Difference between revisions of "PAA 11-11-13"

From New IAC Wiki
Jump to navigation Jump to search
Line 123: Line 123:
 
| 6739 || Logic Unit in OR for Y-88 sample # 130113-004, post amp at '''20''',  rate is about 540 Hz ,HpGe discriminator at 12 mV and NaI at 12 mV.
 
| 6739 || Logic Unit in OR for Y-88 sample # 130113-004, post amp at '''20''',  rate is about 540 Hz ,HpGe discriminator at 12 mV and NaI at 12 mV.
 
|-
 
|-
| 6740 || Logic Unit in AND for Y-88 sample # 130113-004, post amp at '''20''',  rate is about 540 Hz ,HpGe discriminator at 12 mV and NaI at 12 mV.
+
| 6741 || Logic Unit in AND for Y-88 sample # 130113-004, post amp at '''20''',  rate is about 540 Hz ,HpGe discriminator at 12 mV and NaI at 12 mV.
 
|-
 
|-
 
|}
 
|}

Revision as of 22:12, 14 November 2013

Irradiated Y-88 and Au=186 for 4 hours at 23 MeV e- energy and


Samples place in Bin A60


Most upstream Ni target is # 130113-001 Ni target between Y-88 and Au is # 130113-002

Au-186 sample is 130113-003 Y-88 sample is 130113-004


Au-196 measurement using Detector D in position F and MPA 181775.mpa Y-88 measurement using Detector D in position F and MPA 181776.mpa

runlist

Poor Coincidence setup

Run Description
6686 Logic Unit in OR for Y-88 sample 130113-004
6687 Logic Unit in OR for Au-196 sample 130113-004, rate too high for DAQ
6688 Logic Unit in AND for Au-196 sample 130113-004, using Detector B and NaI, rate was 10 Hz, can't see 80 keV peak so I will increase HpGe post amp.
6689 Logic Unit in AND for Au-196 sample 130113-004, increase post amp from 20 to 50, 15 hrs 3x min 51 sec
6690 Logic Unit in OR for Ns-22 sample # 129741, post amp at 50, only 511 peak because postamp has 1 MeV peak in overflow
6691 Logic Unit in OR for Cs-137 sample # 124840, post amp at 50
6692 Logic Unit in OR for Ba-133 sample # 129789, post amp at 50
6693 Logic Unit in OR for Mn-54 sample # 129804, post amp at 50
6694 Changed post amp on HpGe Logic Unit in OR for Mn-54 sample # 129804, post amp at 20, 4 min 37 sec
6695 Logic Unit in OR for Co-60 sample # 129737, post amp at 20,
6696 Logic Unit in OR for Na-22 sample # 129741, post amp at 20,
6697 Logic Unit in OR for Cs-137 sample # 124840, post amp at 20,
6698 Logic Unit in OR for Ba-133 sample # 129789, post amp at 20, 4 min 44 sec
6699 Logic Unit in OR for Ni-57 sample # 130113-001, post amp at 20, 5 min 42 sec
6700 Logic Unit in OR for Ni-57 sample # 130113-002, post amp at 20, 5 min 2 sec
6701 Logic Unit in OR for Y-88 sample # 130113-004, post amp at 20, min sec
6702 Logic Unit in COIN for Y-88 sample # 130113-004, post amp at 20, min sec, rate is about 0.21 Hz, ended early to run nickel
6703 Logic Unit in COIN for Ni-57 sample # 130113-002, post amp at 20, rate is about 0.4 Hz, 22 hrs 55 min 8 sec, TDC16-TDC19 is not peaked around zero , the NIM2ECL translator for TDC19 does not have a square output, the problem was due to using the wrong CFD output
6704 Logic Unit in OR for Ni-57 sample # 130113-002, post amp at 20, rate is about 883 Hz, 4 min 0 sec
6705 Logic Unit in OR for Y-88 sample # 130113-004, post amp at 20, rate is about 770 Hz, 4 min 0 sec,
6706 Logic Unit in OR for Ni-57 sample # 130113-001, post amp at 20, rate is about 902 Hz, 4 min 5sec, Using correct NaI CFD channel, TDC19 should be good now will check with a coincidence run using Co-60 after taking singles measurements on the four irradiation samples
6707 Logic Unit in OR for Au-196 sample # 130113-003, post amp at 20, rate is about 4500 Hz, 37 sec,
6708 Logic Unit in AND for Au-196 sample # 130113-003, post amp at 50, rate is about 11 Hz
6709 Logic Unit in AND for Au-196 sample # 130113-003, post amp at 50, rate is about 150 Hz changed output width of CFDs from 500 ns to be about 1 usec.
6710 Logic Unit in AND for Au-196 sample # 130113-003, post amp at 50, rate is about 150 Hz inserted 1 usec delay on HpGe timing., TDC16-19 peak is -110 with an RMS of 22.5 nsec, adjusted delay near end of run
6711 Logic Unit in AND for Au-196 sample # 130113-003, post amp at 50, rate is about 150 Hz inserted 1 usec delay on HpGe timing., TDC16-19 peak is -60 with an RMS of 22.5 nsec, 51 min 38 sec
6712 Logic Unit in OR for Au-196 sample # 130113-003, post amp at 50, rate is about 4500 Hz
6713 Logic Unit in OR for Au-196 sample # 130113-003, post amp at 50, rate is about 4500 Hz
6714 Logic Unit in OR for Au-196 sample # 130113-003, post amp at 20, rate is about 140 Hz
6715 Logic Unit in AND for Au-196 sample # 130113-003, post amp at 20, rate is about 140 Hz
6716 Logic Unit in AND for Ni-57 sample # 130113-002, post amp at 20, rate is about 10 Hz,
6717 Logic Unit in OR for Ni-57 sample # 130113-002, post amp at 20, rate is about 868 Hz,
6718 Logic Unit in OR for Ni-57 sample # 130113-002, post amp at 50, rate is about 860 Hz,
6719 Logic Unit in AND for Au-196 sample # 130113-003, post amp at 50, take a longer coin run to check if the 188 keV line for the 9.6 hour half life of Au is still measurable after 2 days compare to run 6689 which is mostly accidentals from coincidence mode, ~3 hrs 21 min
6720 Logic Unit in AND for Y-88 sample # 130113-004, post amp at 50, 6 Hz
6721 Logic Unit in AND for Y-88 sample # 130113-004, post amp at 20, 17 hrs 56 min 48 sec, some problem with the HpGe spectrum, the NaI looks fine
6722 -
runs 5703 to 6722 are not useful for coincidence, the delay inserted into the HpGe CFD would ocaisionally more the HpGe singal outside the ADC gate.
6724 Logic Unit in OR for Y-88 sample # 130113-004, post amp at 20, sec, I see peaks in the HpGe histogram, now look at timing
6724 Logic Unit in OR for Y-88 sample # 130113-004, post amp at 20, sec, I see peaks in the HpGe histogram, now look at timing
6729 Logic Unit in AND for Na-22 source # 129743, post amp at 20, rate is about 32 Hz changed output width of CFDs from 1000 ns to be about 500 nsec., HpGe discriminator at 50 mV and NaI at 30 mV
6730 Logic Unit in OR for Ba-133 source # 129789, post amp at 20, rate is about 350 Hz .
6731 Logic Unit in OR for Ba-133 source # 129789, post amp at 50, rate is about 1000 Hz ,HpGe discriminator at 10 mV and NaI at 10 mV. can see ~80 keV peak from Ba-133
6732 Logic Unit in OR for Ba-133 source # 129789, post amp at 50, rate is about 800 Hz ,HpGe discriminator at 20 mV and NaI at 20 mV.
6733 Logic Unit in OR for Ba-133 source # 129789, post amp at 50, rate is about 900 Hz ,HpGe discriminator at 12 mV and NaI at 12 mV.
6734 Logic Unit in OR for Mn-54 source # 129804, post amp at 50, rate is about 540 Hz ,HpGe discriminator at 12 mV and NaI at 12 mV.
6735 Logic Unit in OR for Ni-57 sample # 130113-001, post amp at 50, rate is about 650 Hz ,HpGe discriminator at 12 mV and NaI at 12 mV.
6736 Logic Unit in OR for Ni-57 sample # 130113-002, post amp at 50, rate is about 676 Hz ,HpGe discriminator at 12 mV and NaI at 12 mV.
6737 Logic Unit in OR for Au-196 sample # 130113-003, post amp at 50, rate is about 3670 Hz ,HpGe discriminator at 12 mV and NaI at 12 mV.
6738 Logic Unit in OR for Y-88 sample # 130113-004, post amp at 50, rate is about 608 Hz ,HpGe discriminator at 12 mV and NaI at 12 mV.
6739 Logic Unit in OR for Y-88 sample # 130113-004, post amp at 20, rate is about 540 Hz ,HpGe discriminator at 12 mV and NaI at 12 mV.
6741 Logic Unit in AND for Y-88 sample # 130113-004, post amp at 20, rate is about 540 Hz ,HpGe discriminator at 12 mV and NaI at 12 mV.
 Somehow at delay switch was flipped up to the 1 usec position instead of the 0.1 usec position on the HpGe 
Hazards of taking data in a common use environment I guess
I didn't notice it before because the HpGe output was barely making it into the gate

Improved coincidence setup

Run Description
Logic Unit in AND for Na-22 sample # 130113-003, post amp at 20, rate is about 32 Hz changed output width of CFDs from 500 ns to be about 1 usec.




PAA_CoincidenceCounting#8.2F22.2F13_Activation