Calibration Info
To fully understand the relationship between PMT Counts and total exposure for the Nanodot OSLs, a custom calibration is needed. This calibration is used in lieu of the calibration created by the OSL reader, which is created using the pre-dosed OSLs from the manufacturer. To begin the calibration, a set of fifteen unexposed Nanodot OSLs is chosen at random and exposed to a 9.3Ci source. Using a distance of 30cm from the faceplate of the source and also including the distance from the faceplate to the surface of the source, , the exposure rate is calculated in Roentgen using . The gamma factor used for these calculations is
,
There is an adjustment distance of 11.2cm which accounts for the distance from the faceplate to the surface of the source
Total exposures are calculated by