OSL Reader

From New IAC Wiki
Revision as of 17:19, 1 November 2017 by Harpconn (talk | contribs)
Jump to navigation Jump to search

The OSL reader is a Landauer MicroStar reader. It uses a photo multiplier tube to count the hits off of the crystal structure in the OSLs when an LED is shining through it.

The measured dose is calculated through the following equation:

[math] \frac{PMT Counts}{(Calibration\ factor) (Sensitivity)} = Dose\ \text{(mRad)} [/math]



Thesis