PePPO 03132012
Jump to navigation
Jump to search
Accelerator Configuration
- Accelerator Settings
beam energy | 6.8 MeV |
Beam type: | Electron |
I_average: | 0.3 nA |
I_peak: | 0.05 | A 0.006 duty factor
pulse width: | 10 | s (it seems on 200 s is possible?)
Rate: | 60 Hz |
PEPPO Spectrometer: | 151.3 A |
Set HV to Detectors
Signal channel | HV dac | HV (V) | |
Detectors #1 | A1 | 2.4 | 960 |
Detectors #2 | A2 | 3.0 | 1200 |
Scope Pictures
CH1: Beam Sync CH2: A1 CH3: A2
Target | Pictures |
No Target | |
Chromox Target |
Run Plan
- Signal Definitions for Trigger logic
- SYNC:=60 Hz sync pulse 200 micro seconds long
- Det1:= Annihilation Counter #1 (HV1)
- Det2:=Annihilation Counter #2 (HV2)
Please make the following measurements
Run# | run mode | Run length | Beam | Target | Trig Logic | VME discriminator Threshold (mV) | HV1 (setting/V) | HV2 (setting/V) | PL (setting/ns) | PTW (setting/ns) |
2902 | sample | 5 minutes | e | Chromox | (Det1 OR Det2) | 50 | 2.4/960 | 3/1200 | ||
2903 | sample | 5 minutes | e | Empty | (Det1 OR Det2) | 50 | 2.4/960 | 3/1200 | ||
2905 | sample | 10 minutes | e | Chromox | Det1 AND Det2 | 50 | 2.4/960 | 3/1200 | ||
2904 | sample | 15 minutes | e | Empty | Det1 AND Det2 | 50 | 2.4/960 | 3/1200 | ||
2908 | sample | 30 minutes After Beam Studies end | off | out | (Det1 or Det2) | 50 | 2.4/960 | 3/1200 | ||
2909 | sample | 30 minutes After Beam Studies end | off | out | (Det1 AND Det2) | 50 | 2.4/960 | 3/1200 |