Difference between revisions of "HRRL 03-18-2011"
Jump to navigation
Jump to search
Line 1: | Line 1: | ||
+ | =Tune parameters= | ||
+ | |||
+ | {| border="1" | ||
+ | | Solenoid 1 || 6.8 A | ||
+ | |- | ||
+ | |Solenoid 2 || 10.4 A | ||
+ | |- | ||
+ | |Gun Ver || -0.2 A | ||
+ | |- | ||
+ | | Gun Hor || +0.4 A | ||
+ | |- | ||
+ | |Output Hor || -1.4A | ||
+ | |- | ||
+ | | Output Ver || -0.5 A | ||
+ | |||
+ | |- | ||
+ | | Gun HV || +9.75 (Knob Setting) | ||
+ | |- | ||
+ | | Gun Grid Voltage || 5.25 (Knob Setting) | ||
+ | |- | ||
+ | | RF frequency || 2855.816 MHz | ||
+ | |- | ||
+ | | Modulator HV Power Supply || 4.42 (Knob Setting) | ||
+ | |- | ||
+ | | RF macro Pulse Length (FWHM) || 200 ns | ||
+ | |- | ||
+ | | Peak Current on FC || 37.2 mAmps | ||
+ | |- | ||
+ | | Scope on FC || [[File:Scope_image_of_good_Accelarator_setting_by_Dr._Kim_Mar_17_2011.png | 100 px]] | ||
+ | |- | ||
+ | | e- Beam Energy || 14 MeV | ||
+ | |||
+ | |||
+ | |} | ||
= Quad Scans = | = Quad Scans = | ||
To test reproducibility from yesterday. Scan at 40, 20, 10, 5 mA peak currents. Scan twice on each Current. | To test reproducibility from yesterday. Scan at 40, 20, 10, 5 mA peak currents. Scan twice on each Current. |
Revision as of 17:32, 18 March 2011
Tune parameters
Quad Scans
To test reproducibility from yesterday. Scan at 40, 20, 10, 5 mA peak currents. Scan twice on each Current.
Do Energy Scan
Bend the beam through 45 bending dipole, take images on the scope of FC and YAG images.
D1 current | Energy | FC peak Volt | Pic |
(A) | MeV | mV | |
28 | 11.85 | 39 | |
29 | 12.3 | 74 | |
30 | 12.74 | 93 | |
30.5 | 12.96 | 65 | |
30.7 | large uncertainty | 22 |