= Comparison of old and new calibration coefficients =
 
= Comparison of old and new calibration coefficients =
 +
 +
We used ADC 1D for the run pf_10_10_08_00040
  
 
[[Image:Old_calibration.jpg |500 px]]
 
[[Image:Old_calibration.jpg |500 px]]

Revision as of 18:11, 7 November 2008

We used:

-detector A to calibrate all the TDCs;

-LeCroy octal discriminator to produce standard negative pulses;

-passive delay line to calibrate TDCs in the region < 100 ns;

-delay amplifier TENNELEC TC215 to calibrate TDCs in the region > 250 ns;

Range of all three TDCs was set to 50nsx100 (upper plot) 200nsx10 (lower plot).

ADC-TAC pair was chosen in such a way that we used during the experiment.

ORTEC TAC 567-02 and ADC 1D

567 02.jpg

567 02 2000.jpg

ORTEC TAC 567-01 and ADC 1A

567 01.jpg

567 01 2000.jpg

ORTEC TAC 566-01 and ADC 1C

566 01.jpg

566 01 2000.jpg

Comparison of old and new calibration coefficients

We used ADC 1D for the run pf_10_10_08_00040

Old calibration.jpg

The width of photon peak is about 10.3 ns for old calibration.

New calibration.jpg

The width of photon peak is about 32.3 ns for new calibration.