PAS Feasibility Preliminary Runs
This experimental setup does not contain BorAl
General run plan
- Run source in and source out in experimental setup and background subtract with NaI
- Move the source to different locations along the source tunnel and plot the different distances and count rates
- Run source in and source out in experimental setup and background subtract with He3
- Move the source to different locations along the source tunnel and plot the different distances and count rates
I will upload some models of the setup when I have updated the model to include the source and detector
7/19/21
NaI detector 30cm from converter
NaI Detector, 5mV threshold, +1000V on detector, no converter
Source Distance (cm) | Counts | Run Time (s) | Count Rate (Hz) | Uncertainty | |
---|---|---|---|---|---|
39.5 | 1563 | 100 | |||
35 | 1631 | 100 | |||
30 | 1791 | 100 | |||
25 | 2054 | 100 | |||
20 | 2514 | 100 | |||
15 | 3375 | 100 | |||
10 | 6182 | 100 |
NaI Detector, 5mV threshold, +1000V on detector, with converter
Source Distance (cm) | Counts | Run Time (s) | Count Rate (Hz) | Uncertainty | |
---|---|---|---|---|---|
39.5 | 1489 | 100 | |||
35 | 1581 | 100 | |||
30 | 1729 | 100 | |||
25 | 1936 | 100 | |||
20 | 2401 | 100 | |||
15 | 3249 | 100 | |||
10 | 5969 | 100 |
NaI Detector, 2mV threshold, +1000V on detector, no converter
Source Distance (cm) | Counts | Run Time (s) | Count Rate (Hz) | Uncertainty | |
---|---|---|---|---|---|
39.5 | 2039 | 100 | |||
35 | 2228 | 100 | |||
30 | 2546 | 100 | |||
25 | 2960 | 100 | |||
20 | 3939 | 100 | |||
15 | 5349 | 100 | |||
10 | 10050 | 100 |
NaI Detector, 2mV threshold, +1000V on detector, with converter
Source Distance (cm) | Counts | Run Time (s) | Count Rate (Hz) | Uncertainty | |
---|---|---|---|---|---|
39.5 | 2071 | 100 | |||
35 | 2225 | 100 | |||
30 | 2430 | 100 | |||
25 | 2825 | 100 | |||
20 | 3391 | 100 | |||
15 | 4861 | 100 | |||
10 | 9576 | 100 |
7/20/21
detector is 5cm from converter (5cm from end of detector tunnel towards source)
He3 detector, 1250V, no converter, 250mV threshold
Source Distance (cm) | Counts | Run Time (s) | Count Rate (Hz) | Uncertainty | |
---|---|---|---|---|---|
39.5 | 478 | 100 | |||
35 | 676 | 100 | |||
30 | 1010 | 100 | |||
25 | 1604 | 100 | |||
20 | 2533 | 100 | |||
15 | 4017 | 100 | |||
10 | 6035 | 100 |
He3 detector, 1250V, with converter, 250mV threshold
Source Distance (cm) | Counts | Run Time (s) | Count Rate (Hz) | Uncertainty | |
---|---|---|---|---|---|
39.5 | 146 | 200 | |||
35 | 140 | 200 | |||
30 | 228 | 200 | |||
25 | 186 | 100 | |||
20 | 278 | 100 | |||
15 | 546 | 100 | |||
10 | 729 | 100 |
7/21/21
Source at 10cm inside source tunnel
Running with NaI and ADC setup.
ADC is not calibrated. Running converter in and converter out.
Will swap to He3 detector following these and overlay the source in source out plots from the root files.
I will put all this into a table later
- NaI with converter: Run 7031
- run start: 13:01:51
- run stop: 13:12:07
- NaI without converter: Run 7033
- Run 7032 failed 7 minutes in and was stopped by the computer
- run start: 13:17:30
- run stop: 13:27:27 or 13:37:27
He3 Runs
- He3 no converter: Run 7034
- run start: 15:46:25
- run stop: 15:57:38
- He3 with converter: Run 7035
- run start: 16:01:34
- run stop: 16:11:28
7/22/21
2mV Threshold NaI | 5mV Threshold NaI | He3 Tube |
---|---|---|
Adjusting discriminator threshold for NaI with/without converter