Difference between revisions of "PePPO NaI PositronDetector 4 JLab"
Line 23: | Line 23: | ||
End Time: March 23: 13:10:47 | End Time: March 23: 13:10:47 | ||
+ | |||
+ | Elapsed time: 2421 seconds | ||
[[File:1979_NaI_1185Volts.png| 200px]] | [[File:1979_NaI_1185Volts.png| 200px]] |
Revision as of 23:43, 23 March 2011
Detector setup for a NaI positron detector for PePPO
NaI detector
The PMT is powered by a HV supply which should have a DC ripple less than 200 mV peak-to-peak at 3kV/3mA. A DC ripple will change the gain of the PMT and cause the energy resolution to deteriorate.
The Linearity of the detector system is another way to reduce the energy resolution. The measured PMT output should be linear with the input signal when the gain is constant. This linearity can be measured using 2 radiation sources.
Input HV: +1085 V
Input Current: 2.362 mA
Running CODA: Running CODA
Measurements Using V792 QDC
Run 1979
Start Time : March 23:12:30:26
End Time: March 23: 13:10:47
Elapsed time: 2421 seconds
Run 1980
Remove Na-22 source from top of NaI crystal.
Start Time : March 23:13:11:23
End Time: March 23: 15:56:40
Let's lower the gain of the Phillips 777 amplifier so the 511 peak is around channel 500. This way the 1 MeV photons may come on scale.
Gain
Linearity
Readout
F250
JLab's F250 flash ADC will be used to measure the NaI output