(Created page with 'Oportunistic Beam studies =Trigger Setup= 1.) The V895 was double pulsing and its output pulse maxed out at 40 ns {| border="1" |cellpadding="20" cellspacing="0 |- |[[Ima…') |
|||
Line 36: | Line 36: | ||
| 4073 || sample || 240 || off || empty || (Det1 OR Det2) || 200 || 2.0/0.888 || 2.7/1.2 || || | | 4073 || sample || 240 || off || empty || (Det1 OR Det2) || 200 || 2.0/0.888 || 2.7/1.2 || || | ||
|- | |- | ||
− | | 4074 || sample || || off/tuning || empty || (Det1 AND Det2) || 50 || 2.2/0.888 || 3.0/1.2 || || | + | | 4074 || sample || || off/tuning || empty || (Det1 AND Det2) || 50 || 2.2/0.888 || 3.0/1.2 || || |
|- | |- | ||
| 4075 || sample || || off/tuning || Chromox || (Det1 AND Det2 AND SYNC) || 50 || 2.2/0.888 || 3.0/1.2 || || | | 4075 || sample || || off/tuning || Chromox || (Det1 AND Det2 AND SYNC) || 50 || 2.2/0.888 || 3.0/1.2 || || | ||
Line 49: | Line 49: | ||
| 4081 || sample || 540 sec || e+(3 na)|| Chromox || (Det1 AND Det2 AND SYNC)** || 50 || 2.2/0.888 || 3.0/1.2 || || | | 4081 || sample || 540 sec || e+(3 na)|| Chromox || (Det1 AND Det2 AND SYNC)** || 50 || 2.2/0.888 || 3.0/1.2 || || | ||
|- | |- | ||
− | | 4082 || sample || 23 min || e+(0.3 na)|| Chromox || (Det1 AND Det2 AND SYNC)** || 50 || 2.2/0.888 || 3.0/1.2 || || | + | | 4082 || sample || 23 min || e+(0.3 na)|| Chromox || (Det1 AND Det2 AND SYNC)** || 50 || 2.2/0.888 || 3.0/1.2 || || |
|- | |- | ||
− | | 4083 || sample || sec || e+(0.3 na)|| Chromox || (Det1 AND Det2) || 50 || 2.2/0.888 || 3.0/1.2 || || | + | | 4083 || sample || sec || e+(0.3 na)|| Chromox || (Det1 AND Det2) || 50 || 2.2/0.888 || 3.0/1.2 || || |
|- | |- | ||
| 4084 || sample || sec || e+(0.150 na)|| Chromox || (Det1 AND Det2) || 50 || 2.2/0.888 || 3.0/1.2 || || Changed discriminator to 50 mv for both channels it was at 20 the scalers will count differently now but the trigger theshold is still the same 100 mV | | 4084 || sample || sec || e+(0.150 na)|| Chromox || (Det1 AND Det2) || 50 || 2.2/0.888 || 3.0/1.2 || || Changed discriminator to 50 mv for both channels it was at 20 the scalers will count differently now but the trigger theshold is still the same 100 mV | ||
Line 57: | Line 57: | ||
| 4085 || sample || sec || e+(0.15 na)|| Chromox || (Det1 AND Det2 AND SYNC) || 50 || 2.2/0.888 || 3.0/1.2 || || | | 4085 || sample || sec || e+(0.15 na)|| Chromox || (Det1 AND Det2 AND SYNC) || 50 || 2.2/0.888 || 3.0/1.2 || || | ||
|- | |- | ||
− | | 4086 || sample || sec || e+(0.05 na)|| Chromox || (Det1 AND Det2) || 50 || 2.2/0.888 || 3.0/1.2 || || | + | | 4086 || sample || sec || e+(0.05 na)|| Chromox || (Det1 AND Det2) || 50 || 2.2/0.888 || 3.0/1.2 || || |
|} | |} | ||
Line 72: | Line 72: | ||
Ch4(Green)=Annih. Det. #2 | Ch4(Green)=Annih. Det. #2 | ||
− | |||
− | |||
=Data Analysis= | =Data Analysis= |
Revision as of 06:27, 25 March 2012
Oportunistic Beam studies
Trigger Setup
1.) The V895 was double pulsing and its output pulse maxed out at 40 ns
![]() |
The V895 "Maj" output is shown on Ch1 along with the NaI signal which was discriminated by it |
2.) inserted a gate delay generator to change the V895 output pulse to a microsecond
![]() |
I took the "MAJ" output and used a Phillips Gate Generator to extend the pulse length in order to stop multiple triggers |
Pre-Beam DAT test runs
Run# | run mode | Run length (sec) | Beam | Target | Trig Logic | VME discriminator Threshold (mV) | HV1 (setting/V) A1 | HV2 (setting/V) A3 | PL (setting/ns) | PTW (setting/ns) |
4073 | sample | 240 | off | empty | (Det1 OR Det2) | 200 | 2.0/0.888 | 2.7/1.2 | ||
4074 | sample | off/tuning | empty | (Det1 AND Det2) | 50 | 2.2/0.888 | 3.0/1.2 | |||
4075 | sample | off/tuning | Chromox | (Det1 AND Det2 AND SYNC) | 50 | 2.2/0.888 | 3.0/1.2 | |||
4078 | sample | e+ (0.3 nA) | Chromox | (Det1 AND Det2 AND SYNC)** | 50 | 2.2/0.888 | 3.0/1.2 | |||
4079 | sample | e+ (0.3nA) | Chromox | (Det1 AND Det2)** | 50 | 2.2/0.888 | 3.0/1.2 | |||
4080 | sample | e+(1 na) | Chromox | (Det1 AND Det2 AND SYNC)** | 50 | 2.2/0.888 | 3.0/1.2 | |||
4081 | sample | 540 sec | e+(3 na) | Chromox | (Det1 AND Det2 AND SYNC)** | 50 | 2.2/0.888 | 3.0/1.2 | ||
4082 | sample | 23 min | e+(0.3 na) | Chromox | (Det1 AND Det2 AND SYNC)** | 50 | 2.2/0.888 | 3.0/1.2 | ||
4083 | sample | sec | e+(0.3 na) | Chromox | (Det1 AND Det2) | 50 | 2.2/0.888 | 3.0/1.2 | ||
4084 | sample | sec | e+(0.150 na) | Chromox | (Det1 AND Det2) | 50 | 2.2/0.888 | 3.0/1.2 | Changed discriminator to 50 mv for both channels it was at 20 the scalers will count differently now but the trigger theshold is still the same 100 mV | |
4085 | sample | sec | e+(0.15 na) | Chromox | (Det1 AND Det2 AND SYNC) | 50 | 2.2/0.888 | 3.0/1.2 | ||
4086 | sample | sec | e+(0.05 na) | Chromox | (Det1 AND Det2) | 50 | 2.2/0.888 | 3.0/1.2 |
- stop using V895 and used a discriminator (Phillips 708 set to ~100 mV) and logic unit (phillips 758)
Scope pict traces for TEK0020 -> TEK00
Ch1(Yellow)=Majority
Ch2(Blue)=Beam SYNC
Ch3(Pink)=Annih. Det. #1
Ch4(Green)=Annih. Det. #2
Data Analysis
Replay
running the decoder using the personal account on the opsmdaq0 failed; perhaps due to problems writing its output to the /data disk
to convert a data file into a rood file I needed to login to opsdaq0 under the username "idaq"
then execute the following command
/work1/fanchini/decoder/fadc_decode_Source /data/annihilation/Annihilation_Sample_4086.dat
This created the summary text file and root file for run 4086
MakeClass
Load up the ntuple and click over to the "T" tree
Then onthe command line type
T->MakeClass();