LDS HpGe CPAA Apparatus

From New IAC Wiki
Jump to navigation Jump to search

Setup

6/19/14

Looked at the raw output of both HpGe detectors setup in the LDS.

Their names are Zyad and Tareq.

Zyad is a 90+% efficiency HpGe detector.

The HV on both was set to 2200 by slowing increasing the voltage from zero while keeping the detector output DC offset from exceeding 100 mV. Zyad's DC output is very sensitive in the first 500 Volts or so. Ortec's suggested operating voltage for Zyad +3500 Volts and for Tareq +3000 Volts.


Below are 2 scope picture of the signal coming directly out of the detectors

HpGe LDS Tareq Zyad 6-19-2014.pngHpGe LDS Tareq Zyad B6-19-2014.png


The noise level increases on Zyad to 25 +/- 5 mV when the HV is turned off while Tareq's is 12 +/- 5 mV.

HpGe LDS Tareq Zyad C6-19-2014.png


LDS_HpGe_CPAA#Apparatus