First I change the power supply for HAMAMATSU PMT to the new 2 channel one C.A.E.N N471
 
First I change the power supply for HAMAMATSU PMT to the new 2 channel one C.A.E.N N471
 
  
 
The powers are:
 
The powers are:
Line 20: Line 19:
 
  PS1 and PS2:  -80 mV
 
  PS1 and PS2:  -80 mV
 
  n's: -60 mV
 
  n's: -60 mV
 
  
  
Line 46: Line 44:
  
 
Interesting. The same power for HAMAMATSU PMT, but different counts result.
 
Interesting. The same power for HAMAMATSU PMT, but different counts result.
 +
 +
 +
 +
Now I increased the power supply for HAMAMATSU PMT and adjusted the threshold.
 +
 +
The powers are:
 +
 +
PS1 and PS2:  -1000 V (the old 2-channel TENNELEC TC952)
 +
n's: -1500 V (the new 2-channel C.A.E.N N471)
 +
 +
 +
The thresholds were:
 +
 +
PS1 and PS2:  -80 mV
 +
n's: -500 mV
 +
 +
 +
Let see what we have. First he CF-243 source is far far away (sealed in the safe):
 +
{| border="1" cellspacing="0" style="text-align: center; width: 800px; height: 100px;"
 +
! scope="col" width="50" |
 +
! scope="col" width="50" | counts/min (1 min test)
 +
! scope="col" width="50" | counts/min (1 min test)
 +
! scope="col" width="50" | counts/min (5 min test)
 +
! scope="col" width="50" | counts/min (5 min test)
 +
! scope="col" width="50" | avg counts/min
 +
|-
 +
!PS1
 +
|187||160||166||160.8||168.4
 +
|-
 +
!PS2
 +
|187||170||180.4||184.4||180.4
 +
|-
 +
!PS1+PS2
 +
|
 +
|-
 +
!n's
 +
|205||204||201.2||202.4||203.2
 +
|}

Revision as of 16:36, 10 May 2011

Go Back


continue...the same set up as before, the distances are the same:

Detector1.png


First I change the power supply for HAMAMATSU PMT to the new 2 channel one C.A.E.N N471

The powers are:

PS1 and PS2:  -1000 V (the old 2-channel TENNELEC TC952)
n's: -1100 V (the new 2-channel C.A.E.N N471)


The thresholds were:

PS1 and PS2:  -80 mV
n's: -60 mV


Let see what we have. First he CF-243 source is far far away (sealed in the safe):

counts/min (1 min test) counts/min (1 min test) counts/min (5 min test) counts/min (5 min test) avg counts/min
PS1 187 160 166 160.8 168.4
PS2 187 170 180.4 184.4 180.4
PS1+PS2
n's 205 204 201.2 202.4 203.2

Interesting. The same power for HAMAMATSU PMT, but different counts result.


Now I increased the power supply for HAMAMATSU PMT and adjusted the threshold.

The powers are:

PS1 and PS2:  -1000 V (the old 2-channel TENNELEC TC952)
n's: -1500 V (the new 2-channel C.A.E.N N471)


The thresholds were:

PS1 and PS2:  -80 mV 
n's: -500 mV


Let see what we have. First he CF-243 source is far far away (sealed in the safe):

counts/min (1 min test) counts/min (1 min test) counts/min (5 min test) counts/min (5 min test) avg counts/min
PS1 187 160 166 160.8 168.4
PS2 187 170 180.4 184.4 180.4
PS1+PS2
n's 205 204 201.2 202.4 203.2