LB Feb 2017 Energy Efficiency Calculations

From New IAC Wiki
Jump to navigation Jump to search

Theoretical Activities

Na-22, 9.427micro Ci on July 1, 2008, half life 2.602 +/- 0.002 years, 99.937% for 1274.52 and 178.8 for 511 line , activity in March 31, 2016 =1.196micro Ci

= [math]\left (0.99937 \right )\left ( 1.196 \times 10^{-6} \mbox{Ci} \right) \left (\frac{ (3.7 \times 10^{10} \mbox{Hz}}{\mbox{Ci}} \right)= 44,224 Hz [/math] for the 1274 line
= [math]\left (1.788 \right )\left ( 1.196 \times 10^{-6} \mbox{Ci} \right) \left (\frac{ (3.7 \times 10^{10} \mbox{Hz}}{\mbox{Ci}} \right)= 79122.6 Hz [/math] for the 511 line

Cs-137, 661.660 line, 85.21% * 1.066micro Ci on July 1, 2008, half life 30.0 +/- 0.2 yrs, March 31, 2016 activity = 0.891micro Ci expected rate for 661 line

= [math]\left (0.8521 \right )\left ( 0.891 \times 10^{-6} \mbox{Ci} \right) \left (\frac{ (3.7 \times 10^{10} \mbox{Hz}}{\mbox{Ci}} \right)= 28091.2 Hz [/math]

Mn-54, 11.77 microCi on July 1, 2008, half life =312.20 +/- 0.07 days, 99.975% intensity on 834.826 , March 31, 2016 activity =0.02328micro Ci

= [math]\left (0.99975 \right )\left ( 0.02328 \times 10^{-6} \mbox{Ci} \right) \left (\frac{ (3.7 \times 10^{10} \mbox{Hz}}{\mbox{Ci}} \right)= 861.1 Hz [/math] for the 834 line


Co-60, 10.42micro Ci July 1, 2008, half life 5.271 +/- 0.001 years, 99.0 % for 1173.237 and 99.9824 % for 1332.501, March 31, 2016 activity=3.759micro Ci


= [math]\left (0.99 \right )\left ( 3.759 \times 10^{-6} \mbox{Ci} \right) \left (\frac{ (3.7 \times 10^{10} \mbox{Hz}}{\mbox{Ci}} \right)= 137692.17 Hz [/math] for the 1173 line


= [math]\left (0.999824 \right )\left ( 3.759 \times 10^{-6} \mbox{Ci} \right) \left (\frac{ (3.7 \times 10^{10} \mbox{Hz}}{\mbox{Ci}} \right)= 139058.52 Hz [/math] for the 1332 line

Actual Activites

This is an example of the method that will be used to find the activities in the calibrated source's spectra

The line of interest in Cs-137 the 661.7 line.


Using the program BackgroundSubtract.C in /data/IAC/Se/Feb_2017_Efficiency/Energy_Efficiency and the root command


BackgroundSubtract("r5818.root","r5891.root",412,3369);

The 661.7 keV line can be seen

Cs137 Activity.png

Note that the range along the x-axis is [640:680] and the integral is 208.3.

Taking the ratio of the rate seen in the detector and the theoretical rate calculated above, one can arrive at a detector efficiency of 0.7%

The rest of the data is on the wiki page

LB_Efficiency_Plots_Feb_2017

Energy Efficiencies

Below is a table with the calculated efficiencies

Run Source Energy (keV) Expected Rate (Hz) ROOT Window HpGe Rate (Hz) HpGe X-Cool Detector Efficiency (%)
r5818 Cs-137 661.7 28091.2 [640:680] 208.3 0.7