |
215 |
No |
81 |
60 |
1.35
|
13 |
Unshielded detector at +800V |
235 |
No |
95 |
60 |
1.58
|
14 |
Unshielded detector at +800V |
255 |
No |
89 |
60 |
1.48
|
800V Source In Data |
|
Run Number |
Configuration |
Discriminator Threshold (mV) |
Cs-137 Button Source |
Counts |
Run Time (Sec) |
Count Rate (Hz)
|
1 |
Unshielded detector at +800V |
05 |
Yes |
10055 |
60 |
|
2 |
Unshielded detector at +800V |
15 |
Yes |
5438 |
60 |
|
3 |
Unshielded detector at +800V |
35 |
Yes |
1078 |
60 |
|
4 |
Unshielded detector at +800V |
55 |
Yes |
375 |
60 |
|
5 |
Unshielded detector at +800V |
75 |
Yes |
217 |
60 |
|
6 |
Unshielded detector at +800V |
95 |
Yes |
126 |
60 |
|
7 |
Unshielded detector at +800V |
115 |
Yes |
100 |
60 |
|
8 |
Unshielded detector at +800V |
135 |
Yes |
93 |
60 |
|
9 |
Unshielded detector at +800V |
155 |
Yes |
99 |
60 |
|
10 |
Unshielded detector at +800V |
175 |
Yes |
96 |
60 |
|
11 |
Unshielded detector at +800V |
195 |
Yes |
102 |
60 |
|
12 |
Unshielded detector at +800V |
215 |
Yes |
99 |
60 |
|
13 |
Unshielded detector at +800V |
235 |
Yes |
62 |
60 |
|
14 |
Unshielded detector at +800V |
255 |
Yes |
85 |
60 |
|
800V Plots |
|
Plateau measurements at 800V, source in and source out overlaid. |
Plateau measurements at 800V, source in minus source out.
|
850V
850V Source Out Data |
|
Run Number |
Configuration |
Discriminator Threshold (mV) |
Cs-137 Button Source |
Counts |
Run Time (Sec) |
Count Rate (Hz)
|
1 |
Unshielded detector at +850V |
05 |
No |
6402 |
60 |
106.7
|
2 |
Unshielded detector at +850V |
15 |
No |
3867 |
60 |
64.45
|
3 |
Unshielded detector at +850V |
35 |
No |
2058 |
60 |
34.3
|
4 |
Unshielded detector at +850V |
55 |
No |
1233 |
60 |
20.55
|
5 |
Unshielded detector at +850V |
75 |
No |
787 |
60 |
13.12
|
6 |
Unshielded detector at +850V |
95 |
No |
391 |
60 |
6.52
|
7 |
Unshielded detector at +850V |
115 |
No |
256 |
60 |
4.27
|
8 |
Unshielded detector at +850V |
135 |
No |
193 |
60 |
3.22
|
9 |
Unshielded detector at +850V |
155 |
No |
151 |
60 |
2.52
|
10 |
Unshielded detector at +850V |
175 |
No |
97 |
60 |
1.62
|
11 |
Unshielded detector at +850V |
195 |
No |
105 |
60 |
1.75
|
12 |
Unshielded detector at +850V |
215 |
No |
97 |
60 |
1.62
|
13 |
Unshielded detector at +850V |
235 |
No |
87 |
60 |
1.45
|
14 |
Unshielded detector at +850V |
255 |
No |
101 |
60 |
1.68
|
850V Source In Data |
|
Run Number |
Configuration |
Discriminator Threshold (mV) |
Cs-137 Button Source |
Counts |
Run Time (Sec) |
Count Rate (Hz)
|
1 |
Unshielded detector at +850V |
05 |
Yes |
15370 |
60 |
|
2 |
Unshielded detector at +850V |
15 |
Yes |
10298 |
60 |
|
3 |
Unshielded detector at +850V |
35 |
Yes |
2963 |
60 |
|
4 |
Unshielded detector at +850V |
55 |
Yes |
1366 |
60 |
|
5 |
Unshielded detector at +850V |
75 |
Yes |
842 |
60 |
|
6 |
Unshielded detector at +850V |
95 |
Yes |
402 |
60 |
|
7 |
Unshielded detector at +850V |
115 |
Yes |
281 |
60 |
|
8 |
Unshielded detector at +850V |
135 |
Yes |
196 |
60 |
|
9 |
Unshielded detector at +850V |
155 |
Yes |
143 |
60 |
|
10 |
Unshielded detector at +850V |
175 |
Yes |
98 |
60 |
|
11 |
Unshielded detector at +850V |
195 |
Yes |
88 |
60 |
|
12 |
Unshielded detector at +850V |
215 |
Yes |
101 |
60 |
|
13 |
Unshielded detector at +850V |
235 |
Yes |
104 |
60 |
|
14 |
Unshielded detector at +850V |
255 |
Yes |
101 |
60 |
|
850V Plots |
|
Plateau measurements at 850V, source in and source out overlaid. |
Plateau measurements at 850V, source in minus source out.
|
1000V
1000V Source Out Data |
|
Run Number |
Configuration |
Discriminator Threshold (mV) |
Cs-137 Button Source |
Counts |
Run Time (Sec) |
Count Rate (Hz)
|
1 |
Unshielded detector at +1000V |
05 |
No |
21167 |
60 |
|
2 |
Unshielded detector at +1000V |
10 |
No |
18739 |
60 |
|
3 |
Unshielded detector at +1000V |
15 |
No |
16091 |
60 |
|
4 |
Unshielded detector at +1000V |
20 |
No |
13773 |
60 |
|
5 |
Unshielded detector at +1000V |
25 |
No |
11983 |
60 |
|
6 |
Unshielded detector at +1000V |
30 |
No |
10358 |
60 |
|
7 |
Unshielded detector at +1000V |
35 |
No |
9274 |
60 |
|
8 |
Unshielded detector at +1000V |
45 |
No |
7423 |
60 |
|
9 |
Unshielded detector at +1000V |
55 |
No |
6067 |
60 |
|
10 |
Unshielded detector at +1000V |
65 |
No |
5258 |
60 |
|
11 |
Unshielded detector at +1000V |
75 |
No |
4451 |
60 |
|
12 |
Unshielded detector at +1000V |
80 |
No |
4321 |
60 |
|
13 |
Unshielded detector at +1000V |
85 |
No |
4111 |
60 |
|
14 |
Unshielded detector at +1000V |
90 |
No |
3823 |
60 |
|
15 |
Unshielded detector at +1000V |
95 |
No |
3666 |
60 |
|
16 |
Unshielded detector at +1000V |
100 |
No |
3466 |
60 |
|
17 |
Unshielded detector at +1000V |
105 |
No |
3241 |
60 |
|
18 |
Unshielded detector at +1000V |
110 |
No |
3180 |
60 |
|
19 |
Unshielded detector at +1000V |
115 |
No |
2968 |
60 |
|
20 |
Unshielded detector at +1000V |
120 |
No |
2812 |
60 |
|
21 |
Unshielded detector at +1000V |
125 |
No |
2769 |
60 |
|
22 |
Unshielded detector at +1000V |
130 |
No |
2548 |
60 |
|
23 |
Unshielded detector at +1000V |
135 |
No |
2426 |
60 |
|
24 |
Unshielded detector at +1000V |
155 |
No |
2065 |
60 |
|
25 |
Unshielded detector at +1000V |
175 |
No |
1843 |
60 |
|
26 |
Unshielded detector at +1000V |
195 |
No |
1582 |
60 |
|
27 |
Unshielded detector at +1000V |
215 |
No |
1374 |
60 |
|
28 |
Unshielded detector at +1000V |
235 |
No |
1172 |
60 |
|
29 |
Unshielded detector at +1000V |
255 |
No |
1012 |
60 |
|
1000V Source In Data |
|
Run Number |
Configuration |
Discriminator Threshold (mV) |
Cs-137 Button Source |
Counts |
Run Time (Sec) |
Count Rate (Hz)
|
1 |
Unshielded detector at +1000V |
05 |
Yes |
34443 |
60 |
|
2 |
Unshielded detector at +1000V |
10 |
Yes |
30427 |
60 |
|
3 |
Unshielded detector at +1000V |
15 |
Yes |
26882 |
60 |
|
4 |
Unshielded detector at +1000V |
20 |
Yes |
23779 |
60 |
|
5 |
Unshielded detector at +1000V |
25 |
Yes |
21672 |
60 |
|
6 |
Unshielded detector at +1000V |
30 |
Yes |
19611 |
60 |
|
7 |
Unshielded detector at +1000V |
35 |
Yes |
18115 |
60 |
|
8 |
Unshielded detector at +1000V |
45 |
Yes |
15119 |
60 |
|
9 |
Unshielded detector at +1000V |
55 |
Yes |
12824 |
60 |
|
10 |
Unshielded detector at +1000V |
65 |
Yes |
11575 |
60 |
|
11 |
Unshielded detector at +1000V |
75 |
Yes |
10320 |
60 |
|
12 |
Unshielded detector at +1000V |
80 |
Yes |
9658 |
60 |
|
13 |
Unshielded detector at +1000V |
85 |
Yes |
9328 |
60 |
|
14 |
Unshielded detector at +1000V |
90 |
Yes |
8885 |
60 |
|
15 |
Unshielded detector at +1000V |
95 |
Yes |
8328 |
60 |
|
16 |
Unshielded detector at +1000V |
100 |
Yes |
7906 |
60 |
|
17 |
Unshielded detector at +1000V |
105 |
Yes |
7777 |
60 |
|
18 |
Unshielded detector at +1000V |
110 |
Yes |
7478 |
60 |
|
19 |
Unshielded detector at +1000V |
115 |
Yes |
7313 |
60 |
|
20 |
Unshielded detector at +1000V |
120 |
Yes |
7206 |
60 |
|
21 |
Unshielded detector at +1000V |
125 |
Yes |
6947 |
60 |
|
22 |
Unshielded detector at +1000V |
130 |
Yes |
6936 |
60 |
|
23 |
Unshielded detector at +1000V |
135 |
Yes |
6692 |
60 |
|
24 |
Unshielded detector at +1000V |
155 |
Yes |
2242 |
60 |
|
25 |
Unshielded detector at +1000V |
175 |
Yes |
1878 |
60 |
|
26 |
Unshielded detector at +1000V |
195 |
Yes |
1596 |
60 |
|
27 |
Unshielded detector at +1000V |
215 |
Yes |
1378 |
60 |
|
28 |
Unshielded detector at +1000V |
235 |
Yes |
1280 |
60 |
|
29 |
Unshielded detector at +1000V |
255 |
Yes |
1092 |
60 |
|
1000V Plots |
|
Plateau measurements at 1000V, source in and source out overlaid. |
Plateau measurements at 1000V, source in minus source out.
|
1050V
Operating voltage when detectors used at IAC
1050V Source Out Data |
|
Run Number |
Configuration |
Discriminator Threshold (mV) |
Cs-137 Button Source |
Counts |
Run Time (Sec) |
Count Rate (Hz)
|
1 |
Unshielded detector at +1050V |
05 |
No |
22001 |
60 |
|
2 |
Unshielded detector at +1050V |
10 |
No |
20338 |
60 |
|
3 |
Unshielded detector at +1050V |
15 |
No |
18679 |
60 |
|
4 |
Unshielded detector at +1050V |
20 |
No |
16810 |
60 |
|
5 |
Unshielded detector at +1050V |
25 |
No |
15498 |
60 |
|
6 |
Unshielded detector at +1050V |
30 |
No |
14096 |
60 |
|
7 |
Unshielded detector at +1050V |
35 |
No |
13005 |
60 |
|
8 |
Unshielded detector at +1050V |
45 |
No |
10948 |
60 |
|
9 |
Unshielded detector at +1050V |
55 |
No |
9276 |
60 |
|
10 |
Unshielded detector at +1050V |
65 |
No |
8022 |
60 |
|
11 |
Unshielded detector at +1050V |
75 |
No |
7012 |
60 |
|
12 |
Unshielded detector at +1050V |
80 |
No |
6618 |
60 |
|
13 |
Unshielded detector at +1050V |
85 |
No |
6169 |
60 |
|
14 |
Unshielded detector at +1050V |
90 |
No |
5942 |
60 |
|
15 |
Unshielded detector at +1050V |
95 |
No |
5559 |
60 |
|
16 |
Unshielded detector at +1050V |
100 |
No |
5272 |
60 |
|
17 |
Unshielded detector at +1050V |
105 |
No |
4871 |
60 |
|
18 |
Unshielded detector at +1050V |
110 |
No |
4783 |
60 |
|
19 |
Unshielded detector at +1050V |
115 |
No |
4451 |
60 |
|
20 |
Unshielded detector at +1050V |
120 |
No |
4287 |
60 |
|
21 |
Unshielded detector at +1050V |
125 |
No |
4168 |
60 |
|
22 |
Unshielded detector at +1050V |
130 |
No |
4020 |
60 |
|
23 |
Unshielded detector at +1050V |
135 |
No |
3860 |
60 |
|
24 |
Unshielded detector at +1050V |
155 |
No |
3405 |
60 |
|
25 |
Unshielded detector at +1050V |
175 |
No |
3029 |
60 |
|
26 |
Unshielded detector at +1050V |
195 |
No |
2634 |
60 |
|
27 |
Unshielded detector at +1050V |
215 |
No |
2298 |
60 |
|
28 |
Unshielded detector at +1050V |
235 |
No |
2153 |
60 |
|
29 |
Unshielded detector at +1050V |
255 |
No |
2002 |
60 |
|
1050V Source In Data |
|
Run Number |
Configuration |
Discriminator Threshold (mV) |
Cs-137 Button Source |
Counts |
Run Time (Sec) |
Count Rate (Hz)
|
1 |
Unshielded detector at +1050V |
05 |
Yes |
34250 |
60 |
|
2 |
Unshielded detector at +1050V |
10 |
Yes |
31439 |
60 |
|
3 |
Unshielded detector at +1050V |
15 |
Yes |
29093 |
60 |
|
4 |
Unshielded detector at +1050V |
20 |
Yes |
26912 |
60 |
|
5 |
Unshielded detector at +1050V |
25 |
Yes |
25185 |
60 |
|
6 |
Unshielded detector at +1050V |
30 |
Yes |
23568 |
60 |
|
7 |
Unshielded detector at +1050V |
35 |
Yes |
21921 |
60 |
|
8 |
Unshielded detector at +1050V |
45 |
Yes |
19174 |
60 |
|
9 |
Unshielded detector at +1050V |
55 |
Yes |
17314 |
60 |
|
10 |
Unshielded detector at +1050V |
65 |
Yes |
15339 |
60 |
|
11 |
Unshielded detector at +1050V |
75 |
Yes |
13939 |
60 |
|
12 |
Unshielded detector at +1050V |
80 |
Yes |
13280 |
60 |
|
13 |
Unshielded detector at +1050V |
85 |
Yes |
12649 |
60 |
|
14 |
Unshielded detector at +1050V |
90 |
Yes |
12203 |
60 |
|
15 |
Unshielded detector at +1050V |
95 |
Yes |
11962 |
60 |
|
16 |
Unshielded detector at +1050V |
100 |
Yes |
11217 |
60 |
|
17 |
Unshielded detector at +1050V |
105 |
Yes |
10733 |
60 |
|
18 |
Unshielded detector at +1050V |
110 |
Yes |
10347 |
60 |
|
19 |
Unshielded detector at +1050V |
115 |
Yes |
9953 |
60 |
|
20 |
Unshielded detector at +1050V |
120 |
Yes |
9691 |
60 |
|
21 |
Unshielded detector at +1050V |
125 |
Yes |
9551 |
60 |
|
22 |
Unshielded detector at +1050V |
130 |
Yes |
9147 |
60 |
|
23 |
Unshielded detector at +1050V |
135 |
Yes |
8840 |
60 |
|
24 |
Unshielded detector at +1050V |
155 |
Yes |
7735 |
60 |
|
25 |
Unshielded detector at +1050V |
175 |
Yes |
6970 |
60 |
|
26 |
Unshielded detector at +1050V |
195 |
Yes |
6587 |
60 |
|
27 |
Unshielded detector at +1050V |
215 |
Yes |
5559 |
60 |
|
28 |
Unshielded detector at +1050V |
235 |
Yes |
2341 |
60 |
|
29 |
Unshielded detector at +1050V |
255 |
Yes |
1966 |
60 |
|
1050V Plots |
|
Plateau measurements at 1050V, source in and source out overlaid. |
Plateau measurements at 1050V, source in minus source out.
|
1100V
1100V Source Out Data |
|
Run Number |
Configuration |
Discriminator Threshold (mV) |
Cs-137 Button Source |
Counts |
Run Time (Sec) |
Count Rate (Hz)
|
1 |
Unshielded detector at +1100V |
05 |
No |
|
60 |
|
2 |
Unshielded detector at +1100V |
10 |
No |
|
60 |
|
3 |
Unshielded detector at +1100V |
15 |
No |
|
60 |
|
4 |
Unshielded detector at +1100V |
20 |
No |
|
60 |
|
5 |
Unshielded detector at +1100V |
25 |
No |
|
60 |
|
6 |
Unshielded detector at +1100V |
30 |
No |
|
60 |
|
7 |
Unshielded detector at +1100V |
35 |
No |
|
60 |
|
8 |
Unshielded detector at +1100V |
45 |
No |
|
60 |
|
9 |
Unshielded detector at +1100V |
55 |
No |
|
60 |
|
10 |
Unshielded detector at +1100V |
75 |
No |
|
60 |
|
11 |
Unshielded detector at +1100V |
95 |
No |
|
60 |
|
12 |
Unshielded detector at +1100V |
115 |
No |
|
60 |
|
13 |
Unshielded detector at +1100V |
135 |
No |
|
60 |
|
14 |
Unshielded detector at +1100V |
155 |
No |
|
60 |
|
15 |
Unshielded detector at +1100V |
175 |
No |
|
60 |
|
16 |
Unshielded detector at +1100V |
195 |
No |
|
60 |
|
17 |
Unshielded detector at +1100V |
215 |
No |
|
60 |
|
18 |
Unshielded detector at +1100V |
235 |
No |
|
60 |
|
19 |
Unshielded detector at +1100V |
255 |
No |
|
60 |
|
1100V Source In Data |
|
Run Number |
Configuration |
Discriminator Threshold (mV) |
Cs-137 Button Source |
Counts |
Run Time (Sec) |
Count Rate (Hz)
|
1 |
Unshielded detector at +1100V |
05 |
Yes |
36350 |
60 |
|
2 |
Unshielded detector at +1100V |
10 |
Yes |
32632 |
60 |
|
3 |
Unshielded detector at +1100V |
15 |
Yes |
30656 |
60 |
|
4 |
Unshielded detector at +1100V |
20 |
Yes |
29153 |
60 |
|
5 |
Unshielded detector at +1100V |
25 |
Yes |
|
60 |
|
6 |
Unshielded detector at +1100V |
30 |
Yes |
|
60 |
|
7 |
Unshielded detector at +1100V |
35 |
Yes |
|
60 |
|
8 |
Unshielded detector at +1100V |
45 |
Yes |
|
60 |
|
9 |
Unshielded detector at +1100V |
55 |
Yes |
|
60 |
|
10 |
Unshielded detector at +1100V |
75 |
Yes |
|
60 |
|
11 |
Unshielded detector at +1100V |
95 |
Yes |
|
60 |
|
12 |
Unshielded detector at +1100V |
115 |
Yes |
|
60 |
|
13 |
Unshielded detector at +1100V |
135 |
Yes |
|
60 |
|
14 |
Unshielded detector at +1100V |
155 |
Yes |
|
60 |
|
15 |
Unshielded detector at +1100V |
175 |
Yes |
|
60 |
|
16 |
Unshielded detector at +1100V |
195 |
Yes |
|
60 |
|
17 |
Unshielded detector at +1100V |
215 |
Yes |
|
60 |
|
18 |
Unshielded detector at +1100V |
235 |
Yes |
|
60 |
|
19 |
Unshielded detector at +1100V |
255 |
Yes |
|
60 |
|
1100V Plots |
|
Plateau measurements at 1100V, source in and source out overlaid. |
Plateau measurements at 1100V, source in minus source out.
|
1200V
1200V Source Out Data |
|
Run Number |
Configuration |
Discriminator Threshold (mV) |
Cs-137 Button Source |
Counts |
Run Time (Sec) |
Count Rate (Hz)
|
1 |
Unshielded detector at +1200V |
05 |
No |
24052 |
60 |
|
2 |
Unshielded detector at +1200V |
15 |
No |
20684 |
60 |
|
3 |
Unshielded detector at +1200V |
35 |
No |
18324 |
60 |
|
4 |
Unshielded detector at +1200V |
55 |
No |
17266 |
60 |
|
5 |
Unshielded detector at +1200V |
75 |
No |
15643 |
60 |
|
6 |
Unshielded detector at +1200V |
95 |
No |
14223 |
60 |
|
7 |
Unshielded detector at +1200V |
115 |
No |
12792 |
60 |
|
8 |
Unshielded detector at +1200V |
135 |
No |
11691 |
60 |
|
9 |
Unshielded detector at +1200V |
155 |
No |
10414 |
60 |
|
10 |
Unshielded detector at +1200V |
175 |
No |
9245 |
60 |
|
11 |
Unshielded detector at +1200V |
195 |
No |
8574 |
60 |
|
12 |
Unshielded detector at +1200V |
215 |
No |
7665 |
60 |
|
13 |
Unshielded detector at +1200V |
235 |
No |
7008 |
60 |
|
14 |
Unshielded detector at +1200V |
255 |
No |
6386 |
60 |
|
1200V Source In Data |
|
Run Number |
Configuration |
Discriminator Threshold (mV) |
Cs-137 Button Source |
Counts |
Run Time (Sec) |
Count Rate (Hz)
|
1 |
Unshielded detector at +1200V |
05 |
Yes |
39450 |
60 |
|
2 |
Unshielded detector at +1200V |
15 |
Yes |
34061 |
60 |
|
3 |
Unshielded detector at +1200V |
35 |
Yes |
29029 |
60 |
|
4 |
Unshielded detector at +1200V |
55 |
Yes |
27672 |
60 |
|
5 |
Unshielded detector at +1200V |
75 |
Yes |
25935 |
60 |
|
6 |
Unshielded detector at +1200V |
95 |
Yes |
24104 |
60 |
|
7 |
Unshielded detector at +1200V |
115 |
Yes |
22223 |
60 |
|
8 |
Unshielded detector at +1200V |
135 |
Yes |
20670 |
60 |
|
9 |
Unshielded detector at +1200V |
155 |
Yes |
19168 |
60 |
|
10 |
Unshielded detector at +1200V |
175 |
Yes |
17920 |
60 |
|
11 |
Unshielded detector at +1200V |
195 |
Yes |
16383 |
60 |
|
12 |
Unshielded detector at +1200V |
215 |
Yes |
15066 |
60 |
|
13 |
Unshielded detector at +1200V |
235 |
Yes |
14130 |
60 |
|
14 |
Unshielded detector at +1200V |
255 |
Yes |
13382 |
60 |
|
1200V Plots |
|
Plateau measurements at 1200V, source in and source out overlaid. |
Plateau measurements at 1200V, source in minus source out.
|
1300V
1300V Source Out Data |
|
Run Number |
Configuration |
Discriminator Threshold (mV) |
Cs-137 Button Source |
Counts |
Run Time (Sec) |
Count Rate (Hz)
|
1 |
Unshielded detector at +1300V |
05 |
No |
24651 |
60 |
|
2 |
Unshielded detector at +1300V |
10 |
No |
22487 |
60 |
|
3 |
Unshielded detector at +1300V |
15 |
No |
21029 |
60 |
|
4 |
Unshielded detector at +1300V |
20 |
No |
20535 |
60 |
|
5 |
Unshielded detector at +1300V |
25 |
No |
19313 |
60 |
|
6 |
Unshielded detector at +1300V |
30 |
No |
18947 |
60 |
|
7 |
Unshielded detector at +1300V |
35 |
No |
18654 |
60 |
|
8 |
Unshielded detector at +1300V |
45 |
No |
18405 |
60 |
|
9 |
Unshielded detector at +1300V |
55 |
No |
18078 |
60 |
|
10 |
Unshielded detector at +1300V |
65 |
No |
17830 |
60 |
|
11 |
Unshielded detector at +1300V |
75 |
No |
17463 |
60 |
|
12 |
Unshielded detector at +1300V |
80 |
No |
17303 |
60 |
|
13 |
Unshielded detector at +1300V |
85 |
No |
17542 |
60 |
|
14 |
Unshielded detector at +1300V |
90 |
No |
17342 |
60 |
|
15 |
Unshielded detector at +1300V |
95 |
No |
17276 |
60 |
|
16 |
Unshielded detector at +1300V |
100 |
No |
17350 |
60 |
|
17 |
Unshielded detector at +1300V |
105 |
No |
16885 |
60 |
|
18 |
Unshielded detector at +1300V |
110 |
No |
16842 |
60 |
|
19 |
Unshielded detector at +1300V |
115 |
No |
16801 |
60 |
|
20 |
Unshielded detector at +1300V |
120 |
No |
16599 |
60 |
|
21 |
Unshielded detector at +1300V |
125 |
No |
16283 |
60 |
|
22 |
Unshielded detector at +1300V |
130 |
No |
16292 |
60 |
|
23 |
Unshielded detector at +1300V |
135 |
No |
16165 |
60 |
|
24 |
Unshielded detector at +1300V |
155 |
No |
15329 |
60 |
|
25 |
Unshielded detector at +1300V |
175 |
No |
14452 |
60 |
|
26 |
Unshielded detector at +1300V |
195 |
No |
14130 |
60 |
|
27 |
Unshielded detector at +1300V |
215 |
No |
12954 |
60 |
|
28 |
Unshielded detector at +1300V |
235 |
No |
12155 |
60 |
|
29 |
Unshielded detector at +1300V |
255 |
No |
11392 |
60 |
|
1300V Source In Data |
|
Run Number |
Configuration |
Discriminator Threshold (mV) |
Cs-137 Button Source |
Counts |
Run Time (Sec) |
Count Rate (Hz)
|
1 |
Unshielded detector at +1300V |
05 |
Yes |
36726 |
60 |
|
2 |
Unshielded detector at +1300V |
10 |
Yes |
33627 |
60 |
|
3 |
Unshielded detector at +1300V |
15 |
Yes |
31514 |
60 |
|
4 |
Unshielded detector at +1300V |
20 |
Yes |
30182 |
60 |
|
5 |
Unshielded detector at +1300V |
25 |
Yes |
29448 |
60 |
|
6 |
Unshielded detector at +1300V |
30 |
Yes |
28211 |
60 |
|
7 |
Unshielded detector at +1300V |
35 |
Yes |
27912 |
60 |
|
8 |
Unshielded detector at +1300V |
45 |
Yes |
26890 |
60 |
|
9 |
Unshielded detector at +1300V |
55 |
Yes |
26347 |
60 |
|
10 |
Unshielded detector at +1300V |
65 |
Yes |
26055 |
60 |
|
11 |
Unshielded detector at +1300V |
75 |
Yes |
25154 |
60 |
|
12 |
Unshielded detector at +1300V |
80 |
Yes |
25277 |
60 |
|
13 |
Unshielded detector at +1300V |
85 |
Yes |
25219 |
60 |
|
14 |
Unshielded detector at +1300V |
90 |
Yes |
25103 |
60 |
|
15 |
Unshielded detector at +1300V |
95 |
Yes |
25104 |
60 |
|
16 |
Unshielded detector at +1300V |
100 |
Yes |
25059 |
60 |
|
17 |
Unshielded detector at +1300V |
105 |
Yes |
24800 |
60 |
|
18 |
Unshielded detector at +1300V |
110 |
Yes |
24749 |
60 |
|
19 |
Unshielded detector at +1300V |
115 |
Yes |
24944 |
60 |
|
20 |
Unshielded detector at +1300V |
120 |
Yes |
24233 |
60 |
|
21 |
Unshielded detector at +1300V |
125 |
Yes |
24260 |
60 |
|
22 |
Unshielded detector at +1300V |
130 |
Yes |
23815 |
60 |
|
23 |
Unshielded detector at +1300V |
135 |
Yes |
23766 |
60 |
|
24 |
Unshielded detector at +1300V |
155 |
Yes |
23158 |
60 |
|
25 |
Unshielded detector at +1300V |
175 |
Yes |
22176 |
60 |
|
26 |
Unshielded detector at +1300V |
195 |
Yes |
21476 |
60 |
|
27 |
Unshielded detector at +1300V |
215 |
Yes |
20375 |
60 |
|
28 |
Unshielded detector at +1300V |
235 |
Yes |
19628 |
60 |
|
29 |
Unshielded detector at +1300V |
255 |
Yes |
18512 |
60 |
|
1300V Plots |
|
Plateau measurements at 1300V, source in and source out overlaid. |
Plateau measurements at 1300V, source in minus source out.
|
1400V
1400V Source Out Data |
|
Run Number |
Configuration |
Discriminator Threshold (mV) |
Cs-137 Button Source |
Counts |
Run Time (Sec) |
Count Rate (Hz)
|
1 |
Unshielded detector at +1400V |
05 |
No |
22873 |
60 |
|
2 |
Unshielded detector at +1400V |
10 |
No |
20892 |
60 |
|
3 |
Unshielded detector at +1400V |
15 |
No |
19899 |
60 |
|
4 |
Unshielded detector at +1400V |
20 |
No |
19398 |
60 |
|
5 |
Unshielded detector at +1400V |
25 |
No |
18861 |
60 |
|
6 |
Unshielded detector at +1400V |
30 |
No |
18587 |
60 |
|
7 |
Unshielded detector at +1400V |
35 |
No |
18053 |
60 |
|
8 |
Unshielded detector at +1400V |
45 |
No |
17563 |
60 |
|
9 |
Unshielded detector at +1400V |
55 |
No |
17356 |
60 |
|
10 |
Unshielded detector at +1400V |
65 |
No |
17347 |
60 |
|
11 |
Unshielded detector at +1400V |
75 |
No |
17329 |
60 |
|
12 |
Unshielded detector at +1400V |
80 |
No |
17164 |
60 |
|
13 |
Unshielded detector at +1400V |
85 |
No |
17146 |
60 |
|
14 |
Unshielded detector at +1400V |
90 |
No |
17268 |
60 |
|
15 |
Unshielded detector at +1400V |
95 |
No |
17234 |
60 |
|
16 |
Unshielded detector at +1400V |
100 |
No |
17231 |
60 |
|
17 |
Unshielded detector at +1400V |
105 |
No |
17119 |
60 |
|
18 |
Unshielded detector at +1400V |
110 |
No |
17119 |
60 |
|
19 |
Unshielded detector at +1400V |
115 |
No |
16956 |
60 |
|
20 |
Unshielded detector at +1400V |
120 |
No |
17135 |
60 |
|
21 |
Unshielded detector at +1400V |
125 |
No |
16680 |
60 |
|
22 |
Unshielded detector at +1400V |
130 |
No |
17093 |
60 |
|
23 |
Unshielded detector at +1400V |
135 |
No |
16893 |
60 |
|
24 |
Unshielded detector at +1400V |
155 |
No |
16909 |
60 |
|
25 |
Unshielded detector at +1400V |
175 |
No |
16489 |
60 |
|
26 |
Unshielded detector at +1400V |
195 |
No |
16340 |
60 |
|
27 |
Unshielded detector at +1400V |
215 |
No |
15683 |
60 |
|
28 |
Unshielded detector at +1400V |
235 |
No |
15707 |
60 |
|
29 |
Unshielded detector at +1400V |
255 |
No |
15411 |
60 |
|
1400V Source In Data |
|
Run Number |
Configuration |
Discriminator Threshold (mV) |
Cs-137 Button Source |
Counts |
Run Time (Sec) |
Count Rate (Hz)
|
1 |
Unshielded detector at +1400V |
05 |
Yes |
36948 |
60 |
|
2 |
Unshielded detector at +1400V |
10 |
Yes |
33637 |
60 |
|
3 |
Unshielded detector at +1400V |
15 |
Yes |
31769 |
60 |
|
4 |
Unshielded detector at +1400V |
20 |
Yes |
30477 |
60 |
|
5 |
Unshielded detector at +1400V |
25 |
Yes |
28817 |
60 |
|
6 |
Unshielded detector at +1400V |
30 |
Yes |
28523 |
60 |
|
7 |
Unshielded detector at +1400V |
35 |
Yes |
27931 |
60 |
|
8 |
Unshielded detector at +1400V |
45 |
Yes |
26824 |
60 |
|
9 |
Unshielded detector at +1400V |
55 |
Yes |
26639 |
60 |
|
10 |
Unshielded detector at +1400V |
65 |
Yes |
25810 |
60 |
|
11 |
Unshielded detector at +1400V |
75 |
Yes |
26154 |
60 |
|
12 |
Unshielded detector at +1400V |
80 |
Yes |
26157 |
60 |
|
13 |
Unshielded detector at +1400V |
85 |
Yes |
25888 |
60 |
|
14 |
Unshielded detector at +1400V |
90 |
Yes |
25679 |
60 |
|
15 |
Unshielded detector at +1400V |
95 |
Yes |
25026 |
60 |
|
16 |
Unshielded detector at +1400V |
100 |
Yes |
25533 |
60 |
|
17 |
Unshielded detector at +1400V |
105 |
Yes |
25389 |
60 |
|
18 |
Unshielded detector at +1400V |
110 |
Yes |
25223 |
60 |
|
19 |
Unshielded detector at +1400V |
115 |
Yes |
24894 |
60 |
|
20 |
Unshielded detector at +1400V |
120 |
Yes |
25054 |
60 |
|
21 |
Unshielded detector at +1400V |
125 |
Yes |
25116 |
60 |
|
22 |
Unshielded detector at +1400V |
130 |
Yes |
25117 |
60 |
|
23 |
Unshielded detector at +1400V |
135 |
Yes |
24980 |
60 |
|
24 |
Unshielded detector at +1400V |
155 |
Yes |
24495 |
60 |
|
25 |
Unshielded detector at +1400V |
175 |
Yes |
24411 |
60 |
|
26 |
Unshielded detector at +1400V |
195 |
Yes |
24345 |
60 |
|
27 |
Unshielded detector at +1400V |
215 |
Yes |
23909 |
60 |
|
28 |
Unshielded detector at +1400V |
235 |
Yes |
23306 |
60 |
|
29 |
Unshielded detector at +1400V |
255 |
Yes |
23191 |
60 |
|
1400V Plots |
|
Plateau measurements at 1400V, source in and source out overlaid. |
Plateau measurements at 1400V, source in minus source out.
|
Go Back
NaI_JD-766