Running CODA
Startup the processes
Start the following programs in the order below and in different xterm windows
1.)msqld 2.)minicom
- cntrl-A P E to setup communication speeds
- reboot to be sure ROC is alive
3.)et_start
- you may need to delete the old memory file /tmp/et_sys_DAQ
4.)rcServer
5.) coda_eb -i -s DAQ -n eb1 -t CDEB
6.)coda_er -i -s DAQ -n LDS_ER -t ROC
7.) runcontrol
Select run configuration
Download configuration to ROC
Prestart
Run
debug FAQ
1.) Using local netowrk card
I decided to connect the ROC directly to the second 1 Gigabit ethernet card on the host in order to avoid setting up a firewall which makes our sys admin guys happy and allows CODA to work.
I had to tell the ROC to alias the new host names in the boot script
so in the file CODA/bootscripts/roc1.boot
I added the line
hostAdd "daq1" "10.1.1.1"
You can do this on the roc as well and then on the ROC you can check the lookup tables using the command
-> hostShow hostname inet address aliases -------- ------------ ------- localhost 127.0.0.1 roc1 10.1.1.2 localdaq 10.1.1.1 daq1 value = 0 = 0x0
2.) Incorrect number Argument when downloading
daLogMsg: Downloading configuration "SIS3610" Incorrect number of Arguments passed for ROL = 1 daLogMsg: Incorrect number of Arguments passed for ROL = 1
I had incorrectly connected my EB to my ER system