Calibration Info
Jump to navigation
Jump to search
To fully understand the relationship between PMT Counts and total exposure for the Nanodot OSLs, a custom calibration is needed. This calibration is used in lieu of the calibration created by the OSL reader, which is created using the pre-dosed OSLs from the manufacturer. To begin the calibration, a set of fifteen unexposed Nanodot OSLs is chosen at random and exposed to a 9.3Ci
source. Using a distance of 30cm from the faceplate of the source and also including the distance from the faceplate to the surface of the source, , the exposure rate is calculated in Roentgen using . A gamma factor of and activity of is used for these calculations. Using the exposure rate found in the previous calculation, it is possible to find the total exposure of the Nanodot OSL by integrating the exposure rate over the exposure time, .