CH Weekly Reports
09/19/22
- Investigated non-linearity of peak sensing ADC due to issues arising in energy calibrations and efficiency measurements.
- Modified existing calibration code to use non-linear regression to account for ADC.
- Sourced material needed for CaGaS Phase II setup.
- Calculated ADC dead time for live-time correction to efficiency measurement count rates and calculated dead time per experiment.
- Designed holder for Gd foils needed for CaGaS Phase II experiment. To be 3D printed and tested for tolerances.
09/26/22
- Printed both parts of the Gd foil holder for CaGaS Phase II.
- Parts fit together adequately considering they're 3D printed and not machined.
- M3 bolts threaded through first part of the holder.
- M3 nuts pressed into the second part of the holder.
- Parts printed out of PETG plastic with low density infill (10%).
- Edited rebinning code for energy calibrations for data sets collected with ADC to account for changes in calibration code.
- Energy calibration script now uses 2nd order polynomial regression to account for nonlinearity of ADC
- The coefficients from the regression are passed to another script to rebin ADC data from channel to energy, preserving bin count, but bin width will increase as a function of energy.
- End goal is to pass coefficients into rebinning function and draw energy calibrated plot in the same script.
- Designed Gd2O3 powder holder for CaGaS Phase II.
- Have spray on clear coat to seal plastic after printing is complete in hopes that powder will not stick between printed layers when used.
- Documented Gd holders in wiki with rendered images and a description of intended use.
- Continued to research live time correction for HPGe count rates and think I have applied the correction correctly at the initial measurement level, which will be propagated through to efficiency calculations.
- Researched and tested different integration methods for more accurate measurement of counts in energy peak to compare to Gaussian fit.
- Created tentative list of committee members with major advisor for PhD proposal defense and future dissertation defense.
10/03/22
- Sat in on 30 minute proposal defense presentation by another Ph.D. candidate.
- Began the 3D print of Gd2O3 holder for CaGaS Phase II.
- Both parts printed and fit together tightly. Need to be sealed for use with powder.
- Updated wiki documentation with images of the holders.
- Working with ISU Environmental Health & Safety to set up time to treat and coat lead bricks to build CaGaS Phase II setup.
- Air monitoring and respirators are required per ISU EH&S safety protocols.
- Designed and printed a holder for Cs137 calibration source for storage in lab (signed off with RSO) for easy point calibrations using HPGe.
- Researched dead time adjustment in Practical Gamma-ray Spectrometry by Gordon Gilmore
- Collected 50 data points for ADC gate width, read out controller trigger width, and the delay between end of gate and beginning of read trigger.
- Found dead time per pulse which includes memory storage time, thanks to reading the aforementioned book.
- Worked to re-plot and integrate Eu152 peaks, but source data file seems to be corrupted. Will collect new set of data on Monday (10/10/22) when I can access calibration sources.
- Re-plotting is necessary for new integration with 2nd order polynomial binning and new live time correction.
- Collecting a new Eu152 spectrum is needed before CaGaS Phase II for new energy calibration as NIM bin has been power cycled and spec-amp gain has shifted.
10/10/22
- Moved extra lead for CaGaS Phase II from lead storage at ISU to lab for coating prior to use.
- Coated 30 lead bricks for CaGaS Phase II setup.
- Fixed energy calibration code thanks to input from a colleague.
- Machine precision and how ROOT interprets bins was cause of the issue.
- Designed tools to use to level and compact powder into the powder holder for use during CaGaS Phase II.
- Designed new covers for powder holder to better accommodate for different thicknesses of powder.
- Experimentally confirmed minimum dead and maximum bounds on dead time of data acquisition system per event after questions from colleague and plotted voltage traces for documentation.
- Refit 152Eu energy peaks with gaussians + linear background and recalculated efficiencies using live time correction.
- Finished replotting and calculating uncertainty in HPGe efficiency measurements to present next week, all calibration sources (152Eu, 137Cs, & 60Co).
- Met with collaborators and discussed FY23 work scope.
- Setup recurring meetings biweekly starting 10/19/22.
- Contributed to Paul's SOW for FY23.
10/17/22
- Uploaded new versions of efficiency measurement plots to wiki.
- Wrote up efficiency measurement results, explained in detail how the calculations work.