MiniDC PreampOutput 10 10 07
Revision as of 00:37, 16 October 2007 by Oborn (talk | contribs) (→Scope pictures triggering on preamp output (pin 2))
Drift chamber High voltage settings
HV settings | |
Wire | Volts |
Sense | 1800 |
Field | -800 |
Guard | 800 |
I should have set the Field and Guard wires to 900 volts according to Mac's e-mail reporting his Garfield simulation results. Next time.
Scope pictures triggering on sense wire C2
I set the scope threshold to below -14 mV based on previous measurements.
Scope pictures triggering on preamp output (pin 2)
I found that a trigger threshold of -100 mV discriminated well between noise and pulses consistent with previous ionization events.