Attempted to see Se lines in Horse mineral supplement that contains 16 PPM of Se.
Files on PC stored in
cd:\MPA3(x64)\data\TonyForest\Se_Run_000
calibration at
https://wiki.iac.isu.edu/index.php/SeRun_01-11-16
Calibration for Detector A (Channel 0) on 3-6-16
Run Number | Isotope | Egamma (keV) | Channel |
Calib002 | Na-22 | 511 | 559.4 \pm 1.408 |
Calib002 | Na-22 | 1274.537 | 1376 \pm 1.255 |
Calib004 | Mn-54 | 834.848 | 905.5 \pm 0.9194 |
Calib005 | Ba-133 | 80.9971 | 99.77 \pm 0.6394 |
Calib005 | Ba-133 | 276.3997 | 308.7 \pm 0.8055 |
Calib005 | Ba-133 | 302.8510 | 337 \pm 0.7719 |
Calib005 | Ba-133 | 356.0134 | 393.8 \pm 0.7181 |
Calib005 | Ba-133 | 383.8480 | 423.5 \pm 0.7247 |
Calib006 | Cs-137 | 661.657 | 720.5 \pm 0.8931 |
Run | Acquisition Time(sec) | Description |
AccOnAlInABackgInD001 | ADC1A:14883.26, ADC1D:15020.678 | The Aluminum tube is in detector A, nothing in Detector B, the Accelerator is On |
AccOnAlInASe-AinDetD_001 | ADC1A:4881.622, ADC1D:4835.252 | The Aluminum tube is in detector A, nothing in Detector B, the Accelerator is On |