Se170063 Nickel Foil Thin Gauss Window Expansion

From New IAC Wiki
Revision as of 14:41, 26 October 2017 by Brenleyt (talk | contribs)
Jump to navigation Jump to search

This analysis was done by weighting the histogram by the mass of the sample. The window is [1360,1365] and was expanded by one channel on each side to find the error in the counts. Since the window is smaller, a constant fit was no long acceptable as the constant kept becoming negative and would yield a negative integral, which is not a good representation of background that is clearly there. Instead I used a linear function plus a gaussian to fit, then integrated the line across the window.

5/25/17 5/26/17 5/27/17 5/29/17 5/30/17 5/31/17 6/1/17
Original Window Counts
Original Window Background (Integrated)
Original Window Difference
Expanded Window Counts
Expanded Window Background
Expanded Window Difference
Error in counts
Position 50 cm 30 cm 20cm 20 cm 10cm 10cm 10cm
Efficiency 0.00017 0.000422 0.000847 0.000847 0.0031 0.0031 0.0031
.dat file entry


Below is a plot of the activity as a function of time

170063 Nickel WideGauss HLPlot.png

This gives the constant of 14.5046 +/- 0.00345043. Now find the initial activity

[math] A_t = e^{14.5046} = 1991900.96 Hz [/math]

Since the plot is made with reference to the mixture's measurement time (first point plotted 108 minutes after the mix was measured) means that the constant should be the activity at the time the mixture was being measured.