Because is something wrong I decided to measure the counts rate as function of threshold energy. Maybe my threshold was incorrect.
+
By Dr. Dale suggestions to make threshold adjusting of n's detectors according to <math>(/mbox{RATE_{source}} - <math>/mbox{RATE_{source}})</math> as function of threshold I did the following test.
  
Below are my results for n's detector with HAMAMATSU PMT (-1500 V):
+
Set up like that:
  
[[File:Det1 callibration.png | 1000 px]]
+
[[File:Detector1.png | 400px]]
  
  
 
+
The voltages, thresholds and widths  are:
So it looks like my previous threshold value (-500 mV) was not good. '''Let's set it up the -200 mV.''' And let's see does the n's detector sensitive to the source. The same set up and distances as before.
 
 
{| border="1" cellspacing="0" style="text-align: center; width: 800px; height: 100px;"
 
{| border="1" cellspacing="0" style="text-align: center; width: 800px; height: 100px;"
! scope="col" width="50" |  
+
! scope="col" width="50" |
! scope="col" width="50" | counts/sec (5 min test), no source
+
! scope="col" width="50" | PMT voltage, V
! scope="col" width="50" | counts/sec (5 min test), with source on the top of PS1
+
! scope="col" width="50" | threshold, mV
 +
! scope="col" width="50" | discr. width, ns
 +
|-
 +
!PS1  
 +
| -1000|| -100|| 50
 +
|-
 +
!PS1
 +
| -1000|| -100|| 50
 
|-
 
|-
 
!n's  
 
!n's  
|19.89<math>\pm</math>0.26  (total 5966 cnts) || 58.84<math>\pm</math>0.29  (total 17652 cnts)
+
| -1500|| varying||  10
 
|}
 
|}
 
So now we see the source. The ratio are:
 
 
<math>\frac{Rate_{source}}{Rate_{w/source}} = \frac{58.84}{19.89} = 2.96 </math>
 

Revision as of 21:16, 19 May 2011