Below are the theoretical calculations for the theoretical decay frequencies
Na-22, 10.19 uCi on July 15, 2013, half life 2.602 +/- 0.002 years, 99.937% for 1274.52 keV line and 178.8 for 511 keV line , activity in Feb 4, 2017 = 3.947 uCi
- = [math]\left (0.99937 \right )\left ( 3.947 \times 10^{-6} \mbox{Ci} \right) \left (\frac{ (3.7 \times 10^{10} \mbox{Hz}}{\mbox{Ci}} \right)= 145,946 Hz [/math] for the 1274 keV line
- = [math]\left (1.788 \right )\left ( 3.947 \times 10^{-6} \mbox{Ci} \right) \left (\frac{ (3.7 \times 10^{10} \mbox{Hz}}{\mbox{Ci}} \right)= 261,117.732 Hz [/math] for the 511 keV line
Now compute the expected rate for February 5, 2017.
The activity on Feb 5, 2017 is 3.944 uCi, so the expected rates for the 2 lines are
- = [math]\left (0.99937 \right )\left ( 3.944 \times 10^{-6} \mbox{Ci} \right) \left (\frac{ (3.7 \times 10^{10} \mbox{Hz}}{\mbox{Ci}} \right)= 145,836 Hz [/math] for the 1274 keV line
- = [math]\left (1.788 \right )\left ( 3.944 \times 10^{-6} \mbox{Ci} \right) \left (\frac{ (3.7 \times 10^{10} \mbox{Hz}}{\mbox{Ci}} \right)= 260919.264 Hz [/math] for the 511 keV line
Now compute the expected rate for February 9, 2017
The activity on Feb 9, 2017 is 3.933 uCi, so the expected rate for the 2 lines are
- = [math]\left (0.99937 \right )\left ( 3.933 \times 10^{-6} \mbox{Ci} \right) \left (\frac{ (3.7 \times 10^{10} \mbox{Hz}}{\mbox{Ci}} \right)= 145,429.32 Hz [/math] for the 1274 keV line
- = [math]\left (1.788 \right )\left ( 3.933 \times 10^{-6} \mbox{Ci} \right) \left (\frac{ (3.7 \times 10^{10} \mbox{Hz}}{\mbox{Ci}} \right)= 260,191.55 Hz [/math] for the 511 keV line
Now compute the expected rate for Feb 12, 2017
The activity on Feb 12, 2017 is 3.924 uCi, so the expected rate for the 2 lines are
- = [math]\left (0.99937 \right )\left ( 3.924 \times 10^{-6} \mbox{Ci} \right) \left (\frac{ (3.7 \times 10^{10} \mbox{Hz}}{\mbox{Ci}} \right)= 145096.5316 Hz [/math] for the 1274 keV line
- = [math]\left (1.788 \right )\left ( 3.924 \times 10^{-6} \mbox{Ci} \right) \left (\frac{ (3.7 \times 10^{10} \mbox{Hz}}{\mbox{Ci}} \right)= 259596.144 Hz [/math] for the 511 keV line
Geometric Efficiencies
Below is a table with the calculated efficiencies
Run |
Source |
Energy (keV) |
Position |
Expected Rate (Hz) |
ROOT Window |
HpGe Rate (Hz) |
HpGe X-Cool Detector Efficiency (%)
|
r5825 |
Na-22 |
511 |
5 |
|
r5836
|
r5827
|
r5828
|
r5829
|
r5830
|
r5831
|
r5832
|
r5833
|