Dose Calculations and 137Cs Shot Log

From New IAC Wiki
Jump to navigation Jump to search

Source Calibration and Dose Rate Data

Date Initial Source Strength (Ci) Current Source Strength (Ci) Distance to Faceplate (cm) Exposure Rate (Roentgen/hr)°
6/27/17 20 9.3 30.5 17.691
9/11/17 20 9.3 30.5 17.606

°Calculated by TSO using [math]\dot R = \frac {\Gamma A }{ D^2} [/math] where [math]\boldsymbol{\Gamma} = 3.3 \frac {(cm^2)(R)}{(mCi)(hr)} [/math] , [math]\mathbf{A} = 9.3 Ci [/math] , [math]\mathbf{D}=(\text{Distance to faceplate} + 11.2cm) [/math]

There is an adjustment distance of 11.2cm which accounts for the distance from the faceplate to the [math] ^{137}Cs[/math] source

137Cs Shot Log

Dot OSLs

Date Run Number Run Time (s) Distance to Faceplate Number of OSLs in Stack Total Exposed Dose (R) Measured Dose Adjusted Dose
9/8/17 1 5660 30cm 11 108536 - -
9/8/17 1 5660 30cm 11 108536 - -
9/26/17 2 4500 30cm 11 - - -
9/26/17 2 4500 30cm 11 - - -

Total dose calculated by [math]\int\limits_{t_0}^{t_f}\dot R\ dt [/math]

Nanodot OSL Quality Control

Date Run Number Run Time (s) Distance to Faceplate Number of OSLs Total Exposed Dose (R) Measured Dose Adjusted Dose
10/12/17 1 284.04 30.5cm 3 5242.02 - -

Total dose calculated by [math]\int\limits_{t_0}^{t_f}\dot R\ dt [/math]



Reproducibility with 137Cs