Calibration Info

From New IAC Wiki
Revision as of 22:57, 15 May 2018 by Harpconn (talk | contribs)
Jump to navigation Jump to search

To fully understand the characteristics of the Nanodot OSLs, a custom calibration was needed. This calibration is used in lieu of the calibration created by the OSL reader using the pre-dosed OSLs from the manufacturer. To begin the calibration, a set of fifteen unexposed Nanodot OSLs, chosen at random, were selected and exposed to a 9.3Ci 137Cs source. Using a distance of 30cm from the faceplate of the source, the exposure rate was calculated in Roentgen using

˙R=ΓAD2 where , ,

There is an adjustment distance of 11.2cm which accounts for the distance from the faceplate to the surface of the source

Total exposures are calculated by





Click here for calibration data