CrossTalk
Go back MCNP simulations
Geometry
An array of 6 detectors are placed radially at a distance of 1 meter from an uncorrelated 252-CF source. The image below shows a top down view of the simulation geometry. The detector setups have a vertical extent of 30".
Detector physics
Summary
Neutron energy deposition from collisions with hydrogen and carbon are converted to electron equivalent light output by MCNP-POLIMI. Neutron collisions that all occur whithin a pulse collection time of 10 ns of each other are summed. If a pulse's total light output exceeds 0.03 MeVee, it is considered a detection. The value 0.03 MeVee was chosen as the detector's light output threshold because it corresponds to a mean neutron energy deposit of 0.2 MeV.
Detectpor responce
The electron equivalent (MeVee) conversion functions for neutrons were measured for plastic (BC 420) scintillators as a function of neutron energy deposit,
. For deposit on hydrogen the measurements fit the following quadratic function:
and for deposit on carbon:
A light output of 0.03 MeVee corresponded to an average neutron energy deposition of 0.2 MeV in the scintillator.
Go back MCNP simulations