SeRun 03-07-16
Jump to navigation
Jump to search
Attempted to see Se lines in Horse mineral supplement that contains 16 PPM of Se.
Files on PC stored in
cd:\MPA3(x64)\data\TonyForest\Se_Run_000
calibration at
https://wiki.iac.isu.edu/index.php/SeRun_01-11-16
Calibration for Detector A (Channel 0) on 3-6-16
Run Number | Isotope | Egamma (keV) | Channel |
Calib002 | Na-22 | 511 | 559.4 \pm 1.408 |
Calib002 | Na-22 | 1274.537 | 1376 \pm 1.255 |
Calib003 | Eu-152 | 1408.1 | 1518 \pm 1.417 |
Calib003 | Eu-152 | 963.39 | 1044 \pm 1.198 |
Calib003 | Eu-152 | 778.9 | 845.7 \pm 1.107 |
Calib003 | Eu-152 | 121.78 | 143.4 \pm 10.6816 |
Calib004 | Mn-54 | 834.848 | 905.5 \pm 0.9194 |
Calib005 | Ba-133 | 80.9971 | 99.77 \pm 0.6394 |
Calib005 | Ba-133 | 276.3997 | 308.7 \pm 0.8055 |
Calib005 | Ba-133 | 302.8510 | 337 \pm 0.7719 |
Calib005 | Ba-133 | 356.0134 | 393.8 \pm 0.7181 |
Calib005 | Ba-133 | 383.8480 | 423.5 \pm 0.7247 |
Calib006 | Cs-137 | 661.657 | 720.5 \pm 0.8931 |
A = -12.3372 , B = 0.935388
MPA->Draw("-12.3372+0.935388*evt.Chan>> SeRun_008(8000,0.5,8000.5)","evt.ADCid==0");
Calibration for Detector D (Channel 3) on 3-6-16
Run Number | Isotope | Egamma (keV) | Channel |
Calib002 | Mn-54 | 834.848 | 869.3 \pm 0.5955 |
Calib003 | Ba-133 | 80.9971 | 84.08 \pm 0.6107 |
Calib003 | Ba-133 | 276.3997 | 287.7 \pm 0.6316 |
Calib003 | Ba-133 | 302.8510 | 315.3 \pm 0.6059 |
Calib003 | Ba-133 | 356.0134 | 370.7 \pm 0.606 |
Calib003 | Ba-133 | 383.8480 | 399.6 \pm 0.5947 |
Calib004 | Na-22 | 511 | 532 \pm 1.292 |
Calib004 | Na-22 | 1274.537 | 1327 \pm 0.8383 |
Calib005 | Eu-152 | 1408.1 | 1466 \pm 0.9308 |
Calib007 | Cs-137 | 661.657 | 689 \pm 0.6867 |
A = 0.106296 , B = 0.960308
MPA->Draw("-0.106296+0.960308*evt.Chan>> SeRun_008(8000,0.5,8000.5)","evt.ADCid==3");
Run | Acquisition Time(sec) | Description |
AccOnAlInABackgInD001 | ADC1A:14883.26, ADC1D:15020.678 | The Aluminum tube is in detector A, nothing in Detector B, the Accelerator is On |
AccOnAlInASe-AinDetD_001 | ADC1A:4881.622, ADC1D:4835.252 | The Aluminum tube is in detector A, nothing in Detector B, the Accelerator is On |