Difference between revisions of "PePPO 03132012"
Jump to navigation
Jump to search
Line 137: | Line 137: | ||
[[File:PePPO-03132012-pulses-2905-int_14-Hist.png | 400 px]] | [[File:PePPO-03132012-pulses-2905-int_14-Hist.png | 400 px]] | ||
+ | |||
+ | ==TF Analysis== | ||
+ | |||
+ | |||
+ | Using the following ROOT programs | ||
+ | |||
+ | 03212012_T.C.txt | ||
+ | |||
+ | 03212012_T.h.txt | ||
+ | |||
+ | |||
+ | |||
+ | {| border="1" |cellpadding="20" cellspacing="0 | ||
+ | |- | ||
+ | | Run#|| run mode|| Run length (sec)||Beam||Target ||Trig Logic|| VME discriminator Threshold (mV)|| HV1 (setting/V) || HV2 (setting/V) || Deadtime || | ||
+ | |- | ||
+ | | 2902|| sample || 0 || e<math>^-</math> || Chromox || (Det1 OR Det2) || 50 || 2.4/960 || 3/1200 || 93.1818% <math>\pm</math> 10.3874% || | ||
+ | |- | ||
+ | | 2903|| sample || 203 ||e<math>^-</math> || Empty || (Det1 OR Det2) || 50 || 2.4/960 || 3/1200 || 84.9788% <math>\pm</math>0.106622% || | ||
+ | |- | ||
+ | | 2904|| sample || 452 ||e<math>^-</math> || Empty || Det1 AND Det2 || 50 || 2.4/960 || 3/1200 || 12.2563% <math>\pm</math> 2.32772% || | ||
+ | |- | ||
+ | | 2905|| sample || 556 || e<math>^-</math> || Chromox || Det1 AND Det2 || 50 || 2.4/960 || 3/1200 || 3.16901% <math>\pm</math> 4.2301% || | ||
+ | |- | ||
+ | | 2908|| sample || 150 || off || out || (Det1 or Det2) || 50 || 2.4/960 || 3/1200 || 74.0667% <math>\pm</math> 0.157206% || | ||
+ | |- | ||
+ | | 2909|| sample || 1118 || off || out || (Det1 AND Det2) || 50 || 2.4/960 || 3/1200 || 3.57143% <math>\pm</math>26.9725% || | ||
+ | |- | ||
+ | |||
+ | |} |
Revision as of 02:56, 22 March 2012
Accelerator Configuration
- Accelerator Settings
beam energy | 6.8 MeV |
Beam type: | Electron |
I_average: | 0.3 nA |
I_peak: | 0.05 | A 0.006 duty factor
pulse width: | 10 | s (it seems on 200 s is possible?)
Rate: | 60 Hz |
PEPPO Spectrometer: | 151.3 A |
Set HV to Detectors
Signal channel | HV dac | HV (V) | |
Detectors #1 | A1 | 2.4 | 960 |
Detectors #2 | A2 | 3.0 | 1200 |
Scope Pictures
CH1: Beam Sync CH2: A1 CH3: A2
Target | Pictures |
No Target | |
Chromox Target |
Annihilation Counter Runs
- Signal Definitions for Trigger logic
- SYNC:=60 Hz sync pulse 200 micro seconds long
- Det1:= Annihilation Counter #1 (HV1)
- Det2:=Annihilation Counter #2 (HV2)
data analysis
Negative Energy ?
This run 2902, Int_Ch14 histogram. Why there is negative?
Pulses
run 2905
1st 100 puleses
pulse that only partially read-in:
first 10 pulses, some pulses that only partially read-in:
looks like some pulses are only partially read-in. We need to delay signal more?
One possible cut may to determine if the signal is increasing within the first 100 - 300 $Iterations. This would be like creating a leading edge discriminator in software.
T->Draw("CH14:Iteration$","Iteration$<300 & CH14<100","L",10,1)
looks like it just won't draw. I need to know the to way exclude that event does not satisfy "Iteration$<300 & CH14<100" requirement.
Why CH14 histogram is all positive, but int_14 histogram has negatives?
TF Analysis
Using the following ROOT programs
03212012_T.C.txt
03212012_T.h.txt
Run# | run mode | Run length (sec) | Beam | Target | Trig Logic | VME discriminator Threshold (mV) | HV1 (setting/V) | HV2 (setting/V) | Deadtime | |
2902 | sample | 0 | e | Chromox | (Det1 OR Det2) | 50 | 2.4/960 | 3/1200 | 93.1818% | 10.3874%|
2903 | sample | 203 | e | Empty | (Det1 OR Det2) | 50 | 2.4/960 | 3/1200 | 84.9788% | 0.106622%|
2904 | sample | 452 | e | Empty | Det1 AND Det2 | 50 | 2.4/960 | 3/1200 | 12.2563% | 2.32772%|
2905 | sample | 556 | e | Chromox | Det1 AND Det2 | 50 | 2.4/960 | 3/1200 | 3.16901% | 4.2301%|
2908 | sample | 150 | off | out | (Det1 or Det2) | 50 | 2.4/960 | 3/1200 | 74.0667% | 0.157206%|
2909 | sample | 1118 | off | out | (Det1 AND Det2) | 50 | 2.4/960 | 3/1200 | 3.57143% | 26.9725%