GEM Cosmic V792 Measurements
DAQ setup
Insert block diagram identifying all electronics modules Then insert signal scope picture from Standford pulse generator Now insert the ADC measurements (ADC1,ADC5, and ADC1/ADC5)
GEM Measurements
11-07-08
The pulse coming from the strips is delayed and amplified using the Timing Filter Amplifier Model_474(X6). The amplified pulse is connected to the V792 QDC.
Run 550
Start Time: Nov 6 19:58:03; End Time: Nov 7 09:45:57
12-10-08
The signal below represents a measurement of the Trig out signal from a 10 cm x 10 cm ionization chamber with 3 GEM preamplifiers. The chamber gas a 90/10 % mixture of Arcon and CO2. The Drift HV was set to 3200 Volts and the GEM preamplifier HV distribution network was set to 2900 Volts. The V792 CAEN ADC channels have a gain of 100 fC/channel. The main signal peak appears at channel number 759 and represents an accumulation of
pC of charge.Electronics Settings:
TrigOut goes to Timing Filter Amplifier(Coarse Gain=X6, Fine GAin=7.5ns && Integrate=20ns) and after amplification to ADC
StripOut is used as a trigger and gate. The signal from the strip is amplified by the Timing Filter Amplifier (Coarse Gain=X20, Fine Gain=12.5ns && Integrate=500ns)
The Threshold on DIFF. CFD is set to 0.70 (ULD=9.9~10 && LLD=0.5)
12-10(11)-08
HV Settings:
Using again only TrigOut and the signal from the Strip.
Electronics Setup: done (might change CFD threshold)
12-11-08
HV Settings:
Gas type 90/10 ArCO2.
Electronics Setup
TrigOut and the stripsignal, both are used as a pulse and go to ADC.
The pulse from the Strips is used as a trigger and gate.
The pulse from the trigout is delayed, Amplified by the Timinig Filter Amplifier(Coarse_Gain=X20, Fine_Gain=8.5 && Integrate=20ns).
Signal from the Strips goes to the Timing Filter Amplifier(Coarse_Gain=X20, Fine_Gain=10.5 && Integrate=20ns). Amplified pulse is devided into two signals, One is sued as a trigger and gate and second as a pulse for ADC. The first one(gate && trigger) is amplified using the Timing Filter Amplifier(Coarse_Gain=X2, Fine_Gain=10 && Integrate=50ns) and used in DIFF. CFD for gate and trigger. Second output from the Amplified strips, goes to delay box(TENNELEC TC 215 DELAY AMPLIFIER) and putted to the ADC.
The results shown below are for different Diff. CFD values.
1). Threshold Values: Threshold=1.90, ULD=9.65 && LLD=2.10
Run Number is r613
Start time: Dec 11 16:31:42 End Time: Dec 11 17:32:23
2). Threshold Values: Threshold=3.10, ULD=9.65 && LLD=2.90
Run Number is r622
Start time: Dec 11 19:25:28 End Time: Dec 11 21:53:44
2). Threshold Values: Threshold=3.70, ULD=9.65 && LLD=3.50
Run Number is r624
Start time: Dec 11 21:58:37 End Time: Dec 12 12:19:53
The result shown below is reasonable. The charge that left the last GEM foil is greater than the charge collected by the strips.
12-12-08
Finding out what is gain for each channel using the Stanford pulse generator.
Setting:
Run Number: r634 && r635
Original pulse from stanford pulse generator goes to the ADC: r636 && r637
I used the attenuator for both channels and it was set to the same value: 12dB = 10.7918.
Gain Calculation
Gain_for_Channel_1 = Gain_TrigOut =
Gain_for_Channel_5 = Gain_Strip =
12-13-08
The Stanford Pulse Generator
Gas type ArCO2(90/10)
New measurements. Setting the gain for Strip pulse and TrigOut the same.
Setting Gain the same using The Stanford Pulse Generator:
Run Number 650.dat
GEM Data
Below Channel 1 represents StripOut and Channel 2 trigOut.
- Run Number r651
Start Time: Dec 13 19:43:14 and End Time: Dec 14 00:21:47
Electronics Settings:
TrigOut.)
Timing Filter Amplifier Model 474-06(Coarse Gain=X10, Fine Gain=5 && Integrate=20ns)
StripOut.)
Timing Filter Amplifier Model 474-09(Coarse Gain=X10, Fine Gain=5 && Integrate=20ns)
Pulse for DIFF. CFD
Timing Filter Amplifier Model 474-02(Coarse Gain=X4, Fine Gain=2 && Integrate=20ns)
DIFF. CFD(Threshold= 3, LLD=2.80 && ULD=9.55)
The Amplification on Strip pulse and TrigOut is the same. Signal is amplified and delayed after amplification.
Charge leaving last GEM foil divided by the charge that was collected by the Strips.
- Run Number r653
Electronics Settings are the same, only integration time is set to 50ns for both signals(Strip and TrigOut)
Start Time: Dec 14 00:39:31 and End Time: Dec 14 16:07:21