Timing filter Amp Configuration
Course gain: x6
Fine gain: at minimum
Integrate (nS): 500nS
Diff (nS): Out
Invert/Noninverted: Non-Inv
Dual Timer Configuration
Setup such that input discriminated pulse is at width ~200ns
Out feeds into scalar for counts
Top half of dual timer used to gate signal for scalar in plateau measurements
Plateau Data
700V
700V Source Out Data |
|
Run Number |
Configuration |
Discriminator Threshold (mV) |
Cs-137 Button Source |
Counts |
Run Time (Sec) |
Count Rate (Hz)
|
1 |
Unshielded detector at +700V |
05 |
No |
958 |
60 |
15.97
|
2 |
Unshielded detector at +700V |
15 |
No |
278 |
60 |
4.63
|
3 |
Unshielded detector at +700V |
35 |
No |
107 |
60 |
1.78
|
4 |
Unshielded detector at +700V |
55 |
No |
85 |
60 |
1.42
|
5 |
Unshielded detector at +700V |
75 |
No |
80 |
60 |
1.33
|
6 |
Unshielded detector at +700V |
95 |
No |
80 |
60 |
1.33
|
7 |
Unshielded detector at +700V |
115 |
No |
75 |
60 |
1.25
|
8 |
Unshielded detector at +700V |
135 |
No |
78 |
60 |
1.3
|
9 |
Unshielded detector at +700V |
155 |
No |
82 |
60 |
1.37
|
10 |
Unshielded detector at +700V |
175 |
No |
77 |
60 |
1.28
|
11 |
Unshielded detector at +700V |
195 |
No |
66 |
60 |
1.1
|
12 |
Unshielded detector at +700V |
215 |
No |
68 |
60 |
1.13
|
13 |
Unshielded detector at +700V |
235 |
No |
53 |
60 |
0.88
|
14 |
Unshielded detector at +700V |
255 |
No |
59 |
60 |
0.98
|
700V Source In Data |
|
Run Number |
Configuration |
Discriminator Threshold (mV) |
Cs-137 Button Source |
Counts |
Run Time (Sec) |
Count Rate (Hz)
|
1 |
Unshielded detector at +700V |
05 |
Yes |
1033 |
60 |
|
2 |
Unshielded detector at +700V |
15 |
Yes |
291 |
60 |
|
3 |
Unshielded detector at +700V |
35 |
Yes |
102 |
60 |
|
4 |
Unshielded detector at +700V |
55 |
Yes |
87 |
60 |
|
5 |
Unshielded detector at +700V |
75 |
Yes |
88 |
60 |
|
6 |
Unshielded detector at +700V |
95 |
Yes |
103 |
60 |
|
7 |
Unshielded detector at +700V |
115 |
Yes |
101 |
60 |
|
8 |
Unshielded detector at +700V |
135 |
Yes |
84 |
60 |
|
9 |
Unshielded detector at +700V |
155 |
Yes |
72 |
60 |
|
10 |
Unshielded detector at +700V |
175 |
Yes |
57 |
60 |
|
11 |
Unshielded detector at +700V |
195 |
Yes |
72 |
60 |
|
12 |
Unshielded detector at +700V |
215 |
Yes |
68 |
60 |
|
13 |
Unshielded detector at +700V |
235 |
Yes |
74 |
60 |
|
14 |
Unshielded detector at +700V |
255 |
Yes |
52 |
60 |
|
700V Plots |
|
Plateau measurements at 700V, source in and source out overlaid. |
Plateau measurements at 700V, source in minus source out.
|
750V
750V Source Out Data |
|
Run Number |
Configuration |
Discriminator Threshold (mV) |
Cs-137 Button Source |
Counts |
Run Time (Sec) |
Count Rate (Hz)
|
1 |
Unshielded detector at +750V |
05 |
No |
2211 |
60 |
36.85
|
2 |
Unshielded detector at +750V |
15 |
No |
1125 |
60 |
18.75
|
3 |
Unshielded detector at +750V |
35 |
No |
293 |
60 |
4.88
|
4 |
Unshielded detector at +750V |
55 |
No |
105 |
60 |
1.75
|
5 |
Unshielded detector at +750V |
75 |
No |
80 |
60 |
1.33
|
6 |
Unshielded detector at +750V |
95 |
No |
97 |
60 |
1.62
|
7 |
Unshielded detector at +750V |
115 |
No |
99 |
60 |
1.65
|
8 |
Unshielded detector at +750V |
135 |
No |
89 |
60 |
1.48
|
9 |
Unshielded detector at +750V |
155 |
No |
74 |
60 |
1.23
|
10 |
Unshielded detector at +750V |
175 |
No |
80 |
60 |
1.33
|
11 |
Unshielded detector at +750V |
195 |
No |
80 |
60 |
1.33
|
12 |
Unshielded detector at +750V |
215 |
No |
64 |
60 |
1.07
|
13 |
Unshielded detector at +750V |
235 |
No |
89 |
60 |
1.48
|
14 |
Unshielded detector at +750V |
255 |
No |
69 |
60 |
1.15
|
750V Source In Data |
|
Run Number |
Configuration |
Discriminator Threshold (mV) |
Cs-137 Button Source |
Counts |
Run Time (Sec) |
Count Rate (Hz)
|
1 |
Unshielded detector at +750V |
05 |
Yes |
4055 |
60 |
|
2 |
Unshielded detector at +750V |
15 |
Yes |
1160 |
60 |
|
3 |
Unshielded detector at +750V |
35 |
Yes |
301 |
60 |
|
4 |
Unshielded detector at +750V |
55 |
Yes |
102 |
60 |
|
5 |
Unshielded detector at +750V |
75 |
Yes |
82 |
60 |
|
6 |
Unshielded detector at +750V |
95 |
Yes |
114 |
60 |
|
7 |
Unshielded detector at +750V |
115 |
Yes |
85 |
60 |
|
8 |
Unshielded detector at +750V |
135 |
Yes |
93 |
60 |
|
9 |
Unshielded detector at +750V |
155 |
Yes |
87 |
60 |
|
10 |
Unshielded detector at +750V |
175 |
Yes |
87 |
60 |
|
11 |
Unshielded detector at +750V |
195 |
Yes |
79 |
60 |
|
12 |
Unshielded detector at +750V |
215 |
Yes |
90 |
60 |
|
13 |
Unshielded detector at +750V |
235 |
Yes |
84 |
60 |
|
14 |
Unshielded detector at +750V |
255 |
Yes |
73 |
60 |
|
750V Plots |
|
Plateau measurements at 750V, source in and source out overlaid. |
Plateau measurements at 750V, source in minus source out.
|
800V
800V Source Out Data |
|
Run Number |
Configuration |
Discriminator Threshold (mV) |
Cs-137 Button Source |
Counts |
Run Time (Sec) |
Count Rate (Hz)
|
1 |
Unshielded detector at +800V |
05 |
No |
3866 |
60 |
64.43
|
2 |
Unshielded detector at +800V |
15 |
No |
2356 |
60 |
39.27
|
3 |
Unshielded detector at +800V |
35 |
No |
1091 |
60 |
18.18
|
4 |
Unshielded detector at +800V |
55 |
No |
384 |
60 |
6.4
|
5 |
Unshielded detector at +800V |
75 |
No |
244 |
60 |
4.07
|
6 |
Unshielded detector at +800V |
95 |
No |
167 |
60 |
2.78
|
7 |
Unshielded detector at +800V |
115 |
No |
100 |
60 |
1.67
|
8 |
Unshielded detector at +800V |
135 |
No |
88 |
60 |
1.47
|
9 |
Unshielded detector at +800V |
155 |
No |
89 |
60 |
1.48
|
10 |
Unshielded detector at +800V |
175 |
No |
86 |
60 |
1.43
|
11 |
Unshielded detector at +800V |
195 |
No |
90 |
60 |
1.5
|
12 |
Unshielded detector at +800V |
215 |
No |
81 |
60 |
1.35
|
13 |
Unshielded detector at +800V |
235 |
No |
95 |
60 |
1.58
|
14 |
Unshielded detector at +800V |
255 |
No |
89 |
60 |
1.48
|
800V Source In Data |
|
Run Number |
Configuration |
Discriminator Threshold (mV) |
Cs-137 Button Source |
Counts |
Run Time (Sec) |
Count Rate (Hz)
|
1 |
Unshielded detector at +800V |
05 |
Yes |
10055 |
60 |
|
2 |
Unshielded detector at +800V |
15 |
Yes |
5438 |
60 |
|
3 |
Unshielded detector at +800V |
35 |
Yes |
1078 |
60 |
|
4 |
Unshielded detector at +800V |
55 |
Yes |
375 |
60 |
|
5 |
Unshielded detector at +800V |
75 |
Yes |
217 |
60 |
|
6 |
Unshielded detector at +800V |
95 |
Yes |
126 |
60 |
|
7 |
Unshielded detector at +800V |
115 |
Yes |
100 |
60 |
|
8 |
Unshielded detector at +800V |
135 |
Yes |
93 |
60 |
|
9 |
Unshielded detector at +800V |
155 |
Yes |
99 |
60 |
|
10 |
Unshielded detector at +800V |
175 |
Yes |
96 |
60 |
|
11 |
Unshielded detector at +800V |
195 |
Yes |
102 |
60 |
|
12 |
Unshielded detector at +800V |
215 |
Yes |
99 |
60 |
|
13 |
Unshielded detector at +800V |
235 |
Yes |
62 |
60 |
|
14 |
Unshielded detector at +800V |
255 |
Yes |
85 |
60 |
|
800V Plots |
|
Plateau measurements at 800V, source in and source out overlaid. |
Plateau measurements at 800V, source in minus source out.
|
850V
850V Source Out Data |
|
Run Number |
Configuration |
Discriminator Threshold (mV) |
Cs-137 Button Source |
Counts |
Run Time (Sec) |
Count Rate (Hz)
|
1 |
Unshielded detector at +850V |
05 |
No |
6402 |
60 |
106.7
|
2 |
Unshielded detector at +850V |
15 |
No |
3867 |
60 |
64.45
|
3 |
Unshielded detector at +850V |
35 |
No |
2058 |
60 |
34.3
|
4 |
Unshielded detector at +850V |
55 |
No |
1233 |
60 |
20.55
|
5 |
Unshielded detector at +850V |
75 |
No |
787 |
60 |
13.12
|
6 |
Unshielded detector at +850V |
95 |
No |
391 |
60 |
6.52
|
7 |
Unshielded detector at +850V |
115 |
No |
256 |
60 |
4.27
|
8 |
Unshielded detector at +850V |
135 |
No |
193 |
60 |
3.22
|
9 |
Unshielded detector at +850V |
155 |
No |
151 |
60 |
2.52
|
10 |
Unshielded detector at +850V |
175 |
No |
97 |
60 |
1.62
|
11 |
Unshielded detector at +850V |
195 |
No |
105 |
60 |
1.75
|
12 |
Unshielded detector at +850V |
215 |
No |
97 |
60 |
1.62
|
13 |
Unshielded detector at +850V |
235 |
No |
87 |
60 |
1.45
|
14 |
Unshielded detector at +850V |
255 |
No |
101 |
60 |
1.68
|
850V Source In Data |
|
Run Number |
Configuration |
Discriminator Threshold (mV) |
Cs-137 Button Source |
Counts |
Run Time (Sec) |
Count Rate (Hz)
|
1 |
Unshielded detector at +850V |
05 |
Yes |
15370 |
60 |
|
2 |
Unshielded detector at +850V |
15 |
Yes |
10298 |
60 |
|
3 |
Unshielded detector at +850V |
35 |
Yes |
2963 |
60 |
|
4 |
Unshielded detector at +850V |
55 |
Yes |
1366 |
60 |
|
5 |
Unshielded detector at +850V |
75 |
Yes |
842 |
60 |
|
6 |
Unshielded detector at +850V |
95 |
Yes |
402 |
60 |
|
7 |
Unshielded detector at +850V |
115 |
Yes |
281 |
60 |
|
8 |
Unshielded detector at +850V |
135 |
Yes |
196 |
60 |
|
9 |
Unshielded detector at +850V |
155 |
Yes |
143 |
60 |
|
10 |
Unshielded detector at +850V |
175 |
Yes |
98 |
60 |
|
11 |
Unshielded detector at +850V |
195 |
Yes |
88 |
60 |
|
12 |
Unshielded detector at +850V |
215 |
Yes |
101 |
60 |
|
13 |
Unshielded detector at +850V |
235 |
Yes |
104 |
60 |
|
14 |
Unshielded detector at +850V |
255 |
Yes |
101 |
60 |
|
850V Plots |
|
Plateau measurements at 850V, source in and source out overlaid. |
Plateau measurements at 850V, source in minus source out.
|
1000V
1000V Source Out Data |
|
Run Number |
Configuration |
Discriminator Threshold (mV) |
Cs-137 Button Source |
Counts |
Run Time (Sec) |
Count Rate (Hz)
|
1 |
Unshielded detector at +1000V |
05 |
No |
21167 |
60 |
|
2 |
Unshielded detector at +1000V |
10 |
No |
18739 |
60 |
|
3 |
Unshielded detector at +1000V |
15 |
No |
16091 |
60 |
|
4 |
Unshielded detector at +1000V |
20 |
No |
13773 |
60 |
|
5 |
Unshielded detector at +1000V |
25 |
No |
11983 |
60 |
|
6 |
Unshielded detector at +1000V |
30 |
No |
10358 |
60 |
|
7 |
Unshielded detector at +1000V |
35 |
No |
9274 |
60 |
|
8 |
Unshielded detector at +1000V |
45 |
No |
7423 |
60 |
|
9 |
Unshielded detector at +1000V |
55 |
No |
6067 |
60 |
|
10 |
Unshielded detector at +1000V |
65 |
No |
5258 |
60 |
|
11 |
Unshielded detector at +1000V |
75 |
No |
4451 |
60 |
|
12 |
Unshielded detector at +1000V |
80 |
No |
4321 |
60 |
|
13 |
Unshielded detector at +1000V |
85 |
No |
4111 |
60 |
|
14 |
Unshielded detector at +1000V |
90 |
No |
3823 |
60 |
|
15 |
Unshielded detector at +1000V |
95 |
No |
3666 |
60 |
|
16 |
Unshielded detector at +1000V |
100 |
No |
3466 |
60 |
|
17 |
Unshielded detector at +1000V |
105 |
No |
3241 |
60 |
|
18 |
Unshielded detector at +1000V |
110 |
No |
3180 |
60 |
|
19 |
Unshielded detector at +1000V |
115 |
No |
2968 |
60 |
|
20 |
Unshielded detector at +1000V |
120 |
No |
2812 |
60 |
|
21 |
Unshielded detector at +1000V |
125 |
No |
2769 |
60 |
|
22 |
Unshielded detector at +1000V |
130 |
No |
2548 |
60 |
|
23 |
Unshielded detector at +1000V |
135 |
No |
2426 |
60 |
|
24 |
Unshielded detector at +1000V |
155 |
No |
2065 |
60 |
|
25 |
Unshielded detector at +1000V |
175 |
No |
1843 |
60 |
|
26 |
Unshielded detector at +1000V |
195 |
No |
1582 |
60 |
|
27 |
Unshielded detector at +1000V |
215 |
No |
1374 |
60 |
|
28 |
Unshielded detector at +1000V |
235 |
No |
1172 |
60 |
|
29 |
Unshielded detector at +1000V |
255 |
No |
1012 |
60 |
|
1000V Source In Data |
|
Run Number |
Configuration |
Discriminator Threshold (mV) |
Cs-137 Button Source |
Counts |
Run Time (Sec) |
Count Rate (Hz)
|
1 |
Unshielded detector at +1000V |
05 |
Yes |
34443 |
60 |
|
2 |
Unshielded detector at +1000V |
10 |
Yes |
30427 |
60 |
|
3 |
Unshielded detector at +1000V |
15 |
Yes |
26882 |
60 |
|
4 |
Unshielded detector at +1000V |
20 |
Yes |
23779 |
60 |
|
5 |
Unshielded detector at +1000V |
25 |
Yes |
21672 |
60 |
|
6 |
Unshielded detector at +1000V |
30 |
Yes |
19611 |
60 |
|
7 |
Unshielded detector at +1000V |
35 |
Yes |
18115 |
60 |
|
8 |
Unshielded detector at +1000V |
45 |
Yes |
15119 |
60 |
|
9 |
Unshielded detector at +1000V |
55 |
Yes |
12824 |
60 |
|
10 |
Unshielded detector at +1000V |
65 |
Yes |
11575 |
60 |
|
11 |
Unshielded detector at +1000V |
75 |
Yes |
10320 |
60 |
|
12 |
Unshielded detector at +1000V |
80 |
Yes |
9658 |
60 |
|
13 |
Unshielded detector at +1000V |
85 |
Yes |
9328 |
60 |
|
14 |
Unshielded detector at +1000V |
90 |
Yes |
8885 |
60 |
|
15 |
Unshielded detector at +1000V |
95 |
Yes |
8328 |
60 |
|
16 |
Unshielded detector at +1000V |
100 |
Yes |
7906 |
60 |
|
17 |
Unshielded detector at +1000V |
105 |
Yes |
7777 |
60 |
|
18 |
Unshielded detector at +1000V |
110 |
Yes |
7478 |
60 |
|
19 |
Unshielded detector at +1000V |
115 |
Yes |
7313 |
60 |
|
20 |
Unshielded detector at +1000V |
120 |
Yes |
7206 |
60 |
|
21 |
Unshielded detector at +1000V |
125 |
Yes |
6947 |
60 |
|
22 |
Unshielded detector at +1000V |
130 |
Yes |
6936 |
60 |
|
23 |
Unshielded detector at +1000V |
135 |
Yes |
6692 |
60 |
|
24 |
Unshielded detector at +1000V |
155 |
Yes |
2242 |
60 |
|
25 |
Unshielded detector at +1000V |
175 |
Yes |
1878 |
60 |
|
26 |
Unshielded detector at +1000V |
195 |
Yes |
1596 |
60 |
|
27 |
Unshielded detector at +1000V |
215 |
Yes |
1378 |
60 |
|
28 |
Unshielded detector at +1000V |
235 |
Yes |
1280 |
60 |
|
29 |
Unshielded detector at +1000V |
255 |
Yes |
1092 |
60 |
|
1000V Plots |
|
Plateau measurements at 1000V, source in and source out overlaid. |
Plateau measurements at 1000V, source in minus source out.
|
1050V
Operating voltage when detectors used at IAC
1050V Source Out Data |
|
Run Number |
Configuration |
Discriminator Threshold (mV) |
Cs-137 Button Source |
Counts |
Run Time (Sec) |
Count Rate (Hz)
|
1 |
Unshielded detector at +1050V |
05 |
No |
22001 |
60 |
|
2 |
Unshielded detector at +1050V |
10 |
No |
20338 |
60 |
|
3 |
Unshielded detector at +1050V |
15 |
No |
18679 |
60 |
|
4 |
Unshielded detector at +1050V |
20 |
No |
16810 |
60 |
|
5 |
Unshielded detector at +1050V |
25 |
No |
15498 |
60 |
|
6 |
Unshielded detector at +1050V |
30 |
No |
14096 |
60 |
|
7 |
Unshielded detector at +1050V |
35 |
No |
13005 |
60 |
|
8 |
Unshielded detector at +1050V |
45 |
No |
10948 |
60 |
|
9 |
Unshielded detector at +1050V |
55 |
No |
9276 |
60 |
|
10 |
Unshielded detector at +1050V |
65 |
No |
8022 |
60 |
|
11 |
Unshielded detector at +1050V |
75 |
No |
7012 |
60 |
|
12 |
Unshielded detector at +1050V |
80 |
No |
6618 |
60 |
|
13 |
Unshielded detector at +1050V |
85 |
No |
6169 |
60 |
|
14 |
Unshielded detector at +1050V |
90 |
No |
5942 |
60 |
|
15 |
Unshielded detector at +1050V |
95 |
No |
5559 |
60 |
|
16 |
Unshielded detector at +1050V |
100 |
No |
5272 |
60 |
|
17 |
Unshielded detector at +1050V |
105 |
No |
4871 |
60 |
|
18 |
Unshielded detector at +1050V |
110 |
No |
4783 |
60 |
|
19 |
Unshielded detector at +1050V |
115 |
No |
4451 |
60 |
|
20 |
Unshielded detector at +1050V |
120 |
No |
4287 |
60 |
|
21 |
Unshielded detector at +1050V |
125 |
No |
4168 |
60 |
|
22 |
Unshielded detector at +1050V |
130 |
No |
4020 |
60 |
|
23 |
Unshielded detector at +1050V |
135 |
No |
3860 |
60 |
|
24 |
Unshielded detector at +1050V |
155 |
No |
3405 |
60 |
|
25 |
Unshielded detector at +1050V |
175 |
No |
3029 |
60 |
|
26 |
Unshielded detector at +1050V |
195 |
No |
2634 |
60 |
|
27 |
Unshielded detector at +1050V |
215 |
No |
2298 |
60 |
|
28 |
Unshielded detector at +1050V |
235 |
No |
2153 |
60 |
|
29 |
Unshielded detector at +1050V |
255 |
No |
2002 |
60 |
|
1050V Source In Data |
|
Run Number |
Configuration |
Discriminator Threshold (mV) |
Cs-137 Button Source |
Counts |
Run Time (Sec) |
Count Rate (Hz)
|
1 |
Unshielded detector at +1050V |
05 |
Yes |
34250 |
60 |
|
2 |
Unshielded detector at +1050V |
10 |
Yes |
31439 |
60 |
|
3 |
Unshielded detector at +1050V |
15 |
Yes |
29093 |
60 |
|
4 |
Unshielded detector at +1050V |
20 |
Yes |
26912 |
60 |
|
5 |
Unshielded detector at +1050V |
25 |
Yes |
25185 |
60 |
|
6 |
Unshielded detector at +1050V |
30 |
Yes |
23568 |
60 |
|
7 |
Unshielded detector at +1050V |
35 |
Yes |
21921 |
60 |
|
8 |
Unshielded detector at +1050V |
45 |
Yes |
19174 |
60 |
|
9 |
Unshielded detector at +1050V |
55 |
Yes |
17314 |
60 |
|
10 |
Unshielded detector at +1050V |
65 |
Yes |
15339 |
60 |
|
11 |
Unshielded detector at +1050V |
75 |
Yes |
13939 |
60 |
|
12 |
Unshielded detector at +1050V |
80 |
Yes |
13280 |
60 |
|
13 |
Unshielded detector at +1050V |
85 |
Yes |
12649 |
60 |
|
14 |
Unshielded detector at +1050V |
90 |
Yes |
12203 |
60 |
|
15 |
Unshielded detector at +1050V |
95 |
Yes |
11962 |
60 |
|
16 |
Unshielded detector at +1050V |
100 |
Yes |
11217 |
60 |
|
17 |
Unshielded detector at +1050V |
105 |
Yes |
10733 |
60 |
|
18 |
Unshielded detector at +1050V |
110 |
Yes |
10347 |
60 |
|
19 |
Unshielded detector at +1050V |
115 |
Yes |
9953 |
60 |
|
20 |
Unshielded detector at +1050V |
120 |
Yes |
9691 |
60 |
|
21 |
Unshielded detector at +1050V |
125 |
Yes |
9551 |
60 |
|
22 |
Unshielded detector at +1050V |
130 |
Yes |
9147 |
60 |
|
23 |
Unshielded detector at +1050V |
135 |
Yes |
8840 |
60 |
|
24 |
Unshielded detector at +1050V |
155 |
Yes |
7735 |
60 |
|
25 |
Unshielded detector at +1050V |
175 |
Yes |
6970 |
60 |
|
26 |
Unshielded detector at +1050V |
195 |
Yes |
6587 |
60 |
|
27 |
Unshielded detector at +1050V |
215 |
Yes |
5559 |
60 |
|
28 |
Unshielded detector at +1050V |
235 |
Yes |
2341 |
60 |
|
29 |
Unshielded detector at +1050V |
255 |
Yes |
1966 |
60 |
|
1050V Plots |
|
Plateau measurements at 1050V, source in and source out overlaid. |
Plateau measurements at 1050V, source in minus source out.
|
1100V
1100V Source Out Data |
|
Run Number |
Configuration |
Discriminator Threshold (mV) |
Cs-137 Button Source |
Counts |
Run Time (Sec) |
Count Rate (Hz)
|
1 |
Unshielded detector at +1100V |
05 |
No |
|
60 |
|
2 |
Unshielded detector at +1100V |
10 |
No |
|
60 |
|
3 |
Unshielded detector at +1100V |
15 |
No |
|
60 |
|
4 |
Unshielded detector at +1100V |
20 |
No |
|
60 |
|
5 |
Unshielded detector at +1100V |
25 |
No |
|
60 |
|
6 |
Unshielded detector at +1100V |
30 |
No |
|
60 |
|
7 |
Unshielded detector at +1100V |
35 |
No |
|
60 |
|
8 |
Unshielded detector at +1100V |
45 |
No |
|
60 |
|
9 |
Unshielded detector at +1100V |
55 |
No |
|
60 |
|
10 |
Unshielded detector at +1100V |
75 |
No |
|
60 |
|
11 |
Unshielded detector at +1100V |
95 |
No |
|
60 |
|
12 |
Unshielded detector at +1100V |
115 |
No |
|
60 |
|
13 |
Unshielded detector at +1100V |
135 |
No |
|
60 |
|
14 |
Unshielded detector at +1100V |
155 |
No |
|
60 |
|
15 |
Unshielded detector at +1100V |
175 |
No |
|
60 |
|
16 |
Unshielded detector at +1100V |
195 |
No |
|
60 |
|
17 |
Unshielded detector at +1100V |
215 |
No |
|
60 |
|
18 |
Unshielded detector at +1100V |
235 |
No |
|
60 |
|
19 |
Unshielded detector at +1100V |
255 |
No |
|
60 |
|
1100V Source In Data |
|
Run Number |
Configuration |
Discriminator Threshold (mV) |
Cs-137 Button Source |
Counts |
Run Time (Sec) |
Count Rate (Hz)
|
1 |
Unshielded detector at +1100V |
05 |
Yes |
36350 |
60 |
|
2 |
Unshielded detector at +1100V |
10 |
Yes |
32632 |
60 |
|
3 |
Unshielded detector at +1100V |
15 |
Yes |
30656 |
60 |
|
4 |
Unshielded detector at +1100V |
20 |
Yes |
29153 |
60 |
|
5 |
Unshielded detector at +1100V |
25 |
Yes |
|
60 |
|
6 |
Unshielded detector at +1100V |
30 |
Yes |
|
60 |
|
7 |
Unshielded detector at +1100V |
35 |
Yes |
|
60 |
|
8 |
Unshielded detector at +1100V |
45 |
Yes |
|
60 |
|
9 |
Unshielded detector at +1100V |
55 |
Yes |
|
60 |
|
10 |
Unshielded detector at +1100V |
75 |
Yes |
|
60 |
|
11 |
Unshielded detector at +1100V |
95 |
Yes |
|
60 |
|
12 |
Unshielded detector at +1100V |
115 |
Yes |
|
60 |
|
13 |
Unshielded detector at +1100V |
135 |
Yes |
|
60 |
|
14 |
Unshielded detector at +1100V |
155 |
Yes |
|
60 |
|
15 |
Unshielded detector at +1100V |
175 |
Yes |
|
60 |
|
16 |
Unshielded detector at +1100V |
195 |
Yes |
|
60 |
|
17 |
Unshielded detector at +1100V |
215 |
Yes |
|
60 |
|
18 |
Unshielded detector at +1100V |
235 |
Yes |
|
60 |
|
19 |
Unshielded detector at +1100V |
255 |
Yes |
|
60 |
|
1100V Plots |
|
Plateau measurements at 1100V, source in and source out overlaid. |
Plateau measurements at 1100V, source in minus source out.
|
1200V
1200V Source Out Data |
|
Run Number |
Configuration |
Discriminator Threshold (mV) |
Cs-137 Button Source |
Counts |
Run Time (Sec) |
Count Rate (Hz)
|
1 |
Unshielded detector at +1200V |
05 |
No |
24052 |
60 |
|
2 |
Unshielded detector at +1200V |
15 |
No |
20684 |
60 |
|
3 |
Unshielded detector at +1200V |
35 |
No |
18324 |
60 |
|
4 |
Unshielded detector at +1200V |
55 |
No |
17266 |
60 |
|
5 |
Unshielded detector at +1200V |
75 |
No |
15643 |
60 |
|
6 |
Unshielded detector at +1200V |
95 |
No |
14223 |
60 |
|
7 |
Unshielded detector at +1200V |
115 |
No |
12792 |
60 |
|
8 |
Unshielded detector at +1200V |
135 |
No |
11691 |
60 |
|
9 |
Unshielded detector at +1200V |
155 |
No |
10414 |
60 |
|
10 |
Unshielded detector at +1200V |
175 |
No |
9245 |
60 |
|
11 |
Unshielded detector at +1200V |
195 |
No |
8574 |
60 |
|
12 |
Unshielded detector at +1200V |
215 |
No |
7665 |
60 |
|
13 |
Unshielded detector at +1200V |
235 |
No |
7008 |
60 |
|
14 |
Unshielded detector at +1200V |
255 |
No |
6386 |
60 |
|
1200V Source In Data |
|
Run Number |
Configuration |
Discriminator Threshold (mV) |
Cs-137 Button Source |
Counts |
Run Time (Sec) |
Count Rate (Hz)
|
1 |
Unshielded detector at +1200V |
05 |
Yes |
39450 |
60 |
|
2 |
Unshielded detector at +1200V |
15 |
Yes |
34061 |
60 |
|
3 |
Unshielded detector at +1200V |
35 |
Yes |
29029 |
60 |
|
4 |
Unshielded detector at +1200V |
55 |
Yes |
27672 |
60 |
|
5 |
Unshielded detector at +1200V |
75 |
Yes |
25935 |
60 |
|
6 |
Unshielded detector at +1200V |
95 |
Yes |
24104 |
60 |
|
7 |
Unshielded detector at +1200V |
115 |
Yes |
22223 |
60 |
|
8 |
Unshielded detector at +1200V |
135 |
Yes |
20670 |
60 |
|
9 |
Unshielded detector at +1200V |
155 |
Yes |
19168 |
60 |
|
10 |
Unshielded detector at +1200V |
175 |
Yes |
17920 |
60 |
|
11 |
Unshielded detector at +1200V |
195 |
Yes |
16383 |
60 |
|
12 |
Unshielded detector at +1200V |
215 |
Yes |
15066 |
60 |
|
13 |
Unshielded detector at +1200V |
235 |
Yes |
14130 |
60 |
|
14 |
Unshielded detector at +1200V |
255 |
Yes |
13382 |
60 |
|
1200V Plots |
|
Plateau measurements at 1200V, source in and source out overlaid. |
Plateau measurements at 1200V, source in minus source out.
|
1300V
1300V Source Out Data |
|
Run Number |
Configuration |
Discriminator Threshold (mV) |
Cs-137 Button Source |
Counts |
Run Time (Sec) |
Count Rate (Hz)
|
1 |
Unshielded detector at +1300V |
05 |
No |
24651 |
60 |
|
2 |
Unshielded detector at +1300V |
10 |
No |
22487 |
60 |
|
3 |
Unshielded detector at +1300V |
15 |
No |
21029 |
60 |
|
4 |
Unshielded detector at +1300V |
20 |
No |
20535 |
60 |
|
5 |
Unshielded detector at +1300V |
25 |
No |
19313 |
60 |
|
6 |
Unshielded detector at +1300V |
30 |
No |
18947 |
60 |
|
7 |
Unshielded detector at +1300V |
35 |
No |
18654 |
60 |
|
8 |
Unshielded detector at +1300V |
45 |
No |
18405 |
60 |
|
9 |
Unshielded detector at +1300V |
55 |
No |
18078 |
60 |
|
10 |
Unshielded detector at +1300V |
65 |
No |
17830 |
60 |
|
11 |
Unshielded detector at +1300V |
75 |
No |
17463 |
60 |
|
12 |
Unshielded detector at +1300V |
80 |
No |
17303 |
60 |
|
13 |
Unshielded detector at +1300V |
85 |
No |
17542 |
60 |
|
14 |
Unshielded detector at +1300V |
90 |
No |
17342 |
60 |
|
15 |
Unshielded detector at +1300V |
95 |
No |
17276 |
60 |
|
16 |
Unshielded detector at +1300V |
100 |
No |
17350 |
60 |
|
17 |
Unshielded detector at +1300V |
105 |
No |
16885 |
60 |
|
18 |
Unshielded detector at +1300V |
110 |
No |
16842 |
60 |
|
19 |
Unshielded detector at +1300V |
115 |
No |
16801 |
60 |
|
20 |
Unshielded detector at +1300V |
120 |
No |
16599 |
60 |
|
21 |
Unshielded detector at +1300V |
125 |
No |
16283 |
60 |
|
22 |
Unshielded detector at +1300V |
130 |
No |
16292 |
60 |
|
23 |
Unshielded detector at +1300V |
135 |
No |
16165 |
60 |
|
24 |
Unshielded detector at +1300V |
155 |
No |
15329 |
60 |
|
25 |
Unshielded detector at +1300V |
175 |
No |
14452 |
60 |
|
26 |
Unshielded detector at +1300V |
195 |
No |
14130 |
60 |
|
27 |
Unshielded detector at +1300V |
215 |
No |
12954 |
60 |
|
28 |
Unshielded detector at +1300V |
235 |
No |
12155 |
60 |
|
29 |
Unshielded detector at +1300V |
255 |
No |
11392 |
60 |
|
1300V Source In Data |
|
Run Number |
Configuration |
Discriminator Threshold (mV) |
Cs-137 Button Source |
Counts |
Run Time (Sec) |
Count Rate (Hz)
|
1 |
Unshielded detector at +1300V |
05 |
Yes |
36726 |
60 |
|
2 |
Unshielded detector at +1300V |
10 |
Yes |
33627 |
60 |
|
3 |
Unshielded detector at +1300V |
15 |
Yes |
31514 |
60 |
|
4 |
Unshielded detector at +1300V |
20 |
Yes |
30182 |
60 |
|
5 |
Unshielded detector at +1300V |
25 |
Yes |
29448 |
60 |
|
6 |
Unshielded detector at +1300V |
30 |
Yes |
28211 |
60 |
|
7 |
Unshielded detector at +1300V |
35 |
Yes |
27912 |
60 |
|
8 |
Unshielded detector at +1300V |
45 |
Yes |
26890 |
60 |
|
9 |
Unshielded detector at +1300V |
55 |
Yes |
26347 |
60 |
|
10 |
Unshielded detector at +1300V |
65 |
Yes |
26055 |
60 |
|
11 |
Unshielded detector at +1300V |
75 |
Yes |
25154 |
60 |
|
12 |
Unshielded detector at +1300V |
80 |
Yes |
25277 |
60 |
|
13 |
Unshielded detector at +1300V |
85 |
Yes |
25219 |
60 |
|
14 |
Unshielded detector at +1300V |
90 |
Yes |
25103 |
60 |
|
15 |
Unshielded detector at +1300V |
95 |
Yes |
25104 |
60 |
|
16 |
Unshielded detector at +1300V |
100 |
Yes |
25059 |
60 |
|
17 |
Unshielded detector at +1300V |
105 |
Yes |
24800 |
60 |
|
18 |
Unshielded detector at +1300V |
110 |
Yes |
24749 |
60 |
|
19 |
Unshielded detector at +1300V |
115 |
Yes |
24944 |
60 |
|
20 |
Unshielded detector at +1300V |
120 |
Yes |
24233 |
60 |
|
21 |
Unshielded detector at +1300V |
125 |
Yes |
24260 |
60 |
|
22 |
Unshielded detector at +1300V |
130 |
Yes |
23815 |
60 |
|
23 |
Unshielded detector at +1300V |
135 |
Yes |
23766 |
60 |
|
24 |
Unshielded detector at +1300V |
155 |
Yes |
23158 |
60 |
|
25 |
Unshielded detector at +1300V |
175 |
Yes |
22176 |
60 |
|
26 |
Unshielded detector at +1300V |
195 |
Yes |
21476 |
60 |
|
27 |
Unshielded detector at +1300V |
215 |
Yes |
20375 |
60 |
|
28 |
Unshielded detector at +1300V |
235 |
Yes |
19628 |
60 |
|
29 |
Unshielded detector at +1300V |
255 |
Yes |
18512 |
60 |
|
1300V Plots |
|
Plateau measurements at 1300V, source in and source out overlaid. |
Plateau measurements at 1300V, source in minus source out.
|
1400V
1400V Source Out Data |
|
Run Number |
Configuration |
Discriminator Threshold (mV) |
Cs-137 Button Source |
Counts |
Run Time (Sec) |
Count Rate (Hz)
|
1 |
Unshielded detector at +1400V |
05 |
No |
22873 |
60 |
|
2 |
Unshielded detector at +1400V |
10 |
No |
20892 |
60 |
|
3 |
Unshielded detector at +1400V |
15 |
No |
19899 |
60 |
|
4 |
Unshielded detector at +1400V |
20 |
No |
19398 |
60 |
|
5 |
Unshielded detector at +1400V |
25 |
No |
18861 |
60 |
|
6 |
Unshielded detector at +1400V |
30 |
No |
18587 |
60 |
|
7 |
Unshielded detector at +1400V |
35 |
No |
18053 |
60 |
|
8 |
Unshielded detector at +1400V |
45 |
No |
17563 |
60 |
|
9 |
Unshielded detector at +1400V |
55 |
No |
17356 |
60 |
|
10 |
Unshielded detector at +1400V |
65 |
No |
17347 |
60 |
|
11 |
Unshielded detector at +1400V |
75 |
No |
17329 |
60 |
|
12 |
Unshielded detector at +1400V |
80 |
No |
17164 |
60 |
|
13 |
Unshielded detector at +1400V |
85 |
No |
17146 |
60 |
|
14 |
Unshielded detector at +1400V |
90 |
No |
17268 |
60 |
|
15 |
Unshielded detector at +1400V |
95 |
No |
17234 |
60 |
|
16 |
Unshielded detector at +1400V |
100 |
No |
17231 |
60 |
|
17 |
Unshielded detector at +1400V |
105 |
No |
17119 |
60 |
|
18 |
Unshielded detector at +1400V |
110 |
No |
17119 |
60 |
|
19 |
Unshielded detector at +1400V |
115 |
No |
16956 |
60 |
|
20 |
Unshielded detector at +1400V |
120 |
No |
17135 |
60 |
|
21 |
Unshielded detector at +1400V |
125 |
No |
16680 |
60 |
|
22 |
Unshielded detector at +1400V |
130 |
No |
17093 |
60 |
|
23 |
Unshielded detector at +1400V |
135 |
No |
16893 |
60 |
|
24 |
Unshielded detector at +1400V |
155 |
No |
16909 |
60 |
|
25 |
Unshielded detector at +1400V |
175 |
No |
16489 |
60 |
|
26 |
Unshielded detector at +1400V |
195 |
No |
16340 |
60 |
|
27 |
Unshielded detector at +1400V |
215 |
No |
15683 |
60 |
|
28 |
Unshielded detector at +1400V |
235 |
No |
15707 |
60 |
|
29 |
Unshielded detector at +1400V |
255 |
No |
15411 |
60 |
|
1400V Source In Data |
|
Run Number |
Configuration |
Discriminator Threshold (mV) |
Cs-137 Button Source |
Counts |
Run Time (Sec) |
Count Rate (Hz)
|
1 |
Unshielded detector at +1400V |
05 |
Yes |
36948 |
60 |
|
2 |
Unshielded detector at +1400V |
10 |
Yes |
33637 |
60 |
|
3 |
Unshielded detector at +1400V |
15 |
Yes |
31769 |
60 |
|
4 |
Unshielded detector at +1400V |
20 |
Yes |
30477 |
60 |
|
5 |
Unshielded detector at +1400V |
25 |
Yes |
28817 |
60 |
|
6 |
Unshielded detector at +1400V |
30 |
Yes |
28523 |
60 |
|
7 |
Unshielded detector at +1400V |
35 |
Yes |
27931 |
60 |
|
8 |
Unshielded detector at +1400V |
45 |
Yes |
26824 |
60 |
|
9 |
Unshielded detector at +1400V |
55 |
Yes |
26639 |
60 |
|
10 |
Unshielded detector at +1400V |
65 |
Yes |
25810 |
60 |
|
11 |
Unshielded detector at +1400V |
75 |
Yes |
26154 |
60 |
|
12 |
Unshielded detector at +1400V |
80 |
Yes |
26157 |
60 |
|
13 |
Unshielded detector at +1400V |
85 |
Yes |
25888 |
60 |
|
14 |
Unshielded detector at +1400V |
90 |
Yes |
25679 |
60 |
|
15 |
Unshielded detector at +1400V |
95 |
Yes |
25026 |
60 |
|
16 |
Unshielded detector at +1400V |
100 |
Yes |
25533 |
60 |
|
17 |
Unshielded detector at +1400V |
105 |
Yes |
25389 |
60 |
|
18 |
Unshielded detector at +1400V |
110 |
Yes |
25223 |
60 |
|
19 |
Unshielded detector at +1400V |
115 |
Yes |
24894 |
60 |
|
20 |
Unshielded detector at +1400V |
120 |
Yes |
25054 |
60 |
|
21 |
Unshielded detector at +1400V |
125 |
Yes |
25116 |
60 |
|
22 |
Unshielded detector at +1400V |
130 |
Yes |
25117 |
60 |
|
23 |
Unshielded detector at +1400V |
135 |
Yes |
24980 |
60 |
|
24 |
Unshielded detector at +1400V |
155 |
Yes |
24495 |
60 |
|
25 |
Unshielded detector at +1400V |
175 |
Yes |
24411 |
60 |
|
26 |
Unshielded detector at +1400V |
195 |
Yes |
24345 |
60 |
|
27 |
Unshielded detector at +1400V |
215 |
Yes |
23909 |
60 |
|
28 |
Unshielded detector at +1400V |
235 |
Yes |
23306 |
60 |
|
29 |
Unshielded detector at +1400V |
255 |
Yes |
23191 |
60 |
|
1400V Plots |
|
Plateau measurements at 1400V, source in and source out overlaid. |
Plateau measurements at 1400V, source in minus source out.
|
Go Back
NaI_JD-766