General info on IAC NaI detector
SCIONIX HOLLAND
This is a Crystal with base.
Ruggedized Detectors:
http://www.helgesonsa.es/m02/firmas/SCIONIX/pages/pdf/Ruggedized%20Detector.pdf
LaCl 3:Ce Scintillation Crystals:
http://www.helgesonsa.es/m02/firmas/SCIONIX/pages/pdf/LaCl3%20literature.pdf
File:SCIONIX HOLLAND Crystal info.pdf
Rexon Components Inc
Voltage range: 0 - 2000 V DC (positive)
B76D01 PMT Pinouts
This is a Crystal, and there is a base from same company.
Rexon PB200 Base
Base: http://www.rexon.com/POWER%20SUPPLY2.htm
File:Rexon Base RB200datasheet.pdf
File:Rexon Base RB-200 PMT Base Manual.pdf
PMT:
size: 3" diameter
Saint-Gobain crystals & detectors ltd
This is just a base.
PA-14
Standard Features of Model P-14 Plug-on Voltage Divider
• Aluminum housing
• Positive (+) high voltage operation
• BNC bulkhead connector for signal output
• SHV bulkhead connector for high-voltage supply input
• 800 to 1200 VDC (may be higher or lower depending upon the specific PMT) at <1 mA typical operating power requirements
• Gain potentiometer
• Focus potentiometer
Additional features for model PA-14 plug-on voltage divider/preamplifier
• Standard hard-wired preamplifier power cable terminating in a 9-pin connector (pin 6 for -24VDC, pin 1 common ground)
• Optional BNC (in place of standard 9-pin) connector for preamplifier power input; -24V at 21mA operating power requirements.
Preamplifier specifications
• Noise: Less than 50 microvolts RMS referred to input
• Input: Directly coupled via 1μF capacitor from photomultiplier tube anode to charge integrating preamplifier input circuit, positive pulse output.
• Conversion Gain: 100mV/picocoulomb
• Rise Time: 100 nanoseconds
• Decay Time: 10 microseconds (internally adjustable at the factory)
• Gain Stability: 0.25%/C from 0° to 50°C
• Linearity: 0.25% (preamplifier only)
• Power Requirements: -24V +10% @21mA
Canberra Industries Inc.
This is just a base.
It is not clear from the PMT base we have what model is it, the data got scratched off. The comparison to information given on their website, it seems that the model is: 2007P.
http://www.canberra.com/pdf/Products/Model-2007-2007P-SS-CSP0071.pdf
Ortec Base Images
Crystal size
NaI crystal sizes: 3”x3”
PMT base configureation
IAC NAI_1
analysis commands
/home/daq/CODA/CODAreader/UsingROOTnEVIO/Oleksiy/evio2nt -fr2177.dat > /dev/null
mv r2177 r2177.root
root r2177.root
TH1F * Hist1=new TH1F("Hist1","Hist1",4096,-0.5,4095.5);
ntuple->Draw("ADC1 >> Hist1")