Difference between revisions of "05/11/2011 neutron detector threshold calibration"
Jump to navigation
Jump to search
Line 1: | Line 1: | ||
− | [https://wiki.iac.isu.edu/index.php/N% | + | [https://wiki.iac.isu.edu/index.php/N%27s_detector_calibration Go Back] |
Revision as of 05:09, 14 May 2011
Because is something wrong I decided to measure the counts rate as function of threshold energy. Maybe my threshold was incorrect.
Below are my results for n's detector with HAMAMATSU PMT (-1500 V):
So it looks like my previous threshold value (-500 mV) was not good. Let's set it up the -200 mV. And let's see does the n's detector sensitive to the source. The same set up and distances as before.
counts/sec (5 min test), no source | counts/sec (5 min test), with source on the top of PS1 | |
---|---|---|
n's | 19.89 | 0.26 (total 5966 cnts)58.84 | 0.29 (total 17652 cnts)
So now we see the source. The ratio are: