Forest IonizationChambers

From New IAC Wiki
Jump to navigation Jump to search

An ionization chamber is essentially a gas filled volume which has one or more wires suspended inside with a voltage bias used to collect any ions or liberated electrons within the volume.

Forest IonizationChamber GenericSingleWireChamber.png

There are 4 basic operating regions which are determined by the amount of the voltage across the wire(s).

  1. recombination
  2. ionization
  3. proportional (plataue)
  4. Geiger-Muller


Recombination

Until the voltage reaches a high enough value (threshold), the liberated electrons produces will recombine with the gas before reaching the wire.

Ionization

Once the voltage is increased beyond the above threshold, the ions/ electrons make it to the cathode/anode and produce an electronic pulse on the wire (cathode). The voltage is high enough to attract all of the electron-ion pairs. The pulse height (signal) out of the chamber is independent of the voltage at this point.

quick definitions

Anode
Current IN, positively charged electrode
Cathode
Current OUT, negatively charged electrode

The flow of electrons is from the cathode to the anode.

Proportional

At a certain voltage though the electron-ions pairs gain enough kinetic energy that they can start to ionization the chamber gas as well. Now the pulse height (signal) out of the chamber depends on the Voltage on the wire.

As the electrons get close to the Anode they see a strong electric field and are accelerated a high enough energy to ionize the chamber gas.

This ionization "amplifies" the original ionization signal strength. The amplification is referred to as the Townsend avalanche and can increase the signal strength by 1000 to 10,000.

At some point(Limited proportionality region) though you start to create enough positive ions that the electric field strength decreases and you get less avalanche even though you are increasing the voltage.


Geiger Muller

At some point the high voltage is high enough that the number of positive ion pairs saturates (you don't create any more even though the voltage is increased). This is called the Geiger Muller plateau. If you go too high in voltage you will start to have electrostatic discharge making the device give a signal without having any ionizing radiation.

Forest IonizationChambers Electron-IonPairFormation.png

GM Plateau.jpg


Go Back