Difference between revisions of "FC calibration (by view screen monitoring) 3/1/11"

From New IAC Wiki
Jump to navigation Jump to search
Line 2: Line 2:
  
  
[[File:FC Calibration 3.1.11.png | 1300 px]]
+
[[File:FC Calibration 3.1.11.png | 1200 px]]
  
  

Revision as of 21:52, 1 March 2011

Go Back


FC Calibration 3.1.11.png


Here are:

  1. red line is the fitting line for beam up
  2. green line is the fitting line for beam down
  3. black line is the fitting line for beam up and down

We can see that all three line is almost the same so we can use the fitting line for whole range of kicker magnet which is

[math]x(cm)= 0.059 + 0.028\cdot I(Ampers)[/math]


Go Back