R1S5 SL2 BurnIn
Jump to navigation
Jump to search
Date | Time (24h) | Guard (V) | Sense (V) | Field (V) | Sense ( | A)Field ( | A)comments |
05/28/13 | 14:08 | 100 | 100 | -100 | 0.02 | 0.00 | Confirmed with initial setting of 30/40/50V that there are no crossed connections between wire types and that voltage supply is connected correctly. |
05/28/13 | 16:34 | 536 | 1500 | -750 | 45.5 | 45.5 | Someone upped the voltage to full the full operating level without recording it. This is the current reading right now. |
05/28/13 | 17:56 | 536 | 1500 | -750 | 41.5 | 41.6 | |
05/29/13 | 07:56 | 536 | 1500 | -750 | 22.5 | 22.6 | |
05/29/13 | 15:02 | 536 | 1500 | -750 | 17.9 | 18.0 | |
05/29/13 | 18:07 | 536 | 1500 | -750 | 16.7 | 16.8 | |
05/30/13 | 07:55 | 536 | 1500 | -750 | 10.6 | 10.7 | |
05/30/13 | 12:51 | 536 | 1500 | -750 | 9.5 | 9.6 | |
05/30/13 | 17:23 | 536 | 1500 | -750 | 8.7 | 8.8 | |
06/03/13 | 08:23 | 536 | 1500 | -750 | 1.7 | 1.8 | |
06/04/13 | 07:42 | 536 | 1500 | -750 | 1.4 | 1.5 |