SeRun 03-07-16

From New IAC Wiki
Revision as of 04:04, 10 March 2016 by Foretony (talk | contribs)
Jump to navigation Jump to search

Attempted to see Se lines in Horse mineral supplement that contains 16 PPM of Se.

Files on PC stored in

cd:\MPA3(x64)\data\TonyForest\Se_Run_000


calibration at

https://wiki.iac.isu.edu/index.php/SeRun_01-11-16


Calibration for Detector A (Channel 0) on 3-6-16

Run Number Isotope Egamma (keV) Channel
Calib002 Na-22 511 559.4 \pm 1.408
Calib002 Na-22 1274.537 1376 \pm 1.255
Calib003 Eu-152 1408.1 1518 \pm 1.417
Calib003 Eu-152 963.39 1044 \pm 1.198
Calib003 Eu-152 778.9 845.7 \pm 1.107
Calib003 Eu-152 121.78 143.4 \pm 10.6816
Calib004 Mn-54 834.848 905.5 \pm 0.9194
Calib005 Ba-133 80.9971 99.77 \pm 0.6394
Calib005 Ba-133 276.3997 308.7 \pm 0.8055
Calib005 Ba-133 302.8510 337 \pm 0.7719
Calib005 Ba-133 356.0134 393.8 \pm 0.7181
Calib005 Ba-133 383.8480 423.5 \pm 0.7247
Calib006 Cs-137 661.657 720.5 \pm 0.8931

A = -12.3372 , B = 0.935388

MPA->Draw("-12.3372+0.935388*evt.Chan>> SeRun_008(8000,0.5,8000.5)","evt.ADCid==0");

Calibration for Detector D (Channel 3) on 3-6-16

Run Number Isotope Egamma (keV) Channel
Calib002 Mn-54 834.848 869.3 \pm 0.5955
Calib003 Ba-133 80.9971 84.08 \pm 0.6107
Calib003 Ba-133 276.3997 287.7 \pm 0.6316
Calib003 Ba-133 302.8510 315.3 \pm 0.6059
Calib003 Ba-133 356.0134 370.7 \pm 0.606
Calib003 Ba-133 383.8480 399.6 \pm 0.5947
Calib004 Na-22 511 532 \pm 1.292
Calib004 Na-22 1274.537 1327 \pm 0.8383
Calib005 Eu-152 1408.1 1466 \pm 0.9308
Calib005 Eu-152 963.39 1004 \pm 0.8665
Calib005 Eu-152 778.9 810.9 \pm 0.8554
Calib005 Eu-152 121.78 126.5 \pm 0.9308
Calib007 Cs-137 661.657 689 \pm 0.6867

A = 0.18063 , B = 0.960133

MPA->Draw("0.18063+0.960133*evt.Chan>> SeRun_008(8000,0.5,8000.5)","evt.ADCid==3");


Run Acquisition Time(sec) Description
AccOnAlInABackgInD001 ADC1A:14883.26, ADC1D:15020.678 The Aluminum tube is in detector A, nothing in Detector B, the Accelerator is On
AccOnAlInDetASe-AinDetD_001 ADC1A:4881.622, ADC1D:4835.252 The Aluminum tube is in detector A, nothing in Detector B, the Accelerator is On
Se Test run with Horse Mineral Supplement

PAA_Selenium