Calibration Info

From New IAC Wiki
Revision as of 23:28, 15 May 2018 by Harpconn (talk | contribs)
Jump to navigation Jump to search

To fully understand the relationship between PMT Counts and total exposure for the Nanodot OSLs, a custom calibration is needed. This calibration is used in lieu of the calibration created by the OSL reader, which is created using the pre-dosed OSLs from the manufacturer. To begin the calibration, a set of fifteen unexposed Nanodot OSLs is chosen at random and exposed to a 9.3Ci 137Cesium source. Using a distance of 30cm from the faceplate of the source and also including the distance from the faceplate to the surface of the source, D=(Distance to faceplate+11.2cm)=41.2cm , the exposure rate is calculated in Roentgen using ˙R=ΓAD2. A gamma factor of Γ=0.33(m2)(R)(Ci)(hr) and activity of A=9.3Ci is used for these calculations. Using the exposure rate calculation, it is possible to find the total exposure by integrating over time, tft0˙R dt.





Click here for calibration data