[https://wiki.iac.isu.edu/index.php/N%27s_detector_calibration Go Back]
 
[https://wiki.iac.isu.edu/index.php/N%27s_detector_calibration Go Back]
  
 
+
Set up like that:
continue...the same set up as before, the distances are the same:
 
  
 
[[File:Detector1.png | 400px]]
 
[[File:Detector1.png | 400px]]

Revision as of 20:54, 13 May 2011

Go Back

Set up like that:

Detector1.png


First I change the power supply for HAMAMATSU PMT to the new 2 channel one C.A.E.N N471

The powers are:

PS1 and PS2:  -1000 V (the old 2-channel TENNELEC TC952)
n's: -1500 V (the new 2-channel C.A.E.N N471)

The thresholds are:

PS1 and PS2:  -80 mV
n's: -200 mV


Let's do first the coincidence test PS1+PS2

counts/min (1 min test) counts/min (1 min test) counts/min (5 min test) counts/min (5 min test) avg counts/min
PS1 187 160 166 160.8 168.4
PS2 187 170 180.4 184.4 180.4
n's 205 204 201.2 202.4 203.2