The power supply is connected to a discharge protection circuit. CAEN N470 has a signal Lemo 00 input to activate a kill option automatically when a discharge signal occurs through the detector operation, the kill input is connected to the detector readout plate output,  as a discharge happens and it is wider than 15 us, kill option state turns true, so it will trip all the power supply channels.
 
The power supply is connected to a discharge protection circuit. CAEN N470 has a signal Lemo 00 input to activate a kill option automatically when a discharge signal occurs through the detector operation, the kill input is connected to the detector readout plate output,  as a discharge happens and it is wider than 15 us, kill option state turns true, so it will trip all the power supply channels.
 
= Data Acquisition (DAQ)=
 
 
== LDS Using CAEN V792 QDC==
 
 
 
[[File:LDS_electronics_flow_chart.png |200px]]
 
 
 
Go Back [https://wiki.iac.isu.edu/index.php/Performance_of_THGEM_as_a_Neutron_Detector#Detector_Description]
 

Revision as of 23:42, 17 August 2013

Detector operation

Operating the detector occurs by providing voltage to the HV-circuit connected to the detector components mentioned in the previous section. A CAEN N470 four channel power supply is used for this purpose, using a channel of the power supply allow to increase the voltage difference between the top and bottom side of the first GEM preamplifier reaches up to 350 V then it decreases for the third GEM preamplifier, as the circuit board is provided with a voltage of 2.8 kV.

Another channel of the power supply is used to increase the voltage of the Cathode. When the cathode voltage is 3.1 kV, it will create a drift voltage up to 300V.


Abdel GEM HV distrib 6-19-09.png

The power supply is connected to a discharge protection circuit. CAEN N470 has a signal Lemo 00 input to activate a kill option automatically when a discharge signal occurs through the detector operation, the kill input is connected to the detector readout plate output, as a discharge happens and it is wider than 15 us, kill option state turns true, so it will trip all the power supply channels.